RESPIRA DEVICE ADVANCED - 1

Invasive ventilation based on automation and monitoring of manual resuscitation bag C/ Caracas, 13-15, Nau 6 | 08030 Barcelona Phone +34 931 256 536 | info@gpainnova.com

Open the catalog to page 1
RESPIRA DEVICE ADVANCED - 2

T E C H N I C A L F E AT U R E S Mechanical movement: Vertical displacement with linear actuator with servo engine. Adjustable stroke to different sizes of manual resuscitators, both disposable and reusable. Use both with atmospheric air, as well as with oxygen supply and reservoir. Connectivity: The device can be connected to existing WLAN in medical centers. The data are .xml HL7 format and can be delivered through Ethernet TCP/IP. The device can be connected to own WI-FI with the Connectivity Module and Access Point. Accessories: Battery (SAI) with 45min autonomy Auxiliary ventilator...

Open the catalog to page 2
RESPIRA DEVICE ADVANCED - 3

TECHNICAL FEATURES Ventilation parameters G PAIN NOVA C/ Caracas, 13-15, Nau 6 I 08030 Barcelona Phone +34 931 256 536 | info@gpainnova.com

Open the catalog to page 3
RESPIRA DEVICE ADVANCED - 4

121 This option is applicable to all ventilation modes C/ Caracas, 13-15, Nau 6 | 08030 Barcelona Phone +34 931 256 536 | info@gpainnova.com

Open the catalog to page 4
RESPIRA DEVICE ADVANCED - 5

C O N T R O L V O L U M E V E N T I L AT I O N M O D E Ventilation mode with target tidal volume adjusting stroke of linear actuator. Itis a non assisted ventilation mode designed for sedated patient treatment. The physician set up tidal volume, frequency and I:E Ratio. This ventilation mode includes the possibility of autoadjustment of minimum and maximum tidal volume and PIP pressure. V E N T I L AT I O N PA R A M E T E R S ( C o n t r o l Vo l u m e ) Insufflation: Tidal volume adjustment by compression of the AMBU, from 50 ml to 800ml *(1) Ventilation frequency: Controlled: 10 - 30 rpm...

Open the catalog to page 5
RESPIRA DEVICE ADVANCED - 6

A S S I S T V O L U M E V E N T I L AT I O N M O D E This ventilation mode is usable for weaning ventilation,. The patient trigger the frequency. The physician set up tidal volume, Trigger sensibility and apnea time. This ventilation mode includes the possibility of autoadjustment of minimum and maximum tidal volume and PIP pressure. V E N T I L AT I O N PA R A M E T E R S ( A s s i s t Vo l u m e ) Insufflation: Tidal volume adjustment by compression of the AMBU, from 50 ml to 800ml *(1) Minimum pressure: Pressure limitated by external PEEP valve Maximum pressure: Pressure support based on...

Open the catalog to page 6
RESPIRA DEVICE ADVANCED - 7

A S S I S T - C O N T R O L V O L U M E V E N T I L AT I O N M O D E Weaning mode: the patient is the trigger of ventilation between frequencies. This ventilation mode includes the possibility of autoadjustment of minimum and maximum tidal volume and PIP pressure. V E N T I L AT I O N PA R A M E T E R S ( A s s i s t - C o n t r o l Vo l u m e ) Insufflation: Tidal volume adjustment by compression of the AMBU, from 50 ml to 800ml *(1) Ventilation frequency: Controlled: 10 - 30 rpm Assisted: Up to 60 rpm I:E Ratio: Adjustable from 1:1 to 1:3 with intervals of 0.5 Minimum pressure: Pressure...