Defibrillator/External Pacer Analyzer Technical Data The Impulse 6000D Defibrillator Analyzer and Impulse 7000DP Defibrillator/Transcutaneous Pacer Analyzer Test Systems are rugged, portable precision test instruments that ensure proper operation and ultimate performance of critical life-support cardiac-resuscitation equipment. The Impulse 6000D and Impulse 7000DP test capabilities encompass the spectrum of worldwideestablished pulse shapes, showcase breakthrough AED technology compatibility, and outperform in accuracy and standards. Additionally, the Impulse 7000DP incorporates the tests and the extensive range of test loads and measurement algorithms needed to test external transcutaneous pacemakers. In conjunction with an Impulse 7000DP, the Impulse 7010 Defibrillator Selectable Load Accessory provides multiple loads of 25 Ω, 50 Ω, 75 Ω, 100 Ω, 125 Ω, 150 Ω, 175 Ω, and 200 Ω for defibrillator performance testing. A standard USB interface enables computer control and data transfer, and optional Ansur PC-based automation software increases productivity by outfitting users with an easy-to-use method to standardize testing procedures and capture, print and document data. • Impulse 7010 Defibrillator Selectable Load Accessory provides multiple loads of 25 Ω, 50 Ω, 75 Ω, 100 Ω, 125 Ω, 150 Ω, 175 Ω, and 200 Ω to comply with IEC 60601-2-4 standard (optional) Lown, Edmark, trapezoidal, biphasic and pulsed-biphasic defibrillation technology compatibility • • AED technology compatibility • First-class measurement accuracy ± 1 % of reading + 0.1 J • Intuitive user interface and backlight, easy-to-read display • Portable, rugged, easy to carry • Long-lasting, rechargeable battery • Pacer brand selections • Pacer input protected against defibrillator output (7000DP only) • 10 independent ECG outputs that provide 12 lead combinations for standardized clinical signals • Flexible heart-rate settings (1 BPM step) facilitate rate meter accuracy and alarm testing • DSP-based measurements enable future firmware and waveforms upgrade • Unique integrated posts for secure connections • Two-year extended warranty (no-cost extended warranty available after first-year calibration at any Fluke Biomedical authorized service center) • Optional Ansur test automation software to standardize testing procedures, capture waveforms and test results, and print and document test results • Designed, tested, and built to incomparable Fluke quality standards

Open the catalog to page 1
Impulse 6000D/7000DP - 2013 - 2

General specifications Operating temperature Storage temperature USB device port for computer control Modes of operation Manual and remote Internal rechargeable NiMH battery pack for nine hours (typical) operation after full charge or the battery charger can operate the analyzer and charge the battery simultaneously Battery charger 100 V to 240 V input, 15 V/1.5 A output. For best performance, the battery charger should be connected to a properly grounded ac receptacle Dimensions (WxDxH) Safety standards CE: IEC/EN61010-1 2nd Edition; Pollution degree 2 CAN/CSA-C22.2 No 61010-1; UL61010-1...

Open the catalog to page 2
Impulse 6000D/7000DP - 2013 - 3

ECG waves •  Normal sinus rhythm (NSR): 10 BPM to 180 BPM in 1 BPM steps •  Atrial fibrillation: Coarse and fine Monomorphic ventricular •  tachycardia: 120 BPM to 240 BPM in 5 BPM steps •  Asystole: Flat line Automated defibrillator test ECG waves Normal sinus: 10 BPM to 300 BPM in 1 BPM steps Ventricular fibrillation: Coarse and fine Monomorphic ventricular tachycardia: 120 BPM to 300 BPM in 5 BPM steps Polymorphic ventricular tachycardia: 5 types Asystole: Flat line Lead configuration: 12-lead simulation; RA, LL, LA, RL, V1-6 with independent outputs Lead to lead impedance: 1000 Ω...

Open the catalog to page 3
Impulse 6000D/7000DP - 2013 - 4

Arrhythmia selections Pacer interactive (7000DP only) • Demand: 30 BPM to 360 BPM in 1 BPM steps • Asynchronous • Non-capture • Non-function • Threshold (interactive pacing simulation only): 10 mA to 250 mA in 10 mA steps Supraventricular • Artrial fibrillation course • Atrial fibrillation fine • Atrial flutter • Sinus arrhythmia • Missed beat • Atrial tachycardia • Paroxysmal atrial tachycardia (PAT) • Nodal rhythm • Supraventricular tachycardia Premature • Atrial PAC • Nodal PNC • PVC1 left ventricle • PVC1 LV early • PVC1 LV R on T • PVC2 right ventricle • PVC2 RV early • PVC2 RV R on T...

Open the catalog to page 4
Impulse 6000D/7000DP - 2013 - 5

Transcutaneous pacemaker analyzer technical specifications (7000DP only) Test load Selections Defibrillator input Fixed load: 50 Ω Accuracy: ± 1 %, non-inductive (<2 µH) Power rating: 10 defib pulses of 360 J every 5 minutes Pacemaker input Variable load: 50 Ω to 1500 Ω in 50 Ω steps Accuracy: ± 2 %, non-inductive (< 2 µH) Power rating: 5 W (average), 40 W (peak) @ 1000 Ω Rate: 10 BPM to 300 BPM in 1 BPM steps Amplitude: 1 mV Underdrive rate: 10 BPM minimum Overdrive rate: 300 BPM maximum Sensitivity test Automatic interactive threshold detection Accuracy: ± 5 % of setting Amplitude: 0.05...

Open the catalog to page 5
Impulse 6000D/7000DP - 2013 - 6

Impulse 7010 Defibrillator Selectable Load Accessory General specifications Maximum voltage 5000 V Maximum continuous power 12 W, equivalent to 10 defib pulses of 360 J every 5 minutes Inductance < 2 µH, @25 Ω < 3 µH, @50 Ω < 4 µH, @75 Ω and 100 Ω < 5 µH, @125 Ω < 6 µH, @150 Ω < 7 µH, @175 Ω < 8 µH, @200 Ω Temperature Operating: 10 °C to 40 °C (50 °F to 104 °F) Storage: -20 °C to 60 °C (-4 °F to 140 °F) Calibration interval One-year Electrical specifications (for Load Accessory and Analyzer together) Safety class Complies with EN61010-1 2nd Edition, Class II product Safety and EMC marks...

Open the catalog to page 6
Impulse 6000D/7000DP - 2013 - 7

Ordering information Models 2811928 Impulse 6000D Defibrillator Analyzer 120 V (US) 3077031 Impulse 6000D Defibrillator Analyzer (Schuko) 3077046 Impulse 6000D Defibrillator Analyzer (UK) 3077054 Impulse 6000D Defibrillator Analyzer (Japan) 3085270 Impulse 6000D Defibrillator Analyzer (Australia) 3085281 Impulse 6000D Defibrillator Analyzer (India) 2811919  Impulse 7000DP Defibrillator/ Transcutaneous Pacemaker Analyzer 120 V (US) 3077005 Impulse 7000DP Defibrillator/ Transcutaneous Pacemaker Analyzer (Schuko) 3077010 Impulse 7000DP Defibrillator/ Transcutaneous Pacemaker Analyzer (UK)...

Open the catalog to page 7

All Fluke Biomedical catalogs and technical brochures

  1. IDA-6

    6 Pages

  2. RAYSAFE 452

    9 Pages

  3. RaySafe DXR+

    2 Pages

  4. BP Pump 2

    82 Pages

  5. ESA612

    4 Pages

  6. RaySafe Solo

    5 Pages

  7. PS320

    4 Pages

  8. DPM1B

    2 Pages

  9. 451B

    3 Pages

  10. 451P

    3 Pages

  11. INCU ?

    4 Pages

  12. 76-606DX

    2 Pages

  13. 190M

    8 Pages

  14. RaySafe Xi

    8 Pages

  15. RaySafe X2

    12 Pages

  16. ProSim 4

    6 Pages

  17. ESA615

    5 Pages

  18. VT305

    4 Pages

  19. ProSim

    8 Pages