The document provides detailed information about the Integra® Camino® Flex Ventricular Catheter, a medical device designed for continuous cerebrospinal fluid (CSF) drainage and intracranial pressure (ICP) monitoring. It highlights the catheter's features, benefits, and clinical applications.

Specifications and Features
The Camino Flex Ventricular Catheter is designed with separate lumens for simultaneous CSF drainage and ICP monitoring. It supports a flow rate of approximately 300 ml/hr, significantly higher than the normal CSF production rate of 25 ml/hr. The catheter includes multiple drainage holes to reduce the risk of obstructions.

The sensor located at the catheter tip allows for continuous ICP measurement, independent of fluid flow, ensuring accurate monitoring even if the drainage lumen is occluded. The device is MR conditional at 1.5T and 3.0T.

Clinical Advantages
The catheter provides continuous monitoring with high-fidelity waveforms, capturing critical events and offering at least 24 hours of trending data. This helps in identifying changes in patient compliance and avoiding missed events of elevated ICP.

Designed for ease of use, the catheter requires only a one-time, 30-second auto-zeroing process, eliminating the need for frequent transducer maintenance.

Challenges and Solutions
Common intracranial conditions, such as slit or shifted ventricles, can complicate catheter placement. The Camino Flex Ventricular Catheter addresses these challenges by allowing continuous ICP monitoring even when CSF drainage is not possible.

Continuous digital recordings provide more accurate data compared to manual recordings, which often miss significant episodes of high ICP.

Clinical Studies and References
Several studies are referenced, highlighting the catheter's effectiveness in clinical practice. For instance, a study by Zanier et al. demonstrated the advantages of continuous monitoring over manual recording in detecting ICP episodes.

Ordering Information
The document includes catalog numbers and descriptions for ordering the Camino ICP and Temperature Monitor, Camino Flex Ventricular Catheter, and related accessories.

Conclusion
The Camino Flex Ventricular Catheter, supported by advanced monitoring systems, offers a comprehensive solution for neurosurgeons, enhancing patient outcomes through reliable and continuous ICP monitoring and CSF drainage.
See more

Catalog excerpts

Integra® Camino® Flex-1

Integra® Camino® Flex Ventricular Catheter Limit Uncertainty with Continuous CSF Drainage and ICP Monitoring in One System A Tunneled Ventricular Solution to Continuous Drainage and ICP Monitoring

 Open the catalog to page 1
Integra® Camino® Flex-2

Camino® Flex Ventricular Catheter A Tunneled Ventricular Solution to Continuous Drainage and ICP Monitoring Separate Lumen Design Allows for Simultaneous CSF Drainage and ICP Monitoring • The drainage lumen supports a flow rate of approximately 300 ml/hr5 - Normal CSF production rate: approximately 25 ml/hr 6 • Multiple drainage holes help mitigate the risk of obstructions ICP Sensor with Drainage Holes Silicone Membrane Drainage Lumen Stylet Lumen Braided Tensile Member ICP Sensor Control Wires Sensor Located at the Tip of the Catheter Allows Continuous ICP Measurement in the Ventricle Regardless...

 Open the catalog to page 2
Integra® Camino® Flex-3

Integra® Camino®Flex Ventricular Catheter Flex Ventricular Catheter Elevated ICP and Missed Events Can Be Detected With Continuous Monitoring • Continuous ICP waveforms can provide qualitative information regarding brain compliance.1 • In a retrospective study of 32 traumatic brain injury (TBI) patients, the authors concluded that manually recorded end-hour ICP measurements result in the omission of a significant number of important episodes of high ICP, some for a long duration. Manual recording methods, therefore, provide a clinical picture that may not be accurate or informative of the true...

 Open the catalog to page 3
Integra® Camino® Flex-4

Continuous Monitoring... Always Informed Sensor Located at the Tip of the Catheter Allows Continuous ICP Measurement in the Ventricle Regardless of Fluid Flow • Provides uninterrupted ICP monitoring at the source, even when CSF drainage cannot be established due to slit or shifted ventricles, or becomes occluded due to bloody CSF • Continuous ICP monitoring helps avoid missed events of elevated ICP Camino Flex Ventricular Catheter can continuously monitor the ICP, even when CSF drainage lumen becomes occluded ICP of 25 mmHg Ventricular Cavity Flex Ventricular Catheter Camino Flex Ventricular...

 Open the catalog to page 4
Integra® Camino® Flex-5

Sensor at the Tip Allows Flex Ventricular Catheter to Monitor ICP Independent of Patient Position • The Camino Flex Ventricular Catheter sensor monitors ICP directly at the source, providing an accurate reading regardless of patient position • External transducers will produce inaccurate ICP readings if not properly aligned to the patient’s anatomical landmark such as the Foramen of Monro • When not properly leveled, the ICP reading will be off by approximately 0.736 mmHg, for each centimeter of vertical distance between the patient’s anatomical landmark of leveling and an external transducer7...

 Open the catalog to page 5
Integra® Camino® Flex-6

Integra® Camino® Flex Ventricular Catheter The Complete ICU Solution Supported by both the Camino ICP Monitor (CAM02) and the Camino Advanced Monitor fitted with the Camino Flex Adapter, (CAM01, FLEXADR) the Camino Flex Ventricular Catheter provides neurosurgeons access to Integra’s most advanced monitoring and CSF drainage systems for heightened confidence during challenging procedures. The new Camino Flex Ventricular Catheter, coupled with the Integra External Ventricular Drainage System, offers a unified solution to help clinicians achieve the best outcomes for their patients. Camino Flex...