Introduction
The document discusses an innovative method for dynamic continuous mapping of the corticospinal tract during neurosurgery, particularly for tumor resection. This method, developed by Raabe, utilizes a Mapping Suction Probe to enhance surgical precision and safety.

Application Fields
The primary application is in neurosurgery, focusing on protecting the corticospinal tract and motor cortex during tumor resection.

Methodology
The method involves dynamic continuous subcortical mapping, which allows for safe tumor resection by minimizing the risk of damage to the corticospinal tract. The Mapping Suction Probe enables maximal tumor removal while maintaining low motor thresholds. The procedure starts with a 10 mA current, adjusting based on motor responses, with a rule of thumb that 1 mm distance equals approximately 1 mA.

Equipment
The Mapping Suction Probe combines a surgical suction tube with a monopolar electrical stimulation probe, allowing for simultaneous suction and mapping. It features a 2mm active tip and is compatible with standard monopolar stimulation parameters.

Benefits
The method simplifies maximal tumor resection without changing instruments, supported by continuous audio feedback to indicate motor responses. This approach enhances surgical precision and reduces the risk of postoperative complications.

Features
The probe provides continuous audio feedback, with low-pitched sounds indicating motor responses and high-pitched sounds confirming no motor response, allowing safe continuation of resection.

Conclusion
The Mapping Suction Probe represents a significant advancement in neurosurgical procedures, offering precision, innovation, and improved patient outcomes through enhanced intraoperative neuromonitoring.
See more

Catalog excerpts

Dynamic continuous Mapping of the Corticospinal Tract-1

Intraoperative Neuromonitoring Functional Neurosurgery Pain Treatment Neurological Diagnostics Dynamic continuous Mapping of the Corticospinal Tract by Raabe >> APPLICATION FIELDS Neurosurgery Protection of the corticospinal tract and motor cortex during tumor resection Innovativ Mapping Method with the Mapping Suction Probe by Raabe

 Open the catalog to page 1
Dynamic continuous Mapping of the Corticospinal Tract-2

> Mapping Method by RAABE >> METHOD Dynamic continuous subcortical mapping simplifies a safe tumor resection. The ability to resect a tumor down to low motor thresholds is a significant refinement of the classic subcortical mapping of the corticospinal tract. The use of the Mapping Suction Probe makes it possible to achieve maximal tumor removal minimizing the possibility of damage to the corticospinal tract. >> Mapping starts with 10 mA (ca. 10 mm distance to the corticospinal tract) >> Current intensity is proportional to the distance of the corticospinal tract (Rule of thumb: 1mm ≈ 1 mA) Audio...

 Open the catalog to page 2
Dynamic continuous Mapping of the Corticospinal Tract-3

> All-in-one surgical suction tube & stimulation probe Especially for the new method of dynamic continuous subcortical mapping, the MAPPING SUCTION PROBE was developed in cooperation with Prof. Dr. Raabe. The combination of a surgical suction tube and a monopolar electrical stimulation probe combines suction during tumor resection with simultaneous continuous dynamic mapping of the corticospinal tract. >> BENEFITS >> Subcortical mapping with synchronous suction >> Simplified maximal tumor resection >> No change of instruments during the procedure >> The method is supported by a continuous audio...

 Open the catalog to page 3
Dynamic continuous Mapping of the Corticospinal Tract-4

Intraoperative Neuromonitoring Functional Neurosurgery Pain Treatment Neurological Diagnostics inomed Medizintechnik GmbH Im Hausgruen 29 79312 Emmendingen (GerMANY) >> Partnership >> Precision >> Innovation Art.-No. 525 650 Mapping Suction Probe 120mm, monopolar with connecting cable and black counter electrode More information and further accessories: D030152/ Subject to change without prior notice © copyright by inomed 12.2015 rev 1.1 Mapping Suction Probe – order information: