Technical data Impulse 6000D/7000DP Defibrillator/External Pacer Analyzer The Impulse 6000D Defibrillator Analyzer and Impulse 7000DP Defibrillator/Transcutaneous Pacer Analyzer Test Systems are rugged, portable precision test instruments that ensure proper operation and ultimate performance of critical lifesupport cardiac-resuscitation equipment. The Impulse 6000D and Impulse 7000DP testcapabilities encompass the spectrum of worldwideestablished pulse shapes, showcase breakthrough AED technology compatibility, and outperform in accuracy and standards. Additionally, the Impulse 7000DP incorporates the tests and the extensive range of test loads and measurement algorithms needed to test external transcutaneous pacemakers. In conjunction with an Impulse 7000DP, the Impulse 7010 Defibrillator Selectable Load Accessory provides multiple loads of 25 Ω, 50 Ω, 75 Ω, 100 Ω, 125 Ω, 150 Ω, 175 Ω, and 200 Ω for defibrillator performance testing. A standard USB interface enables computer control and data transfer, and optional Ansur PC-based automation software increases productivity by outfitting users with an easy-to-use method to standardize testing procedures and capture, print and document data. Key features •• Impulse 7010 Defibrillator Selectable Load Accessory provides multiple loads of 25 Ω, 50 Ω, 75 Ω, 100 Ω, 125 Ω, 150 Ω, 175 Ω, and 200 Ω to comply with IEC 60601-2-4 standard (optional) •• Lown, Edmark, trapezoidal, biphasic and pulsed-biphasic defibrillation technology compatibility •• AED technology compatibility •• First-class measurement accuracy ± 1 % of reading + 0.1 J •• Intuitive user interface and backlight, easy-to-read display •• Portable, rugged, easy to carry •• Long-lasting, rechargeable battery •• Pacer brand selections •• Pacer input protected against defibrillator output (7000DP only) •• 10 independent ECG outputs that provide 12 lead combinations for standardized clinical signals •• Flexible heart-rate settings (1 BPM step) facilitate rate meter accuracy and alarm testing •• DSP-based measurements enable future firmware and waveforms upgrade •• Unique integrated posts for secure connections •• Two-year extended warranty (no-cost extended warranty available after first-year calibration at any Fluke Biomedical authorized service center) •• Optional Ansur test automation software to standardize testing procedures, capture waveforms and test results, and print and document test resu
Open the catalog to page 1
Specifications General specifications Operating temperature Storage temperature USB device port for computer control Internal rechargeable NiMH battery pack for nine hours (typical) operation after full charge or the battery charger can operate the analyzer and charge the battery simultaneously Battery charger 100 V to 240 V input, 15 V/1.5 A output. For best performance, the battery charger should be connected to a properly grounded ac receptacle Dimensions (WxDxH) Safety standards CE: IEC/EN61010-1 2nd Edition; Pollution degree 2 CAN/CSA-C22.2 No 61010-1; UL61010-1 C-Tick: Australian EMC...
Open the catalog to page 2
Defibrillator analyzer technical specifications Energy output measurement Compatible defibrillator waveshapes Lown, Edmark, trapezoidal, dc biphasic, and ac pulsed biphasic Autoranged measurement 0.1 J to 360 J: ± (1 % of reading + 0.1 J) 360 J to 600 J: ± (1 % of reading + 0.1 J), typical Note: AC pulsed biphasic waveform has not been approved in the United States. Note: For pulsed biphasic defibrillator, specified accuracy is ± (1.5 % of reading + 0.3 J) on both ranges. Load resistance Resistance Pulse trigger level Pulse width Range Voltage Range Current Range Tilt (biphasic and pulsed...
Open the catalog to page 3
Charge time measurement Range Synchronization test (elective cardioversion) Delay time measurement • Timing window: ECG R-wave peak to the defib pulse peak • Range: -120 ms to 380 ms; measures timing from 120 ms prior to the R-wave peak to up to 380 ms following the R-wave peak • Resolution: 1 ms • Accuracy: ± 1 ms • Normal sinus rhythm (NSR): 10 BPM to 180 BPM in 1 BPM steps • Atrial fibrillation: Coarse and fine • Monomorphic ventricular tachycardia: 120 BPM to 240 BPM in 5 BPM steps • Asystole: Flat line Automated defibrillator test ECG waves Normal sinus Ventricular fibrillation...
Open the catalog to page 4
Normal sinus waves: Lead # Lead I reference Performance waves and R wave detection Lead # Normal sinus waves: Lead # Impulse 6000D/7000DP Defibrillator/External Pacer Analyzer
Open the catalog to page 5
ECG normal sinus Rates ECG high level output (BNC jack) Amplitude Range 0.5 V per mV of reference lead setting Output impedance ECG on defibrillator input load Same as the Lead II amplitude but limited to ± 4 mV ECG performance waves Square wave Triangular wave Sine waves R-wave detection Waveform Impulse 6000D/7000DP Defibrillator/External Pacer Analyze
Open the catalog to page 6
Noise immunity Wave Line frequency Amplitude Range Transvenous pacer pulse simulation Widths Range Amplitudes Range Amplitude of transvenous pacer pulse simulation signals relative to amplitude setting (in percent) Lead II reference Lead # Lead I reference Lead #
Open the catalog to page 7
Arrhythmia selections Pacer interactive (7000DP only) Demand: 30 BPM to 360 BPM in 1 BPM steps Asynchronous Non-capture Non-function Threshold (interactive pacing simulation only): 10 mA to 250 mA in 10 mA steps Artrial fibrillation course Atrial fibrillation fine Atrial flutter Sinus arrhythmia Missed beat Atrial tachycardia Paroxysmal atrial tachycardia (PAT) Nodal rhythm Supraventricular tachycardia Premature Atrial PAC Nodal PNC PVC1 left ventricle PVC1 LV early PVC1 LV R on T PVC2 right ventricle PVC2 RV early PVC2 RV R on T Multifocal PVCs PVCs 6/min PVCs 12/min PVCs 24/min Freq...
Open the catalog to page 8
1° Block 2° Block Type I 2° Block Type II 3° Block Right bundle branch block RBBB Left bundle branch block LBBB Transvenous Paced with selectable pacer spike amplitudes and widths Atrial 80 BPM Async 75 BPM Demand with frequent sinus beats Demand with occasional sinus beats AV sequential Non-capture Non-function Selections for all waves in group Atrial pacer pulse Width Ventricular pacer pulse Width R-wave detection Rate Transcutaneous pacemaker analyzer technical specifications (7000DP only) Test load Selections Defibrillator input Fixed load Power rating 10 defib pulses of 360 J every 5...
Open the catalog to page 9All Fluke Health Solutions catalogs and technical brochures
-
ESA612
6 Pages
-
PS320
5 Pages
-
190M
8 Pages
-
ESA615
7 Pages
-
Gas flow analyzer
2 Pages
-
ProSim 3 and 2
7 Pages
-
Product catalog
40 Pages
-
IDA-6 Flyer
1 Pages
-
ProSim SPOT Light
4 Pages
-
VT900A Datasheet
6 Pages
-
VT900A + VAPOR
7 Pages
-
ESA614 Datasheet
6 Pages
-
Gas Flow Family Brochure
2 Pages
-
RaySafe Solo Specifications
5 Pages
-
451B
3 Pages
-
451P
3 Pages
-
ProSim 4
6 Pages
Archived catalogs
-
QA-ES Series II
4 Pages
-
RAYSAFE 452
9 Pages
-
451P Ion Chamber Survey Meter
30 Pages
-
RaySafe Dose Viewer
10 Pages
-
RaySafe DXR+
2 Pages
-
BP Pump 2
82 Pages
-
Electrical Safety Analyzers
8 Pages
-
RaySafe Solo
5 Pages
-
DPM1B
2 Pages
-
SigmaPace™ 1000
8 Pages
-
965 Stack Monitor
3 Pages
-
INCU ?
4 Pages
-
76-606DX
2 Pages
-
07-CRXW and 07-QRX
4 Pages
-
RaySafe Xi
8 Pages
-
RaySafe X2
12 Pages
-
Impulse 6000D/7000DP - 2013
8 Pages
-
VT305
4 Pages
-
ProSim
8 Pages