1. Catalogs
  2. Hingmed Medical
  3. DBP-01P Meet Specifications of ISO 81060-2:2018 Clinically Validated

DBP-01P Meet Specifications of ISO 81060-2:2018 Clinically Validated

DBP-01P Meet Specifications of ISO 81060-2:2018 Clinically Validated

DBP-01P Meet Specifications of ISO 81060-2:2018 Clinically Validated

Product catalog summary
Scope of the Study
This study evaluates the Hingmed DBP-01P, a Clinical Automatic Blood Pressure Monitor, following the ISO 81060-2:2018 protocol. The device was tested on a general population using a representative sample from mass production.

Methodology
The study was conducted at Villa Maria Hospital in Padova, Italy, using the same arm sequential method for blood pressure measurement. A wide-range cuff was used, suitable for arm sizes from 17 cm to 42 cm. Data handling was overseen by Prof. Paolo Palatini.

Study Criteria
The study involved 85 subjects, with measurements taken at least three times per subject. Observers were required to have a difference of ±4mmHg or less. The study included both male and female participants, aged over 12 years, with specific arm circumference distribution requirements.

Results
The study met all criteria, including:
  • Number of subjects: 85 (PASS)
  • Number of measurements: 7 times (PASS)
  • Observer difference: ±4mmHg or less (PASS)
  • Gender distribution: Male 50.6%, Female 49.4% (PASS)
  • Age: 100% over 12 years (PASS)
  • Arm circumference distribution met specified ranges (PASS)
  • Blood pressure range met specified percentages (PASS)
  • Measurement results: Mean value and standard deviation within acceptable limits for both systolic and diastolic BP (PASS)
Additional subjects were evaluated but excluded due to observer differences, within-subject differences, or withdrawal of consent.

Conclusion
The Hingmed DBP-01P met the ISO 81060-2:2018 protocol specifications, with all criteria satisfied for both systolic and diastolic blood pressure measurements. Scatter plots were used to illustrate the differences between device and observer measurements.

Report Author
Paolo Palatini, MD
Date: November 8, 2022
See more

Catalog excerpts

DBP-01P Meet Specifications of ISO 81060-2:2018 Clinically Validated-1

Validation of the device Hingmed DBP-01P, a Clinical Automatic Blood Pressure Monitor with wide-range cuff, evaluated in a general population according to the ISO 81060-2:2018 protocol. This validation study assessed the performance of the device “DBP-01P”, a Clinical Automatic Blood Pressure Monitor designed and manufactured by Shenzhen Hingmed Medical Instrument Co.,Ltd. according to the ISO 81060-2:2018 protocol. Devices under test (SN: 30120220415002) are representative samples from mass production. The clinical investigation was performed in a general population. Clinical protocol : ISO81060-2:2018 BP measurement: Same arm sequential method Test Device : Hingmed DBP-01 P Cuff Size : One wide-range cuff was used for arm sizes ranging from 17 cm to 42 cm Place of test : Villa Maria Hospital, Via Delle Melette, 20 - 35138 Padova, Italy Hospital Staff: Dr Claudio Fania, MD (Principal investigator); Antonella Giletto, RN; Valentina Zanello, RN (Villa Maria Hospital). Handling and processing of data: Prof. Paolo Palatini (Studium Patavinum, University of Padova, Padova, Italy).

 Open the catalog to page 1
DBP-01P Meet Specifications of ISO 81060-2:2018 Clinically Validated-3

Standard Deviation (Refer to Table 1) Additional subjects evaluated but not included in the analysis (n=10) • Between-observer differences > 4 mmHg (n=5) • Within-subject SBP or DBP differences > 12/8 mmHg (n=3) Clinical test of the BP measurement for Hingmed DBP-01P upper-arm BP monitor: Criterion1 was satisfied because the mean device-observer difference was < 5 mmHg and SD was < 8 mmHg for both systolic BP and diastolic BP. Criterion 2 was also satisfied because SD was below the values reported in table 1 of the protocol for both systolic BP and diastolic BP. The Hingmed DBP-01P met the specifications...

 Open the catalog to page 3
DBP-01P Meet Specifications of ISO 81060-2:2018 Clinically Validated-4

Scatter plots of systolic and diastolic blood pressure differences between the device and the observers (y-axis) against the average of the test device and observer pressure values (x-axis).

 Open the catalog to page 4

All Hingmed Medical catalogs and technical brochures

  1. ABPM software

    39  Pages

  2. ABP-03

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.