PAM-200

PAM-200

PAM-200

Product catalog summary
Overview
The PAM-200 Patient Acquisition Module is designed for accurate gating triggers in MRI environments. It integrates ECG acquisition, optical plethysmography for pulse acquisition, and a pneumatic respiration pad for respiratory acquisition.

Accessories
Various accessories are available, including MRI ECG cables, extension cables, electrodes, and respiration components like bellows and straps.

ECG Patient Acquisition
The module uses a 4-lead configuration with specific vectors to minimize magnetic gradient effects. It provides two simultaneous orthogonal vectors for accurate ECG readings.

Respiratory Acquisition
A pneumatic technique is used, involving a foam-filled bellows that responds to chest expansion. The system transmits respiratory signals digitally and can drive an analog LED indicator.

Peripheral Pulse Acquisition (Pleth)
The module uses an optical-based method to acquire peripheral pulse signals via a finger clip, similar to pulse oximetry. Signals are transmitted digitally and can also drive an analog LED indicator.

Technical Specifications
  • Operating Conditions: MRI Conditional with specific static field, gradient, and SAR limits.
  • Communications: Unidirectional fiber optic communication with a maximum distance of 45 meters.
  • ECG Performance: Detection amplitude range of 0.3mV to 4.5mV, with a digitization rate of 10KHz.
  • Respiratory Performance: Pressure range of 0 to 1 psi, with a digitization rate of 1KHz.
  • Peripheral Pulse Performance: Bandwidth of 10Hz, with a digitization rate of 1KHz.

Mechanical and Power Requirements
The module is housed in a RF-tight aluminum casing and requires specific voltage and current inputs.

Environmental Conditions
Operating temperature ranges from 5 to 40 °C, with specific humidity and altitude conditions.

Regulatory Compliance
The unit complies with UL60601-1, CAN/CSA, CE, and WEEE regulations.

Warranty
The PAM-200 comes with a one-year warranty covering parts and labor.
See more

Catalog excerpts

PAM-200-1

PAM-200 Accurate Gating Trigger for the MRI Environment Precision Engineering The PAM Module is an embedded instrumentation pack which provides an accurate gating trigger for use in an MRI environment. The PAM-200 is a complete solution consisting of ECG Acquisition, an Optical Plethysmograph for pulse acquisition, and a pneumatic respiration pad for respiration acquisition. ECG Patient Acquisition The PAM utilizes a 4 lead configuration in a Right Arm, Left Arm, Right Leg and Left leg setup. The Right Leg is the reference lead with Right Arm, Left Arm and Left Leg as the active leads. The PAM provides for two simultaneous vectors. The first vector is Right Arm and Left Leg. The second vector is Right Leg and Left Leg. The two vectors are orthogonal to each other and will result in one vector less effected by magnetic gradients. Accessories PAM System Module by a flexible plastic tube, there is no electrical connection. The bellows has a stiff backing which aids in the compression action of the chest movement. Positive pressure (exhalation) will result in a positive going signal. The respiratory signal is transmitted in the digital stream to the console computer, but the PAM has provision for an analog signal to drive a LED bar indicator which could be located on the gantry or magnet housing. The PAM utilizes a pneumatic technique which responds to the expansion of the chest during respiration. A foam filled bellows is attached to the patient with a strap, and the air pressure in the bellows changes as the chest moves and compresses the bellows. The internal foam provides the spring action to return the bellows to the pre-compressed state. The bellows is attached to the ©2015 Ivy Biomedical Systems Inc. All Rights Reserved. P/N: 3230-00-16 Rev. 02 Name Integrated MRI ECG Cable/ Lead Set, 6’ ECG Extension Cable, 1 Meter (External Magnet Use only) MRI Electrodes NeuPrep Gel Name Bellows Strap, 42” Hose to Bellows Peripheral Pulse Acquisition (Pleth) Respiratory Acquisition Patient Acquisition Module The PAM has provision to perform cardiac gating by way of an optical based peripheral pulse, which is acquired on the patient finger. The optical emitter and detector are located within the PAM housing and the light energy is coupled to the patient through fiber optic cable. The patient connection is through a finger clip similar to a pulse oximetry sensor. The peripheral pulse signal is transmitted in the digital stream to the console computer, but the PAM has provision for an analog signal to drive a LED bar indicator which could be located on the gantry or magnet housing. Name Part No. Optical Pleth Cable, 2.8 Meters 3168-00-01 Please contact an IVY sales representative at 1-800-247-4614, to learn more about the PAM200 Patient Acquisition Module.

 Open the catalog to page 1
PAM-200-2

Patient Acquisition Module Accurate Gating Trigger for the MRI Environment Technical Specifications MRI Gating Module Operating Conditions Respiratory Performance Respiratory Patient Connection The Ivy Biomedical Systems PAM-200 MRI Gating Module, ECG Cable, Peripheral Pulse Cable and Pneumatic Respiration Pillow have been tested and verified to be MRI Conditional under the following operating conditions: Maximum Static Field: 1.5T Maximum Spatial Gradient: 1580 g/cm Maximum Peak Gradient Strength: 33 mT/m Maximum flux change / time (db/dt): 113.4 T/sec Maximum Wholebody SAR (15min): 3.88 W/kg...

 Open the catalog to page 2