Description
The X11 optometer is a portable light measurement tool designed for field service. It features a compact design, supports up to four photodiodes, and displays data on a backlit four-line display. It is user-friendly with a menu-driven setup and can store settings for future use. It operates on two AA batteries or via USB for remote control.
Specifications
The X11 can function as a photometer, RGB-luminous color meter, radiometer, UV-radiometer, or laser-powermeter. It supports four measurement channels with a dynamic range from 0.1 pA to 200 µA. It offers seven gain ranges with manual or automatic selection and can store calibration data for multiple detector heads.
Interfaces
The device includes a USB interface for data transfer and remote control, and a 9-pin MDSM9 socket for detector connections.
Measurement Range and Calibration
The X11 supports a measurement range from 200 µA to 0.1 pA, with a linearity error of less than ±0.5% and a temperature coefficient of less than ±0.03%/°C. It can store up to 256 calibration datasets in EEPROM.
Menu Guide
The menu system includes options for mode selection, setup, detector selection, offset adjustment, and range setting.
Power Supply and Display
The device operates on two AA batteries or USB power, with a 250-hour operation time with the display backlight off. It features a 97 x 32 pixel LCD graphic display with switchable LED-backlight.
Compatible Detector Heads
The X11 is compatible with various detector heads for measuring different types of illuminance and irradiance, each with specific features and calibration certificates.
Product Overview
Gigahertz-Optik GmbH offers a range of detector heads and accessories for light measurement applications, each with a calibration certificate.
Detector Heads
  • RCH-102: For high-intensity UV or blue light curing processes.
  • PD-9304: Measures low-intensity LASER radiant power.
  • PD-11 Series: Modular heads compatible with integrating spheres.
  • UV-3706: For bilirubin phototherapy irradiance.
  • UV-3711-308: For 308nm Eximer Lasers.
  • UV-3724: For UV-B irradiance measurement.
  • UV-3709: For blue-light hazard measurements.
  • UV-3725: For UV-C 254 nm irradiance.
  • RW-37 with SRT-M37-L: Measures irradiance and radiance.
Software
  • S-SDK-X20: SDK for X20 variants.
  • S-X1: Application software for X1 variants.
Purchasing Information
  • X1-1: Meter with batteries, cable, and manual.
  • X1-1-V03: Optometer for detector head UV-3718.
  • X1-1-V02: Optometer for detector head UV-3726.
Accessories
  • Various adapter cables and hard cases for meters and accessories are available.
See more

Catalog excerpts

X1-1

Gigahertz-Optik GmbH

 Open the catalog to page 1
X1-2

versatile hand-held optometer Hand-held Meter The X11 optometer is one of the most versatile hand-held light measurement instruments available. It combines a powerful electronic design packaged in a light-weight ergonomic housing. Its compact size makes it ideal for field service applications. A unique feature of the X11 is its capability to operate detector heads housing up to four photodiodes with all four signals displayed on the four line display with on/off backlighting. When more than one detector head is used simple and faultless selection of the calibration data is accomplished using...

 Open the catalog to page 2
X1-3

Short description Main features Measurement range typical applications Calibration Detector interface Measurement range General Optometer for the individual configuration as photometer, RGB-luminous Color meter, Radiometer, UV-radiometer, LASER-powermeter etc. with the detector heads supplied. Compact gauge in ergonomic design for one Hand control. Four measurement chanels in multiplex operation for use with single and multi-chanel detectors heads. Back illuminated Display with four lines. Battery powered with two AA cells. USB-Interface. User Software and Software development kit available....

 Open the catalog to page 3
X1-4

Menu guide Menu item 1. Mode 2. Setup 3. Detector 4. Offset 5. Range Submenu item Function CW Measures respective of any offset and calibration factors programmed Dose Accumulates the single readings as exposure for measured quantity CIE Yxy & T Measures the CIE Color Values Yxy and T CIE Yuv & T Measures the CIE Color Values Yuv and T Zero Adjust Performs a zero adjustment of the internal amplifier and ADC Integration Sets the measurement (integration) time Dose Time Sets max. dose measurement time Selects calibration data to calculate the measurement result Performs an automatic offset adjustment...

 Open the catalog to page 4
X1-5

Product Description Show productImage VL-3702 Detector head for the measurement of photopic illuminance in Lux [lx]. Features: f1 < 6 %, f2 < 3 %, 0.5 nA/lx, 20mm height, for the usage with Optometers and amplifiers, calibration certificate Gigahertz-Optik GmbH

 Open the catalog to page 5
X1-6

Product Description Show productImage VL-3701 with SRT-M37-L Gigahertz-Optik GmbH

 Open the catalog to page 6
X1-7

Product Description Show productImage RW-3703 Detector head for the measurement of irradiance in W/m2. Features: spectral responsivity from 400-800nm (VIS), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate. Detector head for the measurement of irradiance in W/m2. http://www.gigahertz-o ptik.de/en-Features: spectral responsivity from 950-1700nm (NIR), cosine us/product/RW-3708 field-of-view, for the usage with optometers and amplifiers, calibration certificate. Detector head for the measurement of irradiance of UV radiation in W/m2. Features: spectral...

 Open the catalog to page 7
X1-8

Product Description Show productImage UV-3716 Detector head for the measurement of irradiance of UV radiation in W/m2. Features: spectral responsivity from 305-400nm (UV-A), cosine field-of-view, for the usage with optometers and amplifiers, calibration certificate. Gigahertz-Optik GmbH

 Open the catalog to page 8
X1-9

Product Description Show productImage RCH-0 Detector head for high intensity irradiation as in UV or blue light curing processes. Features: Separate light integrator and detector with flexible fiber coupling, light, 320-460nm UVABLUE responsivity, wide viewing angle, for the usage with optometers and amplifiers, calibration certificate. Gigahertz-Optik GmbH

 Open the catalog to page 9
X1-10

Product Description Show productImage UV-3725 Detector for the measurement of UV-C 254 nm irradiance in air disinfection applications. Features: wide dynamic range for UV hazard and effective irradiance, wide angle cosine F.O.V. for straylight measurements, for the usage with optometers, calibration certificate. Purchasing information Article-Nr Gigahertz-Optik GmbH

 Open the catalog to page 10
X1-11

Gigahertz-Optik GmbH

 Open the catalog to page 11

All Gigahertz-Optik catalogs and technical brochures

  1. X1-4

    6  Pages

  2. X1-3

    6  Pages

  3. BN-Rxx-D2

    3  Pages

  4. BN-LDSF-2P

    3  Pages

  5. BN-1501

    4  Pages

  6. VL-1101

    5  Pages

  7. UV-3725

    5  Pages

  8. UV-3717

    5  Pages

  9. UV-3711

    5  Pages

  10. UV-3710

    5  Pages

  11. BTS-Solar

    4  Pages

  12. UPB-150-ARTA

    10  Pages

  13. UPB-150-ART

    9  Pages

  14. TPI21-TH

    5  Pages

  15. TFCT25

    4  Pages

  16. X1-RM

    10  Pages

  17. P-2000

    16  Pages

  18. P-9710

    15  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.