SigmaPace™ 1000

SigmaPace™ 1000

SigmaPace™ 1000

Product catalog summary
Overview of SigmaPace™ 1000: The SigmaPace™ 1000 is an advanced external pacemaker analyzer designed for testing both transcutaneous and transvenous pacemakers. It enhances efficiency and accuracy in pacemaker analysis through a variety of tests and features.
Key Features:
  • Performs comprehensive tests including pulse-output, amplitude sensitivity, and refractory tests for both transcutaneous and transvenous pacemakers.
  • Supports demand and asynchronous mode tests, DC load current tests, and output-leakage tests.
  • Displays pacer output on three screens with AV readout for both pacer channels.
  • Offers long-term trend testing for up to 11 days to identify intermittent errors.
  • Includes interactive ECG pacer simulation for training and equipment evaluation.
Specifications:
  • Transcutaneous Pacer Tests: Measures current, rate, width, and energy with high accuracy across test loads from 50 Ω to 1550 Ω.
  • Transvenous Pacer Tests: Provides detailed output pulse measurement with various ranges for current, rate, width, and voltage, including AV interval measurement.
  • Amplitude Sensitivity Tests: Offers adjustable amplitude and width settings for R-, S-, and T-waves.
  • Noise Immunity Tests: Evaluates line frequency noise rejection with 50 Hz and 60 Hz sine waveforms.
  • Refractory Period Tests: Conducts paced and sensed refractory period tests for atrial and ventricular channels.
  • DC Leakage Current: Measures leakage current with high resolution and accuracy.
  • Long-term Testing: Configurable for both pacer types with a maximum pulse count of 999,999.
Accessories: Includes an operator's manual, carrying case, test leads, load test cable, and serial PC interface.
Overview of Universal-Input Battery Charger: This document outlines the technical specifications and ordering information for Fluke Biomedical's Universal-Input Battery Charger, emphasizing quality and regulatory compliance.
Specifications:
  • Power Supply: Universal-Input Battery Charger with USA 120 V AC power cord set.
  • Operating Temperature: 15 °C to 35 °C (59 °F to 95 °F).
  • Storage Temperature: 0 °C to 50 °C (32 °F to 122 °F).
  • Humidity Range: Less than 90% non-condensing.
  • Modes of Operation: Manual and remote via RS-232 serial port.
  • User Interface: 21-character x 8-line LCD readout with adjustable brightness and viewing angle.
  • Power: External battery charger source/power supply 100 to 240 V AC, 50/60 Hz with auto power-off during battery operation.
  • Battery Life: 20 hours.
  • Dimensions: 10.1 cm x 20.3 cm x 5 cm (4 in x 8 in x 2 in).
  • Weight: 0.9 kg (2 lb).
  • Safety Standards: EMC: EN61326-1.1997; UL STD 3101-1; CAN/USA STD C22.2 No.1010 ETL Listed; FDA 510(k) clearance.
Optional Accessories:
  • Electrode Adapters with brand/model-specific connectors and 4 mm banana plugs.
  • Various detachable cord sets for different regions including Japan, Schuko/Euro, UKI, USA, and Australia.
  • Interface Cable for RS-232 connections.
Company Information: Fluke Biomedical is a leading manufacturer of biomedical test and simulation products, providing solutions for medical imaging and oncology quality assurance. The company is ISO 9001 certified, with products that are NIST traceable, calibrated, and certified by UL, CSA, and ETL where required.
Contact Information: Fluke Biomedical, 6045 Cochran Road, Cleveland, OH 44139-3303, USA. Phone: (800) 850-4608, Fax: (440) 349-2307, Email: [email protected], Website: www.flukebiomedical.com
See more

Catalog excerpts

SigmaPace™ 1000-1

External Pacemaker Analyzer Technical Data Fluke Biomedical’s premier SigmaPace™ 1000 analyzes both transvenous and transcutaneous external pacemakers and comes loaded with features to save time and money. This powerful handheld tool conducts the full suite of tests specified by major pacemaker manufacturers in less than half the time it would take using originally prescribed testing methods. Output data is displayed on three selectable screens for easy viewing, including an AV delay time readout providing a performance snapshot for both pacer channels. With capability for long-term trend testing, the SigmaPace 1000 can interrogate a pacer for up to 11 days, capturing data pulse by pulse to detect intermittent and hard-to-find problems. For maximum efficiency, the SigmaPace 1000 doubles as a training tool. Interactive ECG simulation lets users test patient monitoring equipment as well as teach nurses how to operate the pacemaker. • Transcutaneous and transvenous external pacemaker tests • Pulse-output tests (rate, current, volts, energy, pulse width, and AV interval) • Amplitude sensitivity and refractory tests • Demand and asynchronous-mode tests • DC load current test • Output-leakage tests • Line-frequency noise-rejection tests • Wide range of test loads, from 50 Ω to 1500 Ω, specified by manufacturer for transcutaneous pacers • Full range of IEC specified test loads for transvenous pacers 200 Ω, 500 Ω, and 1000 Ω • Pacer output displayed on three different screens • AV readout showing both pacer channels on one screen • Long-term trend test to detect intermittent errors and hard-to-find problems • Interactive ECG pacer simulation with 5-lead output for patient monitor evaluation and pacer operation training • 8-line x 21-character display

 Open the catalog to page 1
SigmaPace™ 1000-2

Specifications Transcutaneous pacer tests Output pulse measurement Current Accuracy Ranges Accuracy Ranges Accuracy Ranges Accuracy Demand and asynchronous mode tests Waveform (physiological simulation) Amplitude Modes of operation Active outputs Pacemaker compatibility 4 mA to 9.99 mA; 10 mA to 99.9 mA; 100 mA to 250 mA ± 2 % of reading or ± 50 μA (whichever is greater) 5 PPM to 99.9 PPM; 100 PPM to 300 PPM ± 0.5 % of reading or ± 0.3 PPM (whichever is greater) 1 mS to 9.99 mS; 10 mS to 99.9 mS ± 0.5 % of reading or ± 14 μS (whichever is greater) 1 μJ to 999 μJ; 1 mJ to 999 mJ; 1.00 J to 1.99...

 Open the catalog to page 2
SigmaPace™ 1000-3

Active outputs 5-lead ECG; ventricular test load Paced refractory period test (PRP) Range Accuracy Physiological simulation Outputs Pacemaker compatibility 20 mS to 500 mS 5 % of reading or 1 mS (whichever is greater) Selection Single pulse, R-wave, SSQ Pulse width 40 mS 5-lead ECG; ventricular test load Pulse rates 30 BPM to 200 BPM Intended type VVI (pace and sense) Sensed refractory period test (SRP) Range Accuracy Physiological simulation Active outputs Pacemaker compatibility 15 mS to 500 mS ± 5 % of reading or ± 1 mS (whichever is greater) Selection Double pulse, R-wave, SSQ Pulse width...

 Open the catalog to page 3
SigmaPace™ 1000-4

Demand/async mode tests Channels Waveform Atrial output Vent output Interactive simulated rates Output Pulse rate Intended pacemaker types Single and dual Sine square (SSQ) Simulated P-wave Width Amplitude Simulated R-wave Width Amplitude Interval Default settings 40 mS 2.5 mV peak AV 90 mS (fixed) Underdrive = NSR @ 85 % of pulse interval/rate; overdrive = NSR @ 115 % of pulse interval/rate Manual NSR simulations adjustable in 1-BPM increments Limits Underdrive (min) = 10 BPM; overdrive (max) = 300 BPM Ventricular channel test load; atrial channel test load 30 PPM to 200 PPM Demand VVI (V-channel...

 Open the catalog to page 4
SigmaPace™ 1000-5

Ventricular channel (physiological simulation) Waveform selections Available test load(s) Amplitude Intended pacemaker type(s) Selection Rate Timing R-wave, S-wave, and T-wave 30 BPM to 120 BPM Waveform delayed from the ventricular demand pacemaker pulse by 80 % of the pulse-to-pulse interval or 400 mS (whichever is shorter) Active output Selected ventricular test load Square (SQU); triangle (TRI); haversine (HSN); sine square (SSQ) (default setting); asymmetrical triangle (ISO) – fixed width: 2 mS rise time/13 mS fall time 200 Ω, 500 Ω (default setting) and 1000 Ω ± 1 % Pacer load selection...

 Open the catalog to page 5
SigmaPace™ 1000-6

Refractory period test (atrial channel) Test selections Period Accuracy Resolution Physiological simulation Additional waveform selections Amplitude Active outputs Intended pacemaker types Pacemaker rates Available test load Paced refractory period; sensed refractory period 20 mS to 500 mS ± 5 % of reading (or ± 1 mS, whichever is greater) ± 1 LSD Selection Square wave (default setting) Atrial channel Simulated P-wave Width 1 mS Amplitude 20 mV peak Active outputs Atrial channel (4 mm banana jacks) only Square (SQU); triangle (TRI); haversine (HSN); sine square (SSQ); asymmetrical triangle (ISO);...

 Open the catalog to page 6
SigmaPace™ 1000-7

Resolution Display format Test selections Test load/input configurations Baseline/test selection Dynamic test dating algorithm Specified applied pacemaker rate 1 LSD (0.1 μA) 3 digits Static Continuous (power OFF) Dynamic Gated (power ON) Atrial + and atrial Ventricular + and ventricular Atrial + and ventricular + 500 Ω Measurement made 400 mS prior to the pacemaker pulse leading edge; 16 measurements averaged at a 4 mS rate for a total of 64 mS 80 PPM Current drain test DC current ranges Polarity Indicator Resolution Display format Accuracy Input dc voltage Selectable test loads Battery test...

 Open the catalog to page 7
SigmaPace™ 1000-8

General information Temperature Humidity range Modes of operation User interface Keys Serial port Dimensions (WxDxH) Weight Safety Operating 15 °C to 35 °C (59 °F to 95 °F) Storage 0 °C to 50 °C (32 °F to 122 °F) < 90 % noncondensing Manual, remote (via standard RS-232 serial port) Display 21-character x 8-line LCD readout; brightness/ viewing angle adjustment Eight push buttons [F-2, F-3 (UP arrow), F-4 (UP arrow), two DOWN arrows, ESCAPE, and ENTER] Type RS-232 Connector type DB-9 (male) Baud Rates 2400, 9600, and 19200 Data control Xon/Xoff External battery charger source/power supply 100...

 Open the catalog to page 8

Archived catalogs

  1. RAYSAFE 452

    9  Pages

  2. RaySafe X2

    12  Pages

  3. RaySafe DXR+

    2  Pages

  4. BP Pump 2

    82  Pages

  5. RaySafe Solo

    5  Pages

  6. DPM1B

    2  Pages

  7. INCU ?

    4  Pages

  8. 76-606DX

    2  Pages

  9. RaySafe Xi

    8  Pages

  10. VT305

    4  Pages

  11. ProSim

    8  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.