Introduction
The document details the validation of the Hingmed WBP-02A, an upper arm oscillometric monitor, for ambulatory blood pressure measurement in a general population. The validation follows the European Society of Hypertension International Protocol revision 2010.

Device Specifications
  • Brand: Hingmed
  • Model: WBP-02A
  • Manufacturer: Shenzhen Hingmed Medical Instrument Co., Ltd
  • Location: Upper Arm
  • Method: Oscillometry
  • Purpose: Ambulatory Blood Pressure Monitoring (ABPM)
  • Operation: Fully Automatic
  • Arm Cuffs: Small Adult (18.0 cm to 26.0 cm), Standard Adult (22.0 cm to 32.0 cm), Large Adult (26.0 cm to 36.0 cm), Other (30.0 cm to 43.0 cm)

Methodology
The study involved 37 participants, with 33 completing the study. Participants were recruited from outpatient clinics and hospital staff. The validation followed the ESH-IP 2010 protocol, with measurements taken by two independent observers.

Results
  • Participants: 14 males, 19 females, aged 32 to 91 years.
  • Arm Circumference: 20.0 cm to 43.0 cm.
  • Blood Pressure Ranges: SBP 96-172 mmHg, DBP 48-124 mmHg.
  • Validation Results: The device passed all validation criteria with SBP and DBP achieving high accuracy levels.

Discussion
The WBP-02A met the ESH-IP standards, demonstrating accuracy across various blood pressure levels. Recruitment challenges were noted for high BP ranges, but discrepancies were minimal and unrelated to BP levels.

Conclusion
The Hingmed WBP-02A is recommended for clinical and personal use in the general population due to its compliance with ESH-IP standards.

Acknowledgements
The study was funded by Hingmed, China, with devices supplied by the manufacturer.
See more

Catalog excerpts

Hingmed WBP-02A ESH validation-1

ESH-IP2 Validation of the Hingmed WBP-02A in oscillometry mode, in a general population. Validation of the Hingmed WBP-02A upper arm oscillometric monitor for ambulatory blood pressure measurement in a general population, according to the European Society of Hypertension International Protocol revision 2010 Claudio Fania,MD, Leonardo Vezzù,RN, Ilaria Lazzaretto,RN, Paolo Palatini, MD Keywords: Validation, device, oscillometric, blood pressure, ambulatory. For Claudio Fania, Leonardo Vezzù, and Ilaria Lazzaretto: Casa di Cura Villa Maria, via delle Melette, 20, 35138 Padova, Italy For Paolo Palatini: via San Fris, 121, 31029 Vittorio Veneto, Italy Correspondence to Doctor Claudio Fania, Casa di Cura Villa Maria, via delle Melette, 20, 35138 Padova, Tel: 049 8711144; e-mail: [email protected] Abstract The Hingmed WBP-02A, an upper arm oscillometric monitor, for ambulatory BP measurement was validated in a general population, according to the European Society of Hypertension International Protocol revision 2010. The protocol requirements were followed precisely. The device passed all of the requirements fulfilling the standards of the protocol. Thus, the WBP-02A device is recommended for clinical use in the adult general population. Device Details Brand Shenzhen Hingmed Medical Instrument Co., Ltd Fully Automatic Small Adult: 18.0 cm to 26.0 cm, Standard Adult: 22.0 cm to 32.0 cm, Large Adult: 26.0 cm to 36.0 cm and other cuffs: 30.0 cm to 43.0 cm Methodology Familiarisation Forty test measurements were carried out. No problems were encountered. Recruitment The WBP-02A device was evaluated in 37 adult individuals with BPs in the ranges required by the protocol. The participants were recruited from the outpatient clinics, wards, and from among the healthcare personnel of the Hospital Villa Maria, Padua, Italy. Four participants were excluded because their BP was within the BP ranges already completed, leaving an overall number of 33 participants. There was some difficulty in recruiting patients with BP in the high ranges, but apart from this there was no problem.

 Open the catalog to page 1
Hingmed WBP-02A ESH validation-2

ESH-IP2 Validation of the Hingmed WBP-02A in oscillometry mode, in a general population. Screening and Recruitment Details Screening and Recruitment Recruitment Ranges Total Screened Total Excluded Ranges Complete Ranges Adjustment Device Failure Poor Quality Sounds Observer Disagreement Other Reasons Cuff Size Unavailable Medium High Total Recruited Medium High The European Society of Hypertension International Protocol revision 2010 for the validation of blood pressure measuring devices in adults was followed precisely.[1] Overseen by an independent supervisor, measurements were recorded by...

 Open the catalog to page 2
Hingmed WBP-02A ESH validation-3

ESH-IP2 Validation of the Hingmed WBP-02A in oscillometry mode, in a general population. Validation Results Part 1 Mean (mmHg) Pass Requirements Pass Requirements Pass Result PASS Discussion The present study demonstrated that the WBP-02A BP monitor met the ESH-IP standards for use in the general adult population because it passed all phases of the 2010 ESH revision Protocol. The recruitment of individuals in the high BP range proved to be difficult and accounted for the extra number of screened participants. Visual inspection of the device– observer discrepancies showed that the error was unrelated...

 Open the catalog to page 3
Hingmed WBP-02A ESH validation-4

ESH-IP2 Validation of the Hingmed WBP-02A in oscillometry mode, in a general population. O’Brien E, Atkins N, Stergiou G, Karpettas N, Parati G, Asmar R, Imai Y, Wang J, Mengden T, Shennan A; on behalf of the Working Group on Blood Pressure Monitoring of the European Society of Hypertension. European Society of Hypertension International Protocol revision 2010 for the Validation of Blood Pressure Measuring Devices In Adults. Blood Press Monit 2010;15:23–38