The GERSTEL Opioids/Cocaine Solution is an automated system designed for the quantitative analysis of opioids, cocaine, and their metabolites in blood serum.
Key Characteristics
  • Automated analysis of opioids, cocaine, and metabolites from 0.6 mL blood serum.
  • Complete automation post-manual protein precipitation and internal standard addition, including Solid Phase Extraction (SPE), eluate evaporation, derivatisation, and GC/MS analysis.
  • Capable of processing 97 samples in one sequence, with a throughput of 29 samples per day.
  • Calibration ranges for various substances: Cocaine (5-300 ng/mL), Benzoylecgonin (50-1500 ng/mL), Methadone (25-1500 ng/mL), Morphine (5-300 ng/mL), Codeine (2.5-150 ng/mL), 6-Monoacetylmorphine (2.5-150 ng/mL), Dihydrocodeine (5-300 ng/mL), 7-Aminoflunitrazepam (2.5-150 ng/mL).
System Components
  • Includes MultiPurpose Sampler MPS Dual Head, SPE for 1 mL cartridges, Multi Position Evaporation Station mVAP, standard agitator, solvent filling stations, trays for SPE cartridges, samples, and eluate vials, standard wash station, and syringes for injection and sample preparation.
  • Consumables required for one sample sequence include columns, liners, vials, caps, and SPE cartridges.
Methods
  • Includes analysis and quantification methods, example sequence for GC MassHunter Software with integrated MAESTRO software, and a detailed description of the analysis method.
System Requirements
  • Requires coupling to an Agilent Technologies 7890 GC / 5977 MSD system, configured separately.
  • Compatible with Agilent Technologies MassHunter software B.07.00.1413 and GERSTEL Cooled Injection System CIS 4/CIS 6.
  • Can be coupled to GC or LC systems from other manufacturers.
Important Notes
  • This method is intended for research use only and is not suitable for diagnostic procedures.
  • Repairs can only be conducted on-site, not at GERSTEL.
See more

Catalog excerpts

Opioids/Cocaine Solution-1

GERSTEL Opioids/Cocaine Solution System Components The GERSTEL Opioids/Cocaine Solution is an automated solution for the quantitative determination of opioids, cocaine and metabolites in blood serum. The components which are part of the system are listed below. All chemicals, the hardware and consumables required for the manual part of the sample preparation as well as the GC/MS analytical system are not included in the scope of delivery. Technische Änderungen vorbehalten. 01-062014 • Analysis of opioids, cocaine and metabolites from blood serum (0.6 mL) • Complete automation of sample preparation and injection into the GC after manual protein precipitation and addition of internal standards: –– Solid Phase Extraction (SPE), –– Evaporation of the eluate, –– Derivatisation, –– Injection into a CIS, –– GC/MS analysis • 97 samples in one sequence • 29 samples per day • Calibration range: –– Cocaine: 5-300 ng/mL, –– Benzoylecgonin: 50-1500 ng/mL, –– Methadone: 25-1500 ng/mL –– Morphine: 5-300 ng/mL –– Codeine: 2.5-150 ng/mL –– 6-Monoacetylmorphine: 2.5-150 ng/mL –– Dihydrocodeine: 5-300 ng/mL –– 7-Aminoflunitrazepam: 2.5-150 ng/mL Also pay attention to the specification sheets of the individual system components. Hardware • MultiPurpose Sampler MPS Dual Head, both towers equipped with gas supply • Solid Phase Extraction SPE for 1 mL cartridges, heating and shaking function are deactivated • Multi Position Evaporation Station mVAP with condenser pump, only one evaporation position can be used • Standard agitator • 2× Solvent Filling Station SFS 2 • Tray for SPE cartridges, 98 positions • Tray on standard tray holder for samples, 98 positions • Tray on standard tray holder for eluate vials, 98 positions • Standard wash

 Open the catalog to page 1
Opioids/Cocaine Solution-2

Opioide/Cocaine Solution • 2.5 mL syringe for sample preparation, right sampler • Consumables for one sample sequence: - column, - liner, - vials, - caps, - SPE cartridges Methods • Analysis method • Quantification method • Example sequence for GC MassHunter Software with integrated MAESTRO software, • Detailed description of analysis method System Requirements For operation of the Opioide/Cocaine solution the system must fulfill the following requirements. • Coupling to an 7890 GC / 5977 MSD system (Agilent Technologies). Must be configured separately (mounting hardware as well) • Agilent Technologies...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.