Puffin

Puffin

Puffin

Product catalog summary
Overview: The document provides technical specifications and features of a T-Piece resuscitation device by International Biomedical, designed to meet AAP resuscitation guidelines. It includes an integrated oxygen blender and suction, auxiliary flow meter, and is suitable for worldwide use.
Technical Specifications:
  • Gas Supply Requirements: Input pressure ranges from 39 to 75 psi (275 to 517 kPa) with a minimum flow of 70 L/min.
  • Suction Specifications: Adjustable PIP with a maximum of 45 ± 5 cmH2O at 10 L/min and PEEP range from 5 L/min to 15 L/min with varying cmH2O.
  • Flow Meter: Flow range is 0 to 15 L/min with accuracy varying by flow rate.
  • Air/O2 Blender: Oxygen range is 21 to 100% with an accuracy of ± 5% full scale.
  • Airway Pressure Manometer: Range is -10 to 80 cmH2O with ± 5% accuracy.
  • Operating Environment: Temperature range is 64ºF to 106ºF (18ºC to 41ºC) with humidity from 5 to 95% (non-condensing).
  • Dimensions and Weight: 13.5 in H x 9.25 in W x 8.0 in D (34.3 cm H x 23.5 cm W x 20.3 cm D), weighing 19 lbs (8.6 kg).
Standard Features:
  • Audible alarms for supply pressure.
  • Recommended patient weight up to 22 lbs (10 kg).
  • Integrated oxygen blender and suction with auxiliary flow meter.
  • Auto switch gas source and suction canister ring mount.
Certifications: ISO 13485 certified.
Contact Information: International Biomedical, Austin, Texas, USA, and Uetendorf, Switzerland.
See more

Catalog excerpts

Puffin-1

Infant Resuscitator T-Piece resuscitation Meets AAP resuscitation guidelines Integrated oxygen blender and suction Auxiliary flow meter

 Open the catalog to page 1
Puffin-2

Technical Specications Standard Features Operating Environment Auto switch gas source Two medical gas ow meters Integrated oxygen blender and suction T-piece resuscitator Airway pressure manometer Peak inspiratory pressure (PIP) control Positive end expiratory pressure (PEEP) control Venturi suction Vacuum manometer Operating temperature 64ºF to 106ºF (18ºC to 41ºC) Storage temperature range -13ºF to 140ºF (-25ºC to 60ºC) Operating humidity 5 to 95% (non-condensing) Storage humidity 5 to 95% (non-condensing) Storage pressure 70 to 106 kPa Audible Alarms Supply pressure alarm Recommended Patient...

 Open the catalog to page 2

Archived catalogs

  1. GlucoScout

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.