190M
1 / 8Pages

Catalog excerpts

190M - 1

Medical ScopeMeter® Portable Oscilloscope Technical Data The 190M: a new generation of medical oscilloscope The 190M Medical ScopeMeter portable oscilloscope is a high-performance test tool built upon the legacy of Fluke and Fluke Biomedical oscilloscopes in partnership with real customers like you. The 190M is available with choice of two or four channels and offers an unprecedented level of performance, ruggedness, and portability. With the combined power of a high-performance oscilloscope, multimeter and paperless recorder in an easy-to-use test tool, the 190M is the one test tool you can rely on to tackle just about any troubleshooting task in the field. To minimize downtime and repair costs, you need to get to the root cause of problems as quickly as possible. The 190M offers a number of unique features to help you quickly set up the scope and diagnose difficult problems like intermittent events, signal fluctuations or drift. Extend your arsenal of troubleshooting capabilities with the new Fluke Biomedical 190M Medical ScopeMeter portable oscilloscope, designed to meet the demands of field service professionals. • ScopeRecord roll mode gives 30,000 points per input channel and capture waveform sample data for up to 40 hours. • Two or four electrically-isolated inputs • TrendPlot, trend measurement readings for up to 22 days Advanced automatic measurements, power (Vpwm, VA, W, PF) and time (mAs, V/s, w/s). • Fast sampling rate, up to 2.5 GS/s on two channels simultaneously with up to 400 ps resolution. • Deep memory: 10,000 samples per channel waveform capture so you can zoom in on the details (scope mode). • Dedicated 5000 count digital multimeter in two-channel model. • Quad meter measurements via scope BNC inputs in four channel model. • Connect-and-View™ triggering for intelligent, automatic triggering on fast, slow and even complex signals. • Frequency spectrum using FFT-analysis. • Large, bright color displaySmart averaging. • Two USB ports make it easy to transfer data to a PC and store unlimited waveforms, screen captures and instrument setups on USB memory devices. • New high-perfomance Li-ion battery technology delivers the longest battery life on the market. • Charge spare battery using optional external battery charger. • Easy-access battery door for quick swaps in the field. • Security slot locks down oscilloscope with Kensington lock while unattended. • Environmentally tested to meet IP-51 and wit

Open the catalog to page 1
190M - 2

Rise time Number of scope inputs 2 input channels plus external trigger Channel architecture All inputs fully insulated from each other and from ground Inputs may be activated in any combination Input coupling AC or DC, with ground level indicator Input sensitivity 2 mV/div to 100 V/div, plus variable attenuation Bandwidth limiter On each input channel, switched separately Input voltage CAT III 1000 V/CAT IV 600 V rated, see general specifications for further details Vertical resolution ± (2.1 % of reading + 0.04 x range/div) @ 5 mV/div to 100 V/div Input impedance Oscilloscope modes...

Open the catalog to page 2
190M - 3

Pulse width triggering (on channel A) Pulse width qualified by time allows for triggering < t, > t, = t, ≠ t, where t is selectable in minimum steps of 0.01 div or 50 ns Time delay 1 Full screen of pre-trigger view or up to 100 screens (= 1,200 divisions) of post- trigger delay Dual slope triggering Triggers on both rising and falling edges alike N-cycle triggering Triggers on Nth occurrence of a trigger event; N to be set in the range 2 to 99 Automatic capture of 100 screens When in oscilloscope mode, the instrument always memorizes the last 100 screens—no specific user setup required....

Open the catalog to page 3
190M - 4

Single vertical line Min/max and average voltage at cursor position; frequency and rms-value of individual frequency component in the FFT resultant Ranges from full record overview to zoom-in up to sample level at any record length Meter Modes Meter inputs Via 4 mm banana inputs, fully isolated from scope inputs and scope ground Maximum resolution Input impedance Advanced meter functions Auto/manual ranging, relative measurements (Zero reference), TrendPlot™ recording The specified accuracy is valid over the temperature range 18 °C to 28 °C Add 10 % of specified accuracy for each degree C...

Open the catalog to page 4
190M - 5

Memory depth 30,000 data points, each holding min/max pair of information Min/max values Min/max values are created at samples that are measured at high sample rate, ensuring capture and display of glitches Recording modes Single sweep, continuous roll, Start-on-trigger (through external), Stopon-trigger (through external) ScopeRecord mode can be stopped by an individual trigger event or by an interruption of a repetitive trigger signal through any input channel (through external on 190M-2 model) Horizontal scale Time from start, time of day Ranges from full record overview to zoom in up to...

Open the catalog to page 5
190M - 6

General specifications Input voltage range Rated maximum floating voltage CAT III 1000 V/CAT IV 600 V (Maximum voltage between any contact and earth-ground voltage level) Maximum probe voltage CAT III 1000 V/CAT IV 600 V (Maximum voltage between any contact and earth-ground voltage level) Maximum BNC input voltage CAT IV 300 V (Maximum voltage on BNC input directly) Maximum voltage on meter input CAT III 1000 V/CAT IV 600 V (Safety designed banana input connectors) Memory save and recall Memory locations (internal) 30 waveform memories plus 10 recording memories plus 9 screen copy memories...

Open the catalog to page 6
190M - 7

Battery type (included) and capacity [+opt. battery] Battery charge indicator Battery has built-in status indicator for use with external charger, next to battery status indicator on instrument screen Battery operating time (with backlight low) Up to 3.5 using BP290 (included), up to 7 hours using BP291 (optional) Up to seven hours using BP291 (included) Battery charging time up to 7 hours using BP291 (optional) Battery power saving functions Auto power-down with adjustable power-down time; auto display off with adjustable power-down time; on-screen battery power indicator Safety Compliance...

Open the catalog to page 7

All Fluke Health Solutions catalogs and technical brochures

  1. ESA612

    6 Pages

  2. PS320

    5 Pages

  3. ESA615

    7 Pages

  4. IDA-6 Flyer

    1 Pages

  5. 451B

    3 Pages

  6. 451P

    3 Pages

  7. ProSim 4

    6 Pages

Archived catalogs

  1. RAYSAFE 452

    9 Pages

  2. RaySafe DXR+

    2 Pages

  3. BP Pump 2

    82 Pages

  4. RaySafe Solo

    5 Pages

  5. DPM1B

    2 Pages

  6. INCU ?

    4 Pages

  7. 76-606DX

    2 Pages

  8. RaySafe Xi

    8 Pages

  9. RaySafe X2

    12 Pages

  10. VT305

    4 Pages

  11. ProSim