Room and Storage Requirements
Before installation, ensure the unit is not placed near windows or in rooms that cannot be darkened. The operational temperature should be between 60° – 80°F (15° – 27°C) with a relative humidity of 35% to 65%. Barometric pressure should range from 11 – 16 psi (76 – 110 kPa). For storage, maintain a temperature between 14° – 140°F (-10° – 60°C) and a relative humidity of 0 – 85%.
Product Description
The iDESIGN® Refractive Studio System uses a Hartmann-Shack sensor to measure the eye's wavefront, determining refractive errors and irregularities. It also measures corneal topography, pupil size, and keratometry, aiding in LASIK treatment calculations.
Specifications for Site Preparation/Installation
The optical head dimensions are 20” x 18” x 27” (50.8 cm x 45.7 cm x 68.6 cm) and weighs 80 lbs (36 kg). The table dimensions are 32.0 in x 26.1 in x 22.1 in – 33.3 in (81.3 cm x 66.3 cm x 56 cm – 84.5 cm) and weighs 102.4 lbs (46.4 kg). Electrical ratings are 100-120/200-240 AC Voltage, 50/60 Hz, 750 VA.
Optical Specifications
The system measures sphere and cylinder in 0.01 D increments, with a spherical equivalent range of -16 to +12 D and a cylinder range of 8 D. Pupil measurements range from 2.0 to 9.5 mm with 0.1 mm resolution. It includes an integrated corneal topographer and pupilometer.
System Specifications
The system is indicated for wavefront-guided PRK and LASIK in patients with specific refractive errors and conditions. It is suitable for patients 18 years or older with stable refraction.
Hardware Components
The system includes a power supply, computer, networking device, and monitor. The CPU is housed inside the system, with a USB port on the right-hand side of the computer case.
Important Safety Information
U.S. Federal Law restricts this device to sale, distribution, and use by or on the order of a licensed eye care practitioner. Refer to the Operator’s Manual for contraindications, warnings, and precautions.
See more

Catalog excerpts

iDESIGN™-1

iDESIGN® Refractive Studio Specifications For Site Preparation/Installation Product Description The iDESIGN® Refractive Studio System measures the wavefront of the eye within a defined range using the HartmannShack sensor. The sensor evaluates the deflection of rays emanating from a small beam of light projected onto the retina. The measurements determine regular (sphero-cylindrical) refractive errors and irregularities (aberrations) that cause reduced visual function. The iDESIGN® Refractive Studio also measures and displays corneal topography, pupil size, and keratometry. Wavefrontlaser assisted in situ keratomileusis (LASIK) treatments can be calculated using measurements obtained from the iDESIGN® Refractive Studio. Treatment calculations for wavefront-guided LASIK include full gradient topography for propagating the wavefront and compensating for the cosine effect (peripheral loss of laser energy due to corneal curvature). Recommended Room Requirements • Do not place the unit near windows or in a room that cannot be sufficiently darkened to allow the patient’s pupils to dilate naturally • Ambient operational temperature range: 60° – 80°F (15° – 27°C) • Humidity: Relative humidity no less than 35% and no greater than 65% (non-condensing) • Barometric pressure range: 11 – 16 psi (76 – 110 kPa) Storing Requirements Before Installation When storing the system before installation, adhere to the following storage requirements: • Storage temperature must be between 14° – 140°F (-10° – 60°C) at a relative humidity no less than 0 – 85% (non-condensing) CAUTION: U.S. Federal Law restricts this device to sale, distribution, and use by or on the order of a physician or other licensed eye care practitioner. ATTENTION: Reference the Operator’s Manual for a complete listing of Contraindications, Warnings and Precautions as well as other Indicati

 Open the catalog to page 1
iDESIGN™-2

System Specifications • Optical Head o Physical dimensions: (L, W, H) 20”, 18”, 27”, (50.8 cm, 45.7 cm, 68.6 cm), including base o Weight: 80 lbs (36 kg) o Enclosure construction: Aluminum and plastic o Class 1 Laser Product: 840 nm, 100uW max • Table o Physical dimensions: (L, W, H) 32.0 in, 26.1 in, 22.1 in – 33.3 in (min-max), (81.3 cm, 66.3 cm, 56 cm – 84.5 cm) o Weight: 102.4 lbs (46.4 kg) o Electrical ratings are 100-120/200-240 AC Voltage, 50/60 Hz, 750 VA Hardware Components The iDESIGN® Refractive Studio hardware components include power supply, computer, networking device, and monitor....

 Open the catalog to page 2

All Johnson & Johnson Vision catalogs and technical brochures

  1. iFS™

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.