MACHINE QA INDEPENDENT & INTEGRATED
10Pages

{{requestButtons}}

Catalog excerpts

MACHINE QA INDEPENDENT & INTEGRATED - 1

MACHINE QA INDEPENDENT & INTEGRATED Machines Software, Detectors, Phantoms & Accessories

Open the catalog to page 1
MACHINE QA INDEPENDENT & INTEGRATED - 2

… Fundamental for Patient Safety Dosimetry QA Linac beam and MLC checked Linac imaging systems checked Morning QA Treatment system ready for the day Treatment Safety Peace of mind Accurate Linac Machine QA is a fundamental requirement for consistently safe and efficient patient treatments. IBA Dosimetry is your proven partner to check and track all Machine QA needs, from daily to annual QA. Our imaging and dosimetry solutions provide highly efficient workflows and accurate analysis of your data. IBA Dosimetry We Protect, Enhance and Save Lives. Innovation leadership in Machine QA IBA...

Open the catalog to page 2
MACHINE QA INDEPENDENT & INTEGRATED - 3

Integrated Software Platform Machines |  Integrated Machine QA Central Database Accept SMARTSCAN Machines Accept Beam Scanning myQA Machines, the complete protocol-based independent machine QA integrated on one platform. Patients Individual QA Tests Cockpit – The leading Quality Assurance Platform. Integrated & Independent ■ Full coverage of tests related to dosimetry, imaging, MLC QA… and more! ■ Easy QA execution with intuitive and consistent software menu for all machine QA software modules (Plug-Ins) ■ Designed to integrate seamlessly with the myQA Plattform ■ Protocol-based machine QA...

Open the catalog to page 3
MACHINE QA INDEPENDENT & INTEGRATED - 4

Machines | Dosimetry Machines |  MLC & VMAT QA New release Dosimetry Plug-In Perform automated dosimetry tests with the StarTrack, MatriXX, or Dolphin detector arrays, or by loading water phantom measurements. Automated MLC stripe tests (‘picket fence test’) The New MLC Plug-In, for instant results and overview ■ Complete Overview: All data on one screen incl. reference image and pass / fail display ■ Completely automatic image pre-processing ■ EPID and film image analysis to determine leaf position accuracy & MLC transmission characteristics Efficiency ■ Acquire all key beam parameters in...

Open the catalog to page 4
MACHINE QA INDEPENDENT & INTEGRATED - 5

Machines |  Imaging QA New release The New 2D-Imaging Plug-In, driving workflow efficiency ■ Test efficiency: Pass / Fail clearly displayed ■ Analysis power with the new spatial resolution MTF chart TG-142 compliant and automated imaging QA for EPIDs and for planar imaging (kV and MV) ■ Automated image analysis avoiding human subjectivity ■ Save time: Complete overview on one screen ■ Instant Confirmation: Clear display of pass / fail TG-142 compliant automated imaging QA for CT and CBCT ■ Workflow ease of use: From image import to the MTF chart ■ Including contrast, contrast to noise...

Open the catalog to page 5
MACHINE QA INDEPENDENT & INTEGRATED - 6

Machines |  Isocenter Check Machines |  Individual QA Tests Doors? Check Check Lasers? Check Check … Check Check Integrated Winston Lutz Test Integrate Any Of Your Specific Tests In myQA Machines Automated checks of the Linac isocenter via the “Winston Lutz” test Your individual QA checks and any other tests are easily integrated with the myQA customizable generic tests functionality ■ Supports most Winston-Lutz phantoms ■ Easily compose any test you need to check and document ■ Supports circular fields (SRS cones) or rectangular fields (MLC) ■ Import of existing data and tests from Excel ■...

Open the catalog to page 6
MACHINE QA INDEPENDENT & INTEGRATED - 7

StarTrack Your High-End Detector for Advanced Machine QA ■ All main tests in one shot: dose, profiles, diagonals energy verification, etc. ■ 453 air-vented ionization chambers with optimized geometry for Machine QA ■ Convenient beam constancy verification in one single shot using specific build-up plates ■ Automatic k(t,p) correction ■ Parallel readout from independent electrometers ■ Instant results and real-time analysis using the Dosimetry plug-in for myQA Machines Left ■ Tabletop or gantry mount (optional) ■ Patented energy verification method MatriXX Your Flexible Detector ■ Detector...

Open the catalog to page 7
MACHINE QA INDEPENDENT & INTEGRATED - 8

Phantoms for Machine QA Cylindrical Phantom Dose Constancy and Isocenter Check ■ erify the mechanical stability of gantry/imager position V (CBCT/ EPID) with a small steel ball insert (Winston-Lutz Test) ■ Measurement of dose constancy in various gantry angles and in rotational beams ■ Adaptors available for most common ion chambers For more information please refer to the IBA white paper ‘Tg-142_Daily Generic Tests’ and ‘Winston-Lutz & Star Shot Test’ Example of a verification image using the IBA Cylindrical Phantom. Disk Phantom Isocenter Verification with Film ■ asy and precise method of...

Open the catalog to page 8
MACHINE QA INDEPENDENT & INTEGRATED - 9

Imaging QA in Radiation Therapy Primus A Test plate for kV planar image QA ■ Easy image QA of your IGRT imaging systems or flat-panel imager (EPID) ■ Verify complete contrast determination, special resolution, s caling discrepancy, uniformity and positioning offsets CT Phantom ■ Innovative 3-part nested PMMA phantom for CTDI measurements ■ Designed to image pediatric and adult head and body ■ According to FDA performance standard for diagnostic X-ray Setup test plate & MagicMaX XR Multi-Detector for QA of kV imager Imaging-Dose Multimeter ■ Fast, simple, and accurate beam analysis and...

Open the catalog to page 9
MACHINE QA INDEPENDENT & INTEGRATED - 10

Independent and Integrated Machine QA Technical Specifications StarTrack Detector Energy Range Electrons: 6–21 MeV. Dose Linearity 0.5 % from 10 cGy to 5 Gy integral dose. 0.5 % from 0.1 Gy/min up to 4 Gy/min dose rate. Sensor Layout Chamber arrays organized along main axes and diagonals, 8 additional chambers for energy constancy check. Spatial Resolution 5mm for horizontal and vertical lines. 7mm for diagonals. Chamber Type Vented plane parallel ionization chambers. Chamber Size Cylindrical, 3 (ø) x 5 (h) mm, sensitive volume 35 mm³. Typical Sensitivity 8 TERA ASICs (each contains 64...

Open the catalog to page 10

All IBA Group catalogs and technical brochures

  1. CYCLONE® IKON

    11 Pages

  2. DOLPHIN

    7 Pages

  3. Sphinx Flyer

    2 Pages

  4. myQA Accept

    6 Pages