SZ-100-1

Nanoparticle Analyzer A highly advanced analyzer solves the mysteries of the nano-world. A single device analyzes the three parameters that characterize nanoparticles: particle size, zeta potential, and molecular weight

 Open the catalog to page 1
SZ-100-2

Newly developed to satisfy the need for devices to simply and accurately evaluate the size and dispersion stability of nanoparticles, the key to nanotechnology advancement: SZ -100 Series Nanoparticle Analyzer   Nanotechnology research and development is a continuously evolving effort to control substances at the atomic and molecular level in order to achieve new and better materials and products. The miniaturization of components – that is, control at the nanolevel – is necessary to achieve faster, higher-performance devices and functions and to reduce energy consumption. Nanotechnology has...

 Open the catalog to page 2
SZ-100-3

Simply fill the sample cell and place the cell in the analyzer. Simple and Convenient Operation [ Cell Set-Up ] [ Start Measurement ] [ Results Display ]

 Open the catalog to page 3
SZ-100-4

Particle Size Measurement Principle Features of HORIBA’s Optical System 1 G (2)(τ) = B+Bf exp (-2D mq 2τ) G(2)( ) :Measured amplitude autocorrelation function B :So-called baseline f :Instrument constant Dm :Particle diffusion coefficient τ q :Scattering vector given by (4πn/λ)sin(θ/2) τ : Delay time D h=kT/(3πηD m) Dh :Particle hydrodynamic size k :Boltzman constant T :Thermodynamic t

 Open the catalog to page 4
SZ-100-5

Zeta Potential Measurement Principle (Laser Doppler Electrophoresis) Molecular Weight Measurement Principle ε ζ :Zeta potential U :Electrical mobility E :Electric field strength n :Solvent refraction index ε :Solvent dielectric constant η :Solvent viscosity fκ

 Open the catalog to page 5
SZ-100-6

Simple and Convenient Operation/Software Functions ●

 Open the catalog to page 6
SZ-100-8

Navigation files turn complex parameter input into simple to use operating procedures. / Store 100 data items on a data list. / Display individual data items with a single mouse click. / Perform pH, temperature, and sample concentration trend measurement. Real-time display of the autocorrelation function / Display of median size, specific surface area, mode size, average size, standard deviation, coefficient of variation, span value, percentage size (max. of 10 items), Z average, polydispersity index, size percentage (max. of 10 items displayed) / Particle distribution graph, autocorrelation function,...