video corpo

Bed-in-bed systems

Bed-in-bed systems
1 / 44 PagesView full catalog

Bed-in-bed systems

Product catalog summary
Preface and General Instructions
The manual introduces Hermann Bock GmbH's nursing care beds, highlighting their compliance with EN 60601-2-52 and EN 60601-1 standards, classifying them as Class 1 medical devices. These beds are designed for elderly care environments, excluding hospitals, and are not intended for patient transport.
Intended Purpose
The beds are suitable for adults over 146 cm in height and weighing between 40 kg and 185 kg, intended for use in elderly homes, nursing homes, rehabilitation facilities, and home care.
Safety Instructions
Proper installation and operation are crucial to avoid dangerous situations. The manual advises disabling the bed's functions when unattended and ensuring no limbs are in the adjustment range during operation. Only trained personnel should operate the bed.
Service Life and Warranty
The expected service life is 2 to 10 years, depending on usage. Unauthorized modifications void the warranty. The product is not approved for the North American market.
Installation Requirements
Floor requirements are specified to prevent damage, and potential electromagnetic interactions with other devices are noted.
Protection and Disposal
The device adheres to Protection Class II standards, ensuring double insulation. Within the EU, it must be disposed of via separated municipal waste.
General Description
The beds are designed for durability and safety, with corrosion protection for metal parts. The lying surface is divided into four functional areas and is adjustable via low-voltage DC motors.
Safety and Hazard Notes
Emphasizes the importance of using original parts, checking for damages, and ensuring secure installation of side rails to prevent injury.
Electric Parts
The drive unit includes a double drive for adjustments, with a protective low voltage of up to 35 VDC. The system includes a 9V battery for emergency lowering.
Drive Systems
Equipped with Limoss drive systems, featuring a double drive for lying surface adjustments and a linear drive for height adjustments.
Hand Control
Features a locking device to prevent unauthorized use and is ergonomically designed with large keys for ease of use.
Technical Data
Provides detailed specifications for different bed models, including dimensions, weight capacities, and power requirements.
Assembly and Operation
Outlines specific requirements for setting up bed-in-bed systems to minimize crushing risks.
Specifications and Safety Notes
Various feet for bed frames ensure compliance with a minimum distance to prevent crushing hazards.
Combiflex BIBS
Designed for home care and rehabilitation, offering comfort and easy operation. Not suitable for hospital use or patient transport.
Assembly Instructions
Includes steps for removing packaging, assembling chassis and lying surface, and securing components.
Steel Side Protection
Increases bed width by 10 mm and ensures proper distances between bed parts and side rails for safety.
Change of Location
Instructions for safely moving the bed, including securing the power cable and inspecting for damage.
Transport, Storage, and Operating Conditions
Specifies temperature and humidity ranges for transport, storage, and operation.
Function Notes
Includes instructions for locking brakes on castors and ensuring minimum mattress height above side rail.
Disposal
Components are recyclable, following WEEE-EC directive 2012/19/EC for disposal of electronic components.
Troubleshooting
Provides solutions for common malfunctions such as power issues and obstructions.
Accessories
Offers a range of accessories to customize beds to individual needs.
Mounting Accessories
Standard accessories include a lifting pole with a safe working load of 75 kg.
Specifications and Components
Describes components like a lifting pole with a triangle handle and a hook-up loop, with specific adjustment and service life details.
Mattress Specifications
Suitable mattresses include foam and latex with specific volumetric weight and dimensions.
Safety and Accessories
Only original Bock accessories should be used to avoid accidents and damages.
Cleaning and Maintenance
Instructions for cleaning bed elements using wipe and spray disinfection methods.
Electromagnetic Emission and Immunity
Guidelines for maintaining separation distances from RF communications equipment to prevent interference.
Regular Inspections
Annual safety-technical inspections are mandated, with a checklist provided for visual inspections, functional testing, and electric measurements.
Inspection and Maintenance Procedures
Includes details on regular inspections, functional testing, and electric measurements to ensure safety and compliance.
See more

Catalog excerpts

Bed-in-bed systems-1

Assembly and operation manual Bed-in-bed systems • combiflex bibs • belluno bibs • unilift

 Open the catalog to page 1
Bed-in-bed systems-2

Dear valued customer, with your decision to purchase a nursing care bed from Hermann Bock GmbH, you are receiving a long-lasting care product with superior functionality at the highest safety level. Our electrically operated nursing care beds guarantee optimal lying comfort and allow professional care at the same time. This product was designed with a focus on the elderly, whose confidence must be reinforced and whose life needs protection. With this health care product, we meet these requirements. We urge you to prevent potential malfunctions and the risk of accidents by complying strictly with...

 Open the catalog to page 2
Bed-in-bed systems-4

Preface and general instructions The various bed systems from Hermann Bock meet special requirements for the use in care and treatment facilities as well as for home care. Reliable functionality and a long product life make each bed particularly valuable. Our beds need little maintenance with proper operation and care. Each bed from Hermann Bock must pass quality testing in a final inspection before it is shipped anywhere. The beds are manufactured according to the current standards for medically used beds and tested accordingly. The beds comply with the EN 60601-2-52 standard. The electrical...

 Open the catalog to page 4
Bed-in-bed systems-5

The Trendelenburg function may be used exclusively under supervision of medical professionals. The beds, which are determined for application environment type 4, are equipped with a hand control which is unable to operate the Trendelenburg function. The nursing care bed is not suitable for use in hospitals. It is also not designed to transport patients. The beds can only be moved within the patient's room - even during patient positioning - for cleaning or better access to the patient, for example. Attention: The beds come with no special connection options for a potential equalisation. Electrical...

 Open the catalog to page 5
Bed-in-bed systems-6

Safety instructions The intended use/operation of all moving parts is as important for the safety of the person in need of care as well as for the relatives and the caregivers/nursing staff to avoid potentially dangerous situations. This requires the correct installation and operation of the bed. The individual physique of the person in need of care as well as type and the extent of their disability must be taken into account by all means when operating the bed. Avoid dangers, accidental motor adjustments and incorrect operation by using the disabling function. When the operator, e.g. the nursing...

 Open the catalog to page 6
Bed-in-bed systems-7

The permitted person’s weight depends on the total weight of the equipment that has been mounted to the bed (mattresses and other electronic medical devices). For safe working load, please refer to the type plate on the lying surface frame of the bed. 1.4 Service life / warranty This nursing care bed was developed, designed and manufactured for safe operation over a long period of time. With proper operation and maintenance, this nursing care bed has an expected service life of 2 to 10 years. The service life depends on operating conditions and frequency. In the furnishing sector, for example,...

 Open the catalog to page 7
Bed-in-bed systems-8

Each nursing care bed is marked with an individual type plate. Either there is only one sticker on your nursing care bed (Fig. 1, previous nursing care bed models) or there are two separate type plates (Fig. 2, current nursing care bed models) on your nursing care bed. Type plate for previous nursing care bed models Model: xxx Date of manufacturing: xx.xx.xxxx Serial no. xxxxxxxxx-xxx XXX V — XX HZ max. X A Duty cycle xx % (x min ON / xx min OFF) Motor protection class IPX4 Hermann Bock GmbH - Nickelstr 12 D-33415 Veil / TeJ. ♦ 49(0)1805/262500 (2) Manufacture date: Day, month and year (3) Serial...

 Open the catalog to page 8
Bed-in-bed systems-9

Type plate for current nursing care bed models Model designation in German (DE) and English (EN) Serial number Item numbers Mains voltage, mains frequency, power input, Duty cycle (INT) Drive protection class Date of manufacture (year-month) Patient weight Safe working load Machine readable serial number Patient population Follow the instructions appropriate for mattress size and thickness Conformity mark according to the medical device directive Protection class II (double insulation, insulated for protection) Within the European Union, this product must be disposed via the separated municipal...

 Open the catalog to page 9
Bed-in-bed systems-10

General description of the functions Construction design and function Corrosion protection The Hermann Bock GmbH nursing care beds are developed and constructed in such a way that they can function long and safely. For this reason, all materials that may corrode are protected accordingly. All metal parts are equipped with a surface protection. The steel parts are either galvanised or stove-enamelled with a PES powder coating and the aluminium profiles are anodised. The lying surface with 4 function areas The lying surface consists as standard of a slatted comfort frame (can alternatively be fitted...

 Open the catalog to page 10
Bed-in-bed systems-11

The chassis The height adjustment of the beds is carried out via a base frame with single or double drive. The surface of the tubular steel structure is stove-enamelled with a PES-powder coating. The side rail The bed-in-bed systems can be equipped with a clip-on steel side guard on both sides. Hazard note from Bock If the bed-in-bed system is to be additionally equipped with a side rail, the mattress bracket on the lying surface must be removed. Hazard note from Bock When the lying surface is lowered, there is an increased risk of crushing in the area of the four integrated castors or in the...

 Open the catalog to page 11
Bed-in-bed systems-12

All dimensions in mm. (*) Depending on the length of the lying surface. Distance between the head part and the side rail Height 1 of side rail Height 2 of side rail Width of the side rail Distance between the elements within the side rail Distance between the lying surface and the upper edge of the side rail Height of the top edge of the side rail above the mattress without compression Thickness of the mattress for the intended use Smallest dimension between side rail and lying surface (or the panel, if any) Distance between the foot part and the side rail

 Open the catalog to page 12

All Hermann Bock GmbH Pflegebetten und Objekteinrichtung catalogs and technical brochures

  1. stellar

    6  Pages

  2. domiflex

    48  Pages

  3. Furniture

    24  Pages

  4. Chairs, Tables

    19  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.