video corpo

HI 3850

HI 3850

Product catalog summary
Introduction
Thank you for choosing a Hanna Product. This document provides instructions for using the HI 3850 Ascorbic Acid Test Kit, designed to determine the Vitamin C content in beverages through a drop-count iodometric titration method.
Kit Contents
  • HI 3850A-0 Ascorbic Acid Reagent, 100 mL
  • Starch Indicator, 25 mL with dropper
  • HI 3850C-0 Ascorbic Acid Reagent, 100 mL
  • 2 calibrated plastic vessels (50 mL)
  • 1 plastic pipette (3 mL)
  • 2 plastic pipettes (1 mL)
  • 1 graduated plastic test tube with cap
Significance and Use
The test kit is used to measure the ascorbic acid content in juice-based beverages. Ascorbic acid reacts with potassium iodate in acidic conditions, and the amount of potassium iodate consumed is measured iodometrically. The test is suitable for colored samples, but reducing substances may interfere with results.
Specifications
  • Range: 10 to 200 ppm Ascorbic Acid
  • Smallest Increment: 10 ppm
  • Analysis Method: Drop-Count Iodometric Titration
  • Sample Size: 10 mL
  • Number of Tests: Approximately 100
  • Case Dimensions: 235x175x115 mm
  • Shipping Weight: 519 g
Instructions
  1. Fill the test tube with 10 mL of the sample and dilute to 50 mL with deionized water in a calibrated vessel. Prepare a reference sample similarly.
  2. Add 1 mL of HI 3850A-0 reagent to one sample and mix.
  3. Add 4 drops of Starch Indicator and mix.
  4. Add HI 3850C-0 reagent dropwise, swirling and counting drops, until a persistent blue color develops. Compare with the reference sample.
  5. Calculate the concentration of Ascorbic Acid by multiplying the number of drops by 10 to get ppm.
Health and Safety
Handle chemicals with care and refer to Health and Safety Data Sheets before testing.
References
Skoog, West, Holler; Fundamentals of Analytical Chemistry; Saunders College Publishing International Edition (5th ed.)
See more

Catalog excerpts

HI 3850-1

SPECIFICATIONS other calibrated vessel and keep it as a reference for the initial color. HI 3850 Ascorbic Acid Test Kit Range developed. Compare the solution to the unreacted sample to be sure of the color change. Count the drops needed to obtain the color change. To calculate the concentration of Ascorbic Acid multiply by 10 the number of drops of HI 3850C-0 titration reagent used: • Using the plastic dropper add 1 mL of HI 3850A-0 reagent to one sample only. Swirl to mix. SIGNIFICANCE AND USE REFERENCES IIIHkNNk Dear Customer, Thank you for choosing a Hanna Product. Please read the instructions carefully before using the chemical test kit. It will provide you with the necessary information for correct use of the kit. Remove the chemical test kit from the packing material and examine it carefully to make sure that no damage has occurred during shipping. If there is any noticeable damage, notify your Dealer or the nearest Hanna office immediately. Each kit is supplied with: • HI 3850A-0 Ascorbic Acid Reagent, 1 bottle (100 mL); • Starch Indicator, 1 bottle with dropper (25 mL); • HI 3850C-0 Ascorbic Acid Reagent, 1 bottle (100 mL); • 2 calibrated plastic vessels (50 mL); • 1 plastic test tube, graduated with cap. Note: Any damaged or defective item must be returned 2 in its original packing materials. Ascorbic acid (Vitamin C) is added as a flavoring and preservative agent to juice-based beverages. This kit is especially designed for the determination of the content in Vitamin C of beverages. It is based on drop-count titration and is practical also with intensively colored samples. Reducing substances interfere in the determination of Ascorbic Acid with this test kit. Note: mg/L is equivalent to ppm (parts per million). CHEMICAL REACTION Ascorbic acid (C6H8O6) undergoes an oxidation reaction with potassium iodate in acidic condition. The quantity of potassium iodate consumed is then determined iodometrically. INSTRUCTIONS READ THE ENTIRE INSTRUCTIONS BEFORE USING THE KIT • Fill the graduated plastic test tube up to the 10 mL mark with the sample and pour it in the calibrated vessel. Add deionized water up to the 50 mL mark to dilute. Similarly prepare another sample using the Skoog, West, Holler; Fundamentals of Analytical Chemistry; Saunders College Publishing International Edition (5th ed.) HEALTH AND SAFETY The chemicals contained in this test kit may be hazardous if improperly handled. Read Health and Safety Data Sheets before performing the test. Add drops of HI 3850C-0 reagent, while swirling and counting the drops, until a persistent blue color is