IDA-1S - 3 Pages

Add to MyMedicalExpo favorites
IDA-1S

Catalog excerpts

One-Channel Infusion Device Analyzer Technical Data Ensure infusion pumps are tested accurately and quickly with real time results using the IDA-1S One-Channel Infusion Device Analyzer, the newest member of the Fluke Biomedical IDA family. The IDA-1S is a portable, battery-operated instrument that allows for speedy verification of infusion device performance. The IDA-1S measures the flow rate and volume delivered, and the pressure generated in occlusions or blockages of the fluid line. The IDA-1S is based on sophisticated measurement technology trusted by biomedical professionals around the world. It is easy to set up and requires little or no training to use. The IDA-1S can be used to test a wide variety of infusion pumps and an auto-start feature simplifies syringe pump testing and other tests that have long startup times. Key features: • Integrated carrying handle and lightweight (2.7 lb) for easy portability • Battery powered with up to 10 hours of continuous operation for on-the-go operation • LCD touch screen for ease of use • Average and instantaneous flow measurement • Occlusion pressure measurements to 45 psi • Maximize accuracy with Autostart mode enabling unit to begin testing only when fluid is detected • Compatible with a wide variety of infusion pumps • Based on technology that is proven and trusted worldwide • On-board memory allows test results storage instantly • Hydrograph graphical software to control unit,display results and print results via PC • Global sales, service and support

 Open the catalog to page 1

Technical specifications Flow rate measurement Technique Flow is calculated by measuring volume over time 1 % of reading ± 1 LSD for flows of 16 ml/h to 200 ml/h for volumes over 20 ml; otherwise 2 % of reading ± 1 LSD for volumes over 10 ml under laboratory conditions Volume measurement Technique Volume is measured directly by the measuring module in minimum sample sizes of 60 μl 1 % of reading ± 1 LSD for flow rates of 16 ml/h to 200 ml/h for volumes over 20 ml. Otherwise 2 % of reading ± 1 LSD for volumes over 10 ml under laboratory conditions Pressure measurement Technique (Occlusion...

 Open the catalog to page 2

Ordering information Models/descriptions 4468525 IDA-1S One-Channel Infusion Device Analyzer Standard accessories 4418071 Hydrograph Software and Users Manual 4497350 20 ml syringe 4480194 3-way plastic Luerlock 4478942 Drain tube (1 m) 4541948 Micro-90 bottle (225 ml) 1740487 USB data transfer cable 2461300 4-plug mains adapter kit for US, UK, EURO, AUS 4329971 Power supply cord without main adapters, univ, wall mount 4481150 NiMH replacement battery Fluke Biomedical. Better products. More choices. One company. Fluke Biomedical Regulatory Commitment As a medical test device manufacturer,...

 Open the catalog to page 3

All Fluke Biomedical catalogs and technical brochures

  1. VT650

    5 Pages

  2. RaySafe Solo

    5 Pages

  3. INCU II

    4 Pages

  4. QA-ES III

    5 Pages

  5. PS320

    4 Pages

  6. DPM1B

    2 Pages

  7. 451B

    3 Pages

  8. 451P

    3 Pages

  9. INCU ?

    4 Pages

  10. 76-606DX

    2 Pages

  11. 190M

    8 Pages

  12. RaySafe Xi

    8 Pages

  13. RaySafe X2

    12 Pages

  14. ESA609

    4 Pages

  15. ProSim 4

    6 Pages

  16. ESA615

    5 Pages

  17. VT305

    4 Pages

  18. ProSim

    8 Pages

Archived catalogs