Fetal Simulator Technical Data The PS320 simulates fetal and maternal ECG as well as uterine activity to test and troubleshoot fetal electronic monitors and to train clinical staff. The unit is battery operated and small enough to fit in a pocket so mobile technicians and clinical instructors can take it anywhere. The PS320 simulates several fetal parameters, including twins, as well as a wide range of clinical scenarios for training labor-and-delivery staff in how to recognize normal and abnormal responses. An optional mechanical heart creates fetal heart sounds for testing fetal monitor ultrasound cables and transducers. PS320 offers an easy user interface, providing a 2 x 16-character LCD display with adjustable contrast. The unit operates on a 9 V battery with low-battery monitoring or functions with the supplied battery eliminator. Key features • Mechanical heart for ultrasound simulation • TOCO simulation (External or IUP) • Ultrasound simulation (including twins) • Maternal ECG simulation • Fetal ECG (tracks ultrasound #1) • Internal (DECG) and external fetal ECG • Uterine-activity selections • Fetal beat-to-beat variability • Periodic and non-periodic fetal ECG changes • Arrhythmia selections • Compact, lightweight, and pocket-size plastic housing • Battery operated with status indications • Special kits available with all required accessories and cables to test fetal monitors for specified manufacturers

Open the catalog to page 1
PS320 - 2

Technical specifications Fetal ECG Static rates 30 BPM, 60 BPM, 90 BPM, 120 BPM, 150 BPM, 180 BPM, 210 BPM, and 240 BPM ECG sensitivity 50 µV, 100 µV, 200 µV, 0.5 mV, 1 mV, and 2 mV US-1 tracks primary fetal ECG rates. US-2 tracks secondary fetal activity for either independent “normal” or “twins” simulation, US-2 rate is fixed at 140 BPM Fetal patterns Note: US-1 and fetal ECG track these selections. US-2 is in normal pattern, except during TREND #1 selection. Trend #1: Twin fetal patterns Normal: Normal pattern Tachycardia: Tachycardia pattern Bradycardia: Bradycardia pattern Arrhythmias:...

Open the catalog to page 2
PS320 - 3

The PS320 turns on in the following state: • Fetal ECG static rate @ 150 BPM • US-1 tracks @ 150 BPM • US-2 normal pattern • Pressure sensitivity @ 5 µV/mmHg • Pressure/Toco set to zero • Maternal ECG rate @ 80 BPM • ECG sensitivity @ 1 mV • Toco wave is normal duration @ 0 to 50 divisions (i.e. 0 mmHg to 50 mmHg) Controls Display 2-line x 16-character LCD with keypad RS-232 Bidirectional interface, 9600 Baud 9 V battery/battery eliminator; low battery indication set at 6 V Plastic case Dimensions (LxWxH) Temperature Operating 15 °C to 35 °C (59 °F to 95 °F), Storage 0 °C to 50 °C (32 °F to...

Open the catalog to page 3
PS320 - 4

Ordering information continued Standard accessories 2631717 Electronic-Version Users Manual (CD-ROM) 2631693 Printed-Version Users Manual 2647372 Battery Eliminator Universal 90 V to 240 V N/A 9 V Battery Optional accessories 2647372 Battery Eliminator 100 V ac to 240 V ac 2462177 Carrying Case, double pocket 2462478 HP/Ag/Philips 50 Series -Ultrasound Cable 2462491 HP/Ag/Philips 50 Series TOCO -External Cable 2462528 HP/Ag/Philips 50 & 8040 Series -IUP Cable 2462469 Corometrics -IUP Cable 2462484 Corometrics -Ultrasound Cable 4021537 Corometrics 120 Ultrasound Cable 2462519 Corometrics...

Open the catalog to page 4

All Fluke Biomedical catalogs and technical brochures

  1. RAYSAFE 452

    9 Pages

  2. RaySafe DXR+

    2 Pages

  3. BP Pump 2

    82 Pages

  4. ESA612

    4 Pages

  5. RaySafe Solo

    5 Pages

  6. DPM1B

    2 Pages

  7. 451B

    3 Pages

  8. 451P

    3 Pages

  9. INCU ?

    4 Pages

  10. 76-606DX

    2 Pages

  11. 190M

    8 Pages

  12. RaySafe Xi

    8 Pages

  13. RaySafe X2

    12 Pages

  14. ProSim 4

    6 Pages

  15. ESA615

    5 Pages

  16. VT305

    4 Pages

  17. ProSim

    8 Pages

Archived catalogs