ESA615
5Pages

{{requestButtons}}

Catalog excerpts

ESA615 - 1

Electrical Safety Analyzer Technical Data The ESA615 Electrical Safety Analyzer brings fast and simple automated testing to a portable analyzer for healthcare technology management professionals that perform electrical safety testing on medical equipment both in the field and in facilities. Whether it is simple testing or comprehensive analysis, the ESA615 can do it all. This multifaceted device performs all primary electrical safety tests, including line (mains) voltage, ground wire (protective earth) resistance, insulation resistance, device current, and lead (patient) leakage tests. It also offers ECG simulation and point–to-point voltage, leakage, and resistance tests. Versatile to global electrical safety standards of choice, ESA615 tests to ANSI/AAMI ES1, NFPA99, IEC62353 (VDE751) IEC60601-1 (2nd and 3rd editions) and AS/NZS 3551. The ESA615 is an allin-one solution with a multimeter, safety analyzer and ECG simulator in a single test instrument. Key features • On-board automation with automated test sequences for rapid testing and easy compliance to key global electrical safety standards (ANSI/AAMI ES1 (NFPA-99), IEC62353 (VDE751), IEC60601-1 2nd and 3rd editions, and AS/NZS 3551) • Portable, ergonomic design with an integrated handle and tilt stand • Human-factors-designed user interface for streamlined testing • ECG waveform tests and dual-lead measurements combine the functionality of a simulator, multimeter and safety analyzer in a single test tool • 20 A at 120 V current capability • Five applied parts jacks and easy ECG snap connection; optional expander box for up to 12-lead ECG testing • Easy data entry through barcode scanner, external keyboard or on-board keypad • Wireless communication plus removable memory card for fast and convenient data storage and exchange • Replaceable mains fuses keep your unit in the field and out of the repair shop • Custom language selections include English, French, German, Spanish, Italian and Portuguese • Large, easy-to-read display with adjustable contrast • Optional Ansur automation software completely integrates medical device testing including electrical safety, visual inspection and other performance parameters for total digital data management. Tethers with other Ansur-automated test devices for harmonized workflow and reporting. • Rigorously tested for rugged field applications, with CE and CSA in addition to Fluke quality—ruggedness for long-term reliability • Two-year extended warranty (no-cost, available after first-year calibration at any authorized Fluke Biomedical Service Center) • Global support network delivering prompt service and peace of mind to Fluke Biomedical customers worldwide

Open the catalog to page 1
ESA615 - 2

Automated: Pre-set templates allow you to test to the global standard of your choices at the push of a button. Let the unit guide you through the test steps quickly and accurately in three simple steps: 1. Select the test sequence corresponding to your standard of choice 2. Initiate the automated test sequence. 3. Store the results on-board or wirelessly export to your PC in seconds. It is that easy. The automated test sequences are easily customized to suit individual testing requirements. Portable: The ESA615 is a small, lightweight analyzer with an integrated handle to carry from place...

Open the catalog to page 2
ESA615 - 3

Frequency response/accuracy ± (1 % of reading + (1 μA or 1 LSB, whichever is greater)) ± (2 % of reading + (1 μA or 1 LSB, whichever is greater)) ± (4 % of reading + (1 µA or 1 LSD, whichever is greater)) ± (5 % of reading + (1 μA or 1 LSB, whichever is greater)) Accuracy for Isolation, MAP, Direct AP, Alternative AP, and Alternative Equipment leakage tests all ranges are: • At 120 V ac + (2.5 µA or 1 LSD, whichever is greater) • At 230 V ac additional ± 3.0 % and + (2.5 µA or 1 LSD, whichever is greater) For Alternative equipment, Alternative AP, and Direct AP leakage tests, the leakage...

Open the catalog to page 3
ESA615 - 4

ECG performance waveforms Accuracy ±2% ± 5 % for amplitude of 2 Hz Square wave only, fixed @ 1 mV Lead II configuration Ventricular fibrillation Square wave (50 % duty cycle) Sine wave Triangle wave Pulse (63 ms pulse width) Test standards Available selections Built-in test sequences IEC60601-1 3rd Edition Patient Monitor, Defibrillator, Infusion Pump, Ultrasound Device, Generic Device and System Patient Monitor, Defibrillator, Infusion Pump, Ultrasound Device and Generic Device Patient Monitor, Defibrillator, Infusion Pump, Ultrasound Device and Generic Device Patient Monitor,...

Open the catalog to page 4
ESA615 - 5

Ordering information Models/descriptions 4132046 ESA615 US, 115 V 4162180 ESA615 US, 115 V with Ansur automation software 4132054 ESA615-01 France/Belgium, 230V 4162198 ESA615 France/Belgium, 230 V with Ansur automation software 4132093 ESA615-02 Europe, 230 V 4162211 ESA615 Europe, 230 V with Ansur automation software 4132101 ESA615-03 Israel, 230 V 4162227 ESA615 Israel, 230 V with Ansur automation software 4132112 ESA615-05 Australia/China, 230 V 4162230 ESA615 Australia/China, 230 V with Ansur automation software 4132120 ESA615-06 UK, 230 V 4162248 ESA615 UK, 230 V with Ansur automation...

Open the catalog to page 5

All Fluke Biomedical catalogs and technical brochures

  1. IDA-6

    6 Pages

  2. RAYSAFE 452

    9 Pages

  3. RaySafe DXR+

    2 Pages

  4. BP Pump 2

    82 Pages

  5. ESA612

    4 Pages

  6. RaySafe Solo

    5 Pages

  7. PS320

    4 Pages

  8. DPM1B

    2 Pages

  9. 451B

    3 Pages

  10. 451P

    3 Pages

  11. INCU ?

    4 Pages

  12. 76-606DX

    2 Pages

  13. 190M

    8 Pages

  14. RaySafe Xi

    8 Pages

  15. RaySafe X2

    12 Pages

  16. ProSim 4

    6 Pages

  17. VT305

    4 Pages

  18. ProSim

    8 Pages

Archived catalogs