This section presents the comparative analysis between the colloidal gold detection reagent and the reference RT-PCR detection reagent based on samples collected and tested within seven days of symptom onset.
These results indicate a high level of agreement between the colloidal gold detection reagent and the RT-PCR method, demonstrating the reagent's reliability for early detection.
The document references components and kit specifications but does not provide detailed information. Further details on the components, their specifications, and quantities are necessary for a complete understanding.
Consistency evaluation results: Based on the comparison results of sampling and testing within seven days of onset, our colloidal gold detection reagent is compared with the contrast RT-PCR detection reagent. The positive coincidence rate is 97.46% (92.19%~99.34%), and the negative coincidence rate is 99.45% (96.53%). ~99.97%), the total coincidence rate is 98.46%.