Product catalog summary
Circuit Components and Composition
  • Pack Components: Includes a 1.1 m heated inspiratory limb, 0.3 m incubator extension, 1.6 m heated expiratory tube with Evaquatechnology, 0.6 m humidifier connection tube with nitric oxide port, swivel wye-piece, MR290V autofeed humidification chamber, pressure line, Dräger adaptors, and label kit change out.
  • Materials: Made from polypropylene, polycarbonate, thermoplastic elastomer, various polyethylene types, polyether ester elastomer, silicone, nylon, brass, styrene resin, cyanoacrylate, and DEHP-free PVC. Not made with natural rubber latex.
  • Quantity: 10 circuits per carton.
  • Carton Dimensions and Weight: 530 mm x 210 mm x 400 mm, weighing 3.99 kg.
  • Carton Material: Cardboard box.
  • Manufacturing Mode: Produced in a controlled working environment.
  • Disposal: Incineration or according to hospital protocol.
Performance Specifications
  • Resistance to Flow: Inspiratory limb - 0.09 ± 0.04 cmH₂O; Expiratory limb - 0.07 ± 0.04 cmH₂O @ 2.5 L/min.
  • Flow Rate: 4 - 15 L/min.
  • Circuit Length: Both inspiratory and expiratory limbs are 1.6 m.
  • Minimum Tube Internal Diameter: 10.2 mm.
  • Ambient Range: 20-26 °C / 68-79 °F.
  • Compliance: 1.30 ± 0.15 mL/cmH₂O @ 60 cmH₂O.
  • Humidifier Compatibility: MR850 humidifier, invasive mode.
  • Compressible Volume: 758 mL.
  • Duration of Use: 7 days, single patient use.
  • Recommended Gas Source: Air/Oxygen.
  • Interface Connections: ISO 5356-1 Conical Connectors.
  • Shelf Life: 5 years.
  • Gas Leakage: <30 mL/min @ 60 cmH₂O.
  • Storage Temperature: -10 °C to 50 °C.
  • Tidal Volume: <185 mL.
Chamber MR290
  • Interface Connections: ISO 5356-1 Conical connector (22 mm Male).
  • Maximum Chamber Operating Pressure: 8 kPa.
  • Maximum Peak Flow: 180 L/min for 30 secs.
  • Gas Leakage: < 10 mL/min @ 8 kPa.
  • Materials: ABS, Polystyrene, Polyethylene, Thermoplastic Elastomer, Aluminium, Ink, Polypropylene, Adhesive, Polycarbonate, Silicone, DEHP-free PVC.
  • Compliance: 0.4 mL/cmH₂O.
  • Compressible Volume: 280 mL.
  • Resistance to Flow: At 60 L/min: 0.52 cmH₂O.
Regulatory
  • Classification: AU IIa; EU IIa; CA II; USA II.
  • Country of Origin: New Zealand or Mexico.
  • Intended Use: For delivering respiratory gases to infant patients dependent on mechanical respiratory support.
  • Notified Body Identification Number: TÜV SÜD Product Services GmBH CE0123.
  • GTIN Number: 09420012430205 (EA), 09420012424365 (PAC).
  • UNSPSC Number: 42272224.
  • GDMN Code: 37706.
  • Biocompatibility: Meets standards ISO 10993-3, ISO 10993-5, ISO 10993-6, ISO 10993-10.

Evaqua and F&P are trademarks of Fisher & Paykel Healthcare Limited.

See more

Catalog excerpts

Evaqua™ 2-1

850 RT265 | SPECIFICATIONS Infant Ventilator Circuit (> 4 L/min) Dual Heated with MR290 Autofeed Chamber CIRCUIT COMPONENTS AND COMPOSITION Pack Components 1.1 m heated inspiratory limb, 0.3 m incubator extension, 1.6 m heated expiratory tube with Evaquatechnology, 0.6 m humidifier connection tube with nitric oxide port, swivel wye-piece, MR290V autofeed humidification chamber, pressure line, Drager adaptors, label kit change out Tidal Volume www.fphcare.com Fisher &Faykel H EALTHCARE

 Open the catalog to page 1
Evaqua™ 2-2

Evaqua and F&P are trademarks of Fisher & Paykel Healthcare Limited 619410 REV B © 2020 Fisher & Paykel Healthcare Limited

 Open the catalog to page 2

All Fisher & Paykel Healthcare catalogs and technical brochures

Archived catalogs

  1. HC150

    2  Pages

  2. OSA Catalogue

    70  Pages

  3. AIRVO 2

    2  Pages

  4. OSA Catalog

    25  Pages

  5. ThermoSmart™

    16  Pages

  6. F&P Neowrap

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.