Precision Intra & Inter-assay:
Intra-assay Precision: Two samples with concentrations of 400 and 1500 pg/ml were tested 30 times within the same plate by the same operator.
Inter-assay Precision: The same samples were tested in 10 separate runs by at least three different operators.

Sensitivity:
Limit of detection (LOD) is 14,025 pg/ml, indicating the lowest concentration significantly different from a blank.

Assay Range:
The assay range is from 125 pg/ml to 3000 pg/ml, with ULOQ and LLOQ representing the highest and lowest limits of quantification, respectively.

Precaution:
This kit is for research use only and is not designed for diagnostic or therapeutic procedures.

Spike Recovery:
Recombinant Calponin 1 was spiked in serum at two concentrations over the assay range, showing recovery rates of 97.2% and 95% for samples of 400 and 1500 pg/ml, respectively.

Dilution Linearity:
Serum samples with high concentrations of Calponin 1 were serially diluted, exhibiting linear dilution between 2-30. Samples should be diluted to at least 1/2 (MRD = 2).

Population and Clinical Data:
100 serum samples from healthy donors (50 female, 50 male, aged 18-63) were evaluated. The population mean was 141 +/- 18.72 pg/mL.
See more

Catalog excerpts

Human Calponin 1 Sandwich ELISA Kit-1

SPECIFICATION SHEET Human Calponin 1 Sandwich ELISA Kit Calponin 1 measurement in Human serum samples ■ Sensitivity: Limit of detection (LOD) 14,025 pg/ml. Lowest concentration significantly different from a blank. ■ Assay Range : The assay range [LLOQ - ULOQ] : 125 pg/ml to 3000 pg/ml. ULOQ & LLOQ = Highest and Lowest limit of quantification respectively. ■ Precision Intra & Inter-assay : Firalis SAS - www.firalis.com - 35 rue du Fort - F-68330 - Huningue FRANCE +33 389 911 320 Intra-assay Precision : Two samples of known concentration (400 and 1500 pg/ml) were tested 30 time within the same plate and by the same operator. Inter-assay Precision : Two samples of known concentration (400 and 1500 pg/mi) were tested in 10 separate runs. Assays were performed by at least three different operators.

 Open the catalog to page 1
Human Calponin 1 Sandwich ELISA Kit-2

( FIRALIS BRIDGING SCIENCE TO CLINICS SPECIFICATION SHEETHuman Calponin 1 Sandwich ELISA Kit Calponin 1 measurement in Human serum samples ■ Dilution Linearity : Serum samples containing high concentrations of Calponin 1 were serially diluted with assay diluent. Samples exhibits a linear dilution between 2-30. Samples should be diluted to at least 1/2 (MRD = 2) ■ Spike Recovery : Recombinant Calponin 1 was spiked in serum (n=10) at two concentrations (Low, High) over the range of the assay. ■ Population and clinical data : 100 serum samples from apparently healthy donors (50 female + 50 male)...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.