Overview: The SafeSet™ Blood Sampling System is designed to prevent blood backflow and air bubbles during blood sampling procedures. It includes specific instructions for setup and use with both blood tube holders and syringes.
Preventing Blood Backflow:
  • Ensure all connections are tight.
  • Keep the SafeSet reservoir in the locked position.
  • Pressurize the IV container according to hospital protocol.
  • Switch vented caps to non-vented caps.
Preventing Air Bubbles:
  • Position the SafeSet reservoir tip upwards with the plunger pulled back to 3 mL during priming.
  • Tap the reservoir and transducer to eliminate air bubbles.
  • Invert and tap blood sampling ports to remove air bubbles.
  • Draw back slowly at no more than 1 mL per second when drawing clearing volume.
  • Reinfuse blood slowly at no more than 1 mL per second.
Using Blood Tube Holder:
  • Turn the one-way stopcock to the 'off' position after obtaining a sample.
  • Remove the blood collection tube before disconnecting from the sampling port.
  • Disconnect the Blood Tube Holder and sampling device together.
  • Turn the stopcock to the 'on' position and flush the line.
Using Syringe:
  • Turn the stopcock to 'off' before removing the syringe.
  • Disconnect the syringe and sampling device together while slightly pulling back on the plunger.
  • Turn the stopcock to 'on' and flush the line.
Setting Up TransPac® IV with SafeSet:
  • Check all connections for tightness.
  • Pull the reservoir plunger back 2-3 mL from the locked position.
  • Spike the fluid source and hang the IV solution bag.
  • Prime the system using gravity pressure.
  • Remove the vented cover, activate the flush device, and fill the line slowly.
  • Place a non-vented cap on the side port.
  • Ensure the reservoir tip is pointed up and prime the reservoir tip.
  • Zero balance the monitor as per manufacturer’s recommendations.
  • Check for air bubbles before connecting to the patient catheter.
Obtaining a Blood Sample:
  • Release the reservoir lock and pull back the plunger slowly to aspirate a clearing volume.
  • Turn the stopcock off to prevent clearing volume contamination.
  • Cleanse the cannula access device or LAD port with disinfectant.
  • Attach the SafeSet blunt cannula to the blood collection device.
  • Aspirate the required blood sample.
  • Turn the stopcock on and reinfuse the blood slowly.
  • Cleanse the sample port with disinfectant.
  • Activate the flush device to clear the line of blood.
See more

Catalog excerpts

SafeSet™-1

Blood Sampling System To prevent blood from backing up in line remember to: • • • • Ensure all connections in line are tightened. Ensure SafeSet reservoir is in the LOCKED position. Pressurize IV container per hospital protocol. Change vented (white) caps to non-vented (yellow) caps. To prevent air bubbles in line remember to: • Ensure SafeSet reservoir tip is pointed up (one-way stopcock on top) with plunger pulled back to 3 mL when priming. • Tap reservoir and transducer to ensure that all air bubbles have been eliminated when priming. • Invert and tap blood sampling ports to remove air bubbles when priming. • raw back SLOWLY no faster than 1 mL per second on reservoir when drawing D clearing volume. • einfuse the patient’s blood slowly, no faster than 1 mL per second, by R pressing the plunger back to the closed and locked position. When using a blood tube holder for blood sampling: • Once blood sample has been obtained, turn one-way stopcock between sampling port and patient to the “off” or perpendicular position. • ext, completely remove blood collection tube from Blood Tube Holder BEFORE N disconnecting from sampling port. • isconnect Blood Tube Holder and blunt sampling device TOGETHER from D sampling port. • urn one-way stopcock between port and patient to the “on” position and flush T line. When using a syringe for blood sampling: • Once the blood sample has been obtained, and prior to removing syringe from sampling port, turn one-way stopcock between sampling port and patient to the “off” or perpendicular position. • Next, disconnect syringe and blunt sampling device TOGETHER (or single syringe if using luer activated device (LAD) kit) from sampling port while slightly pulling back on syringe plunger. • urn one-way stopcock between port and patient to the “on” pos

 Open the catalog to page 1
SafeSet™-2

To set up TransPac® IV with SafeSet: • Check all connections for tightness. • Pull SafeSet reservoir plunger tip back 2-3 mL from the closed and locked position. • Spike the fluid source and insert the IV solution bag into the pressure administration cuff and hang it from the IV pole. (ICU Medical recommends that the system is primed using only gravity pressure.) • Remove the white vented cover from the stopcock next to the transducer, activate the flush device (squeeze or pigtail), and fill the line slowly through the transducer and out the side port (zeroing port) of the stopcock closest to the...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.