Bed-in-bed systems
44Pages

{{requestButtons}}

Catalog excerpts

Bed-in-bed systems - 1

Assembly and operation manual Bed-in-bed systems • combiflex bibs • belluno bibs • unilift

Open the catalog to page 1
Bed-in-bed systems - 2

Dear valued customer, with your decision to purchase a nursing care bed from Hermann Bock GmbH, you are receiving a long-lasting care product with superior functionality at the highest safety level. Our electrically operated nursing care beds guarantee optimal lying comfort and allow professional care at the same time. This product was designed with a focus on the elderly, whose confidence must be reinforced and whose life needs protection. With this health care product, we meet these requirements. We urge you to prevent potential malfunctions and the risk of accidents by complying strictly...

Open the catalog to page 2
Bed-in-bed systems - 4

Preface and general instructions The various bed systems from Hermann Bock meet special requirements for the use in care and treatment facilities as well as for home care. Reliable functionality and a long product life make each bed particularly valuable. Our beds need little maintenance with proper operation and care. Each bed from Hermann Bock must pass quality testing in a final inspection before it is shipped anywhere. The beds are manufactured according to the current standards for medically used beds and tested accordingly. The beds comply with the EN 60601-2-52 standard. The...

Open the catalog to page 4
Bed-in-bed systems - 5

The Trendelenburg function may be used exclusively under supervision of medical professionals. The beds, which are determined for application environment type 4, are equipped with a hand control which is unable to operate the Trendelenburg function. The nursing care bed is not suitable for use in hospitals. It is also not designed to transport patients. The beds can only be moved within the patient's room - even during patient positioning - for cleaning or better access to the patient, for example. Attention: The beds come with no special connection options for a potential equalisation....

Open the catalog to page 5
Bed-in-bed systems - 6

Safety instructions The intended use/operation of all moving parts is as important for the safety of the person in need of care as well as for the relatives and the caregivers/nursing staff to avoid potentially dangerous situations. This requires the correct installation and operation of the bed. The individual physique of the person in need of care as well as type and the extent of their disability must be taken into account by all means when operating the bed. Avoid dangers, accidental motor adjustments and incorrect operation by using the disabling function. When the operator, e.g. the...

Open the catalog to page 6
Bed-in-bed systems - 7

The permitted person’s weight depends on the total weight of the equipment that has been mounted to the bed (mattresses and other electronic medical devices). For safe working load, please refer to the type plate on the lying surface frame of the bed. 1.4 Service life / warranty This nursing care bed was developed, designed and manufactured for safe operation over a long period of time. With proper operation and maintenance, this nursing care bed has an expected service life of 2 to 10 years. The service life depends on operating conditions and frequency. In the furnishing sector, for...

Open the catalog to page 7
Bed-in-bed systems - 8

Each nursing care bed is marked with an individual type plate. Either there is only one sticker on your nursing care bed (Fig. 1, previous nursing care bed models) or there are two separate type plates (Fig. 2, current nursing care bed models) on your nursing care bed. Type plate for previous nursing care bed models Model: xxx Date of manufacturing: xx.xx.xxxx Serial no. xxxxxxxxx-xxx XXX V — XX HZ max. X A Duty cycle xx % (x min ON / xx min OFF) Motor protection class IPX4 Hermann Bock GmbH - Nickelstr 12 D-33415 Veil / TeJ. ♦ 49(0)1805/262500 (2) Manufacture date: Day, month and year (3)...

Open the catalog to page 8
Bed-in-bed systems - 9

Type plate for current nursing care bed models Model designation in German (DE) and English (EN) Serial number Item numbers Mains voltage, mains frequency, power input, Duty cycle (INT) Drive protection class Date of manufacture (year-month) Patient weight Safe working load Machine readable serial number Patient population Follow the instructions appropriate for mattress size and thickness Conformity mark according to the medical device directive Protection class II (double insulation, insulated for protection) Within the European Union, this product must be disposed via the separated...

Open the catalog to page 9
Bed-in-bed systems - 10

General description of the functions Construction design and function Corrosion protection The Hermann Bock GmbH nursing care beds are developed and constructed in such a way that they can function long and safely. For this reason, all materials that may corrode are protected accordingly. All metal parts are equipped with a surface protection. The steel parts are either galvanised or stove-enamelled with a PES powder coating and the aluminium profiles are anodised. The lying surface with 4 function areas The lying surface consists as standard of a slatted comfort frame (can alternatively be...

Open the catalog to page 10
Bed-in-bed systems - 11

The chassis The height adjustment of the beds is carried out via a base frame with single or double drive. The surface of the tubular steel structure is stove-enamelled with a PES-powder coating. The side rail The bed-in-bed systems can be equipped with a clip-on steel side guard on both sides. Hazard note from Bock If the bed-in-bed system is to be additionally equipped with a side rail, the mattress bracket on the lying surface must be removed. Hazard note from Bock When the lying surface is lowered, there is an increased risk of crushing in the area of the four integrated castors or in...

Open the catalog to page 11
Bed-in-bed systems - 12

All dimensions in mm. (*) Depending on the length of the lying surface. Distance between the head part and the side rail Height 1 of side rail Height 2 of side rail Width of the side rail Distance between the elements within the side rail Distance between the lying surface and the upper edge of the side rail Height of the top edge of the side rail above the mattress without compression Thickness of the mattress for the intended use Smallest dimension between side rail and lying surface (or the panel, if any) Distance between the foot part and the side rail

Open the catalog to page 12

All Hermann Bock GmbH Pflegebetten und Objekteinrichtung catalogs and technical brochures

  1. stellar

    6 Pages

  2. domiflex

    48 Pages

  3. Chairs, Tables

    19 Pages

  4. Furniture

    24 Pages