APPLICATION Rotating a node x-ray tube specifically designed for surgical C-arms, CHAU ACTE RISTKS -TZM target with large thermal capacity and 3" diameter -Focal track in tungsten-rhenium alloy to prevent cracking • High-thermal emittance oxide coating on anode and rotor provides high anode heat dissipation rate -Severe tests during conditioning assure best performances RTM ROTATING ANODE X-RAY TUBE FOR SURGICAL C-ARM SPECIFICATIONS Maximum peak voltage 130 kv Anode angle and diameLer iov-73 mm Maxi mi 1 id anode d issi pa r Ion 500 W (4 is qod H Li/m n) Anode heat storage capacity 215 kJ £300 kHU} Rotating anode sp«d3000 rpm FOCAL SPOTS AND POWERS
Open the catalog to page 1IAE spa ITALY via Fabio Fibi. 53 20031 COftMANO Ml tel +39 0266303255 fax+39 026152544 [email protected] www.iae.it lAE FRANCE sas 90,. BI vd St .Michel 91150 ETAMPES tel+33 164943044 fax+33 164949956 [email protected] Shou Id y cu still have quett iorrs a bout the coi n pot ibil ity ot a re place me nt tu be u n it co II aur eu stomc r sc rv ice: -+39 Ol 66303255 § Cepvrvtj-TT I^EiP* rtflUil 1M3 DoH nnov h1 WltbWilt nbtttc