Data Sheet MR-guided focused ultrasound Compatible with select GE HealthCare and Siemens Healthineers MR scanners1 Contact your Insightec representative for a current list of compatible scanners.

Open the catalog to page 1
Data Sheet - 2

System Components Helmet System Stereotactic Frame Includes the helmet-shaped focused ultrasound transducer on a mechanical positioning unit. The Helmet System is transfered (from the Storage Transfer Cart) to the MRI table and locked into place using the included table adapter baseplate. Ensures the patient skull is immobilized during the procedure for treatment safety. Front End Unit Contains the programmable power module that shapes and drives the acoustic beam, as well as the water system that is used to cool the patient’s scalp during the procedure. It is installed in the MR room next...

Open the catalog to page 2
Data Sheet - 3

Operational Specifications Focused Ultrasound Transducer Helmet-shaped phased array transducer with 1,024 elements, which are automatically and individually adjusted to refocus the ultrasound beams to a common focal point ■ Ultrasound waves from 8- to 60-seconds duration are delivered to ablate the target area by multiple sonications Four mechanical degrees of freedom to access the treatment target: linear motion along three axes: superiorinferior, right-left, and anterior-posterior axes and angular motion of the pitch axis Focal spot location is controlled by electronic steering Main...

Open the catalog to page 3
Data Sheet - 4

Operator console is powered from a grounded wall outlet, 50/60 Hz, 115/230 VAC. Equipment cabinet is powered by 3ø 480VAC x 50A or 3ø 400/380 VAC x 60A 50/60 Hz from a permanent connection and protected by a dedicated circuit breaker. The workstation tower, host computer, and the 27-inch monitor will receive power from 110VAC outlet. It is recommended to have an emergency generator 110VAC power outlet. It is recommended that the site adds a UPS unit to receive power via the 110VAC emergency outlet. This should be connected to the workstation tower computer. The recommended specification for...

Open the catalog to page 4

All InSightec catalogs and technical brochures