Modular Patient Monitor: Innocare-T12 Plus
Size and Weight: The monitor measures 318mm x 264mm x 152mm and weighs 4.5 kg. It includes two standard module slots.
Power Supply: Operates on AC 100-240V, 50/60Hz with an input current of 1.1-0.5A, classified under Safety Category I.
Display: Features a 12.1” Color TFT-LCD with a resolution of 800x600 pixels.
Battery: Equipped with a rechargeable Lithium battery (11.1V/4.0AH) providing 210 minutes of operation under normal use.
Recorder (Optional): Utilizes a thermal dot array method with a paper width of 50 mm and a length of 15m, supporting real-time, periodic, and alarm recording.
Alarm System: Includes low, medium, and high-level alarms with auditory and visual indications. Patient physiological alarms are indicated by yellow and red lights, while technical alarms use blue lights.
Input Devices: Standard configuration includes a touch screen and knob, with support for mouse and keyboard input.
System Output & Interfaces: Includes Ethernet (RJ45), Defibrillation Output (BNC), Nurse Call (RJ11), VGA port, two USB1.1 ports, and optional SD memory card and analog output.
Trend & Reviewing: Long trend storage for 168 hours with a minimum resolution of 1 minute, high-resolution trend for 2 hours with a 5-second resolution, and storage for 1000 NIBP measurements.
Environmental Conditions: Operates between 0°C and +40°C with storage temperatures from -20°C to +50°C. Humidity ranges from 15% to 85% during operation and 10% to 93% during storage.
Safety and Standards: Approved by IEC60601-1 and CE marked according to MDD93/42/EEC.
Performance Specifications:
- ECG: Supports 3, 5, and 12-lead inputs with various lead selections and gain settings. Features include a CMRR of 105dB, frequency response modes, and defibrillator proof protection.
- HR: Measures heart rate for adults (10-300 bpm) and pediatric/neonates (10-350 bpm) with a resolution of 1 bpm.
- ST Segment: Measurement range from -2.0mV to 2.0mV with an accuracy of ±0.02mV or ±10%.
- RESP: Uses thoracic impedance with a measurement range of 0-150 rpm and accuracy of ±2 rpm or ±2%.
- NIBP: Automatic oscillometry with adult, child, and neonate measurement ranges. Features include manual, auto, and STAT modes with overpressure protection.
- SpO2: Digital and optional Nellcor and Masimo modules with measurement ranges of 0-100% and varying accuracies.
- TEMP: Measures temperature with a range of 0.0°C to 50.0°C and accuracy of ±0.1°C.
- IBP: Direct invasive pressure measurement with a range of -50 to +350 mmHg and various transducer sites.
- EtCO2: Available in sidestream, mainstream, and microstream methods with measurement ranges and accuracies specified for each.
- Anesthetic Gas: Measures CO2, N2O, and anesthetic gases with specified ranges and accuracies.
- C.O.: Thermal dilution method with a measurement range of 0.1 L/min to 20 L/min.
Standard Configuration: Includes a 12.1” TFT-LCD display, two module slots, touch screen, Ethernet socket, defibrillation output, nurse call socket, VGA port, two USB ports, and a rechargeable battery.
Options: Additional modules for CO2, AG, C.O., IBP, Temp, and SpO2, as well as external display, wireless LAN, and extensive memory card options.
Contact Information: Innomed Medical Developing and Manufacturing Inc., Budapest, Hungary. Phone: (+36-1) 460-9200, Email:
[email protected], Web: www.innomed.hu