INCU ?

INCU ?

INCU ?

Product catalog summary
Key Features:
The INCU™ Incubator Analyzer is designed for simultaneous measurement of humidity, airflow, sound, and four independent temperatures. It supports 24-hour continuous testing on battery and 35-hour on main power. The device is compatible with both closed, forced-convection incubators and open infant warmers, and can operate as a stand-alone unit or with a PC for automated testing. The Windows® compatible INCU software facilitates data collection, analysis, and documentation, providing numerical and full-color graphical reports.
Importance:
Faulty incubator controls can jeopardize the health of critical-care babies, increase hospital stays, and healthcare costs, making thorough testing essential. The INCU™ Analyzer is the only device that measures seven key parameters simultaneously, ensuring accurate results and saving time.
Technical Specifications:
  • Power Supply: 100 V ac to 240 V ac, 47 Hz to 63 Hz, < 60 VA power consumption.
  • Battery: Rechargeable sealed lead-acid, 6 V, 7 Ah, with a low-battery alert.
  • Sound Level: 30 dbA to 80 dbA range, 0.1 dbA resolution, ±5 dbA accuracy.
  • Humidity: 0% to 100% RH range, 0.1% RH resolution, ±5% RH accuracy.
  • Temperature: 5 °C to 70 °C range, 0.1 °C resolution, ±0.5 °C accuracy.
  • Airflow: 0.1 m/s to 0.7 m/s range, 0.01 m/s resolution, ±0.1 m/s accuracy.
Software Requirements:
Compatible with Windows 95B, 98, NT® 4.0, or 2000, requiring a Pentium 133 MHz processor, 16 MB free hard disk space, and a CD ROM drive.
About Fluke Biomedical:
Fluke Biomedical is a leading manufacturer of biomedical test and simulation products, providing solutions for regulatory compliance and quality assurance in medical imaging and oncology.
Ordering Information:
Model 2250148 includes standard accessories such as a user manual, carrying case, airflow sensor, universal AC battery charger, and more.
See more

Catalog excerpts

INCU ?-1

Incubator Analyzer Technical Data Faulty incubator controls jeopardize the health of critical-care babies, lengthen hospital stays, and increase healthcare costs, making thorough incubator testing essential. The INCU™ Incubator Analyzer is the only device on the market today that accurately measures seven key parameters at once in a single, integrated unit, saving time and ensuring more accurate results. Designed around AAMI and IEC standards that specify incubator and radiant warmer sound levels and thermal characteristics, the INCU simultaneously measures airflow, relative humidity, sound, and four independent temperatures. Adjustable measurement intervals allow technicians to configure the unit to meet their facilities’ testing requirements. Technicians are free to do other work while the device collects and records data. The INCU is compatible with both closed, forced-convection incubators, and open infant warmers, including transportable, air-controlled, and baby-controlled units. Lightweight and portable, the INCU fits inside the incubator and operates as a stand-alone device or with a personal computer for automated testing. In stand-alone mode, the INCU displays measured parameters repeatedly in cycle fashion. INCU software lets technicians upload setup parameters and download the test results to a PC file, or print the data in reports with fullcolor charts and graphs. Key features • Simultaneous measurement of humidity, airflow, sound, and 4 independent temperatures • 24-hour continuous testing (battery); 35-hour continuous testing (main power) • Battery operated • Adjustable measurement intervals • Compatible with closed, forced-convection incubators and open infant warmers • Stand-alone measurement or automated testing with PC • Windows® compatible INCU software for easy data collection, analysis, and documentation • Numerical and full-color graphical reports

 Open the catalog to page 1
INCU ?-2

Technical specifications Temperature measurement Power supply Input voltage range 100 V ac to 240 V ac Accuracy + 0.5 °C (+ 0.9 °F) + 1 LSB of range from 25 °C to 40 °C (77 °F to 104 °F) Input frequency range 47 Hz to 63 Hz Conduction One sensor in contact with mattress Convection Inside the incubator Three sensors (T1, T3, T4) Inside or outside the incubator One sensor (T2) Fuse rating 2 A slow blow Battery Rechargeable sealed lead-acid type NP7-6 YUASA, 6 V, 7 Ah; operates for 24 hours continuously; low-battery alert Radiation One sensor for radiation head that can be used also for external...

 Open the catalog to page 2
INCU ?-3

Temperature Storage -20 °C to 50 °C (-4 °F to 122 °F) Operating 10 °C to 40 °C (50 °F to 104 °F) Dimensions (LxWxH) INCU software PC system requirements • Pentium 133 MHz microprocessor • 16 MB free hard disk space • SVGA color monitor • Serial port or USB port (may require serial to USB adapter, ICON USB 2.0 to RS-232 cable recommended (sourced separately) • Mouse compatible with Microsoft® operating system • 1 CD ROM disk drive • Microsoft Windows 95B, Windows 98, Windows NT® 4.0, or Windows 2000 • Laser or inkjet printer INCU™ Incubator Analyzer

 Open the catalog to page 3
INCU ?-4

Ordering information Model 2250148 INCU™ Incubator Analyzer Standard accessories 2206965 User Manual 2248900 Soft-Sided Carrying Case 2239025 Airflow Sensor 2391761 Universal AC Battery Charger with worldwide mains adapter set 2391866 Serial Cable DB9F to DB9F 2391789 Serial Adapter DB9M to DB25F 2213928 Outside-Temperature Probe Holder 2239002 Adapter for Radiant Infant Warmer Assembly 2213919 INCU PC Software (1 CD) About Fluke Biomedical Fluke Biomedical. Fluke Biomedical is the world’s leading manufacturer of quality biomedical test and simulation products. In addition, Fluke Biomedical provides...

 Open the catalog to page 4

Archived catalogs

  1. RAYSAFE 452

    9  Pages

  2. RaySafe X2

    12  Pages

  3. RaySafe DXR+

    2  Pages

  4. BP Pump 2

    82  Pages

  5. RaySafe Solo

    5  Pages

  6. DPM1B

    2  Pages

  7. 76-606DX

    2  Pages

  8. RaySafe Xi

    8  Pages

  9. VT305

    4  Pages

  10. ProSim

    8  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.