Technical Specifications:
  • Operates under tap water pressure with a purification process involving a sediment pre-filter (5µm), an integrated module (sediment-carbon-softening), and reverse osmosis.
  • Device efficiency is 10 dm3/h with a retention level of 97-99%.
  • Equipped with a 24V pump to increase feed water pressure and includes automatic features.
  • Features a nozzle with a minimum range of 2 meters and a mobile, adjustable inox arm for water collection points.
  • Optional installation with a pressure tank and connection to devices like washers and autoclaves.
  • Maintenance is user-friendly with easy disposable replacements.
  • Operates with cold water (5-40ºC) and requires a power supply of 230V/50Hz.
  • Dimensions: 235x470x570 mm.
Monitoring Functions:
  • Equipped with a 24V microprocessor control system featuring a color display with a touch panel.
  • Includes a conductometer for measuring conductivity and temperature, with thermal compensation.
  • Displays system status, membrane module retention level, and maintenance dates.
  • Provides alarms for module replacements and includes graphic and sound signals.
  • Built-in RS 232 and USB connections for maintenance adjustments.
Protection Functions:
  • Pump shutdown if feed water pressure is too low or the tank is full.
  • Thermal protection for the RO module, with automatic shutdown if water temperature is outside 4˚C to 40˚C.
  • Automatic system shutdown upon any alarm occurrence.
Purity and Usage:
  • Produces demineralized water meeting ISO 3696:1999A standard for third class water purity.
  • Conductivity is less than 1200 µS/cm, with pressure over 3.0 bar and hardness under 250 mg CaCO3/dm3.
  • Suitable for mediums, buffer solutions, reagents, and can supply laboratory devices.
Connections Required:
  • Cold tap water connection (½” or ¾”), 230V socket, and a drain.
See more

Catalog excerpts

TECHNICAL 10/C-2

Technical specification: • Works under tap water pressure. • Water purification process levels: - sediment pre-filter 5µm, - integrated module (sediment-carbon-softening) – module A, - reverse osmosis. • Device efficiency: 10 dm3/h. • Equipped with a pump 24V (increasing the feed water pressure) with automatics. • Retention level: 97-99%. • Unattended and automated. • Equipped with a nozzle with a minimum range of 2 meters. • Mobile, adjustable inox arm holding water collection points – available adjustment ranges: up/down, front/back, left/right • Optional installing with a pressure tank. •...

 Open the catalog to page 2
TECHNICAL 10/C-3

Purity class: Demineralized water fits the ISO 3696:1999A standard for third class of water purity (it depends on the quality of feed water, retention level is 97-99%). Conductivity < 1200 µS/cm Pressure > 3,0 bar Temperature: 5-40°C Hardness < 250 mg CaCO3/dm3 Fe < 0,2 mg/dm3 Usage: May be used for mediums, buffer solutions, reagents. Can supply other laboratory devices: washers, analyzers, autoclaves, polishing systems etc. - Prefilters: • Sediment prefilter 5µm, • Module A Required connections: - cold tap water connection ½” lub ¾”, - 230V socket, - drain. Models produced from April 2013 r....