The QuickStart Flow Sensor is engineered for high precision in vitro diagnostics, biotechnology, and point-of-care applications. It measures flow rates from 0 to 1000 µL/min with a design that minimizes contamination and bubble entrapment. Available in standalone or manifold mount options, it includes accessories for quick validation.

Internal Volume
  • I2C FS1000F/UART FS1000F (standalone): 32 µL
  • I2C FS1000M/UART FS1000M (manifold): 24 µL

Wetted Materials
  • PEEK Housing
  • Viton O-ring
  • Borosilicate glass

Product Specifications
  • Output Signal: Digital
  • Operating Voltage: 3.3V – 5.0V
  • Communication Bus: I2C/UART
  • Full Scale Flow Rate: 1000 µL/min
  • Sensor Output Limit: 1500 µL/min
  • Accuracy: 5% of measured value or 0.25% of full scale
  • Repeatability Error: 1% of measured value or 0.05% of full scale
  • Temperature Coefficient: 0.1% of measured value/°C or 0.005% of full scale/°C
  • Flow Detection Response Time: 40ms
  • Operating Temperature: +5°C to +50°C
  • Storage Temperature: -30°C to +60°C (dry)
  • Humidity: 0% to 95% RH, non-condensing
  • Max Pressure: 43 psi (3 barg)
  • Burst Pressure: 101 psi (7 barg)
  • Body Materials: PEEK

Additional Notes
The sensor is fully calibrated with DI H2O and temperature compensated. It is currently an unreleased prototype, and specifications may change.
See more

Catalog excerpts

QuickStart™ Flow Sensor-1

Unreleased part. For prototype use only. Specifications subject to change. Wetted Materials > PEEK Housing > Viton O-ring > Borosilicate glass QuickStart Flow Sensor The QuickStart flow sensor is specifically designed for in vitro Diagnostics, Biotechnology and Point-of-Care applications, with high precision measurements within a flow rate range of 0-1000 ^L/min. The flow through design reduces crossover contamination and bubble entrapment. The QuickStart flow sensor is available as a stand-alone or manifold mount and available with accessories to validate the sensor for your application in a short amount of time. Product Specifications Output Signal: Digital Operating Voltage: 3.3V - 5.0V Digital Communication Bus: I2C/UART Full Scale Flow Rate: 1000pl/min Sensor Output Limit: 1500pl/min Accuracy below full scale: 5% of measured value or 0.25% of full scale (whichever error is larger)* Repeatability error from zero to full scale: 1% of measured value or 0.05% of full scale (whichever error is larger) Temperature Coefficient - additional error /°C (from 23 °C): 0.1% of measured value /°C or 0.005% of full scale /°C (whichever error is larger) Flow Detection Response Time: 40ms Operating Temperature: +5°C to +50°C Ambient Storage Temperature: -30°C to +60°C (flow path dry) Operating/Storage Humidity: 0% to 95% RH, non-condensing Recommended Max Pressure: 43 psi (3 barg) Burst Pressure: 101 psi (7 barg) Body materials: PEEK *Fully calibrated with DI H2O, and temperature compensated For ordering, technical support, and contact information please visit www.idex-hs.com ©2017 IDEX Health & Science IDX2554