Overview of Oxygen Use in BUSTER ICU Cage
Each BUSTER ICU Cage is equipped with an oxygen accessories kit (J-1057Q) that includes various components such as Venturi devices, a nebuliser, a T piece, a cap, and oxygen tubing. These components are essential for administering oxygen therapy and nebulisation.

Contents of the Oxygen Accessories Kit
  • 6 Venturi devices (green, red, yellow, white, blue, and a blank)
  • 1 Nebuliser
  • 1 T piece
  • 1 Cap
  • 1 2.1m oxygen tubing

Instructions for Use
For Nebulisation:
  1. Attach the blank Venturi to one end of the oxygen tubing and connect it to the oxygen supply.
  2. Unscrew the nebuliser, add medication, and reassemble it vertically.
  3. Connect the oxygen tubing to the nebuliser base and attach the nebuliser to the T piece.
  4. Connect the T piece to the BUSTER ICU Cage and close the other arm with the cap.
  5. Turn on the oxygen at a sufficient flow rate to nebulise the medication.
For Oxygen Therapy:
  1. Attach the appropriate coloured Venturi to the oxygen tubing.
  2. Connect the Venturi to the T piece and attach it to the ICU Cage.
  3. Close the other arm of the T piece with the cap.
  4. Turn on the oxygen to the appropriate flow rate, with a minimum of 2 litres per minute.

Rationale for Controlled Oxygen Therapy
Excess oxygen can be harmful, potentially causing toxicity to the lungs. It is advised to avoid inspired oxygen levels greater than 60% and to reduce supplemented oxygen as soon as possible.

Venturi Principles of Operation
The Venturi device functions similarly to a carburetor, mixing oxygen with air to deliver a consistent oxygen/air mixture to the ICU Cage. The mixture's enrichment level is determined by the Venturi's hole size.

Choice of Venturi
Different medical conditions require varying levels of oxygen enrichment:
  • 24-28% for conditions like fractured ribs or diaphragmatic hernia.
  • 35-40% for conditions like chronic bronchitis or mild lung contusion.
  • 60% for severe conditions like pneumonia or congestive cardiac failure.

Calculating Total Gas Mixture Flow Rates
The BUSTER ICU Cage volumes are approximately 50 litres (small), 120 litres (medium), and 460 litres (large). The flow rates should initially be high to fill the cage with the appropriate oxygen concentration. Calculate the minute tidal volume and aim to change at least twice this volume per minute. Use the provided chart to determine the oxygen flow rate.

Example Calculation
Using a 24% Venturi at a minimum O2 flow rate of 2 litres/min generates a total flow rate of 52 litres/min. Increasing the O2 flow to 4 litres/min results in a total flow rate of 104 litres/min, calculated by adding the air entrained (100 litres/min) to the oxygen flow (4 litres/min).
See more

Catalog excerpts

Buster ICU Cage Oxygen-1

Each BUSTER ICU Cage comes with an oxygen accessories kit (J-1057Q). 6 X Venturi - 1 of each colour green, red yellow, white, blue and a blank (no holes) Attach the blank Venturi (no holes) to one end of the oxygen tubing. This then attaches to the oxygen supply. Unscrew the top half of the nebuliser and add the chosen medication. Keep the nebuliser vertical whilst screwing back together. Attach the end of the oxygen tubing to the base of the nebuliser. Attach the top of the nebuliser to the vertical arm of the T piece Attach one of the horizontal arms of the T piece to the BUSTER ICU Cage. Close the other horizontal arm with the cap provided. Turn on the oxygen at sufficient flow rate to nebulise the chosen medication. Attach the appropriate coloured Venturi to the other end of the oxygen tubing. Attach the Venturi to the vertical arm of the T piece. Push one of the horizontal arms of the T piece into the port on the BUSTER ICU Cage. Close the other arm of the T piece with the cap provided. Turn on the oxygen to the appropriate flow rate. The minimum oxygen flow rate is 2 litres per minute. Jorgensen Laboratories • 1450 N. Van Buren Avenue, Loveland, Colorado 80538 • [email protected] • www.JorVet.com

 Open the catalog to page 1
Buster ICU Cage Oxygen-2

Rationale for Controlled Oxygen Therapy Excess oxygen can be harmful. Like many drugs, oxygen can be associated with toxicity and excess oxygen can be harmful to lungs (pulmonary endothelium). The use of inspired oxygen greater than 60% should be avoided if possible. Decreasing the amount of supplemented oxygen at the earliest possible time should be routine practice. The Venturi works in a similar fashion to a carbureter in a car engine. Oxygen is delivered into the Venturi barrel. As the oxygen passes through the barrel it draws in a calculated amount of air which mixes with the oxygen. The...

 Open the catalog to page 2
Buster ICU Cage Oxygen-3

Calculating Total Gas Mixture Flow Rates The approximate volume of each BUSTER ICU Cage is as follows: Small 50 litres, Medium 120 litres and Large 460 litres. Select the Venturi for the condition using Diagram 1. Run the flow rates high initially to fill the ICU cage with oxygen enriched air at the appropriate concentration. Calculate minute tidal volume (1 Oml/kg x breaths per minute). Aim to change at least twice the minute tidal volume of air/oxygen mixture per minute. Calculate the oxygen flow rate using the chart below.

 Open the catalog to page 3
Buster ICU Cage Oxygen-4

At the minimum 02 flow rate of 2 litres/min, this Venturi will generate a total flow rate of 52 litres/min. To calculate the total flow rate from the same Venturi if 02 flow is increased to say 4 litres/min, simply look for 4 litres/min on the vertical axis of the graph and follow the horizontal line unit it intersects the 24% line. At this point of the intersection look to see what air flow rate on the horizontal axis is appropriate. In this case it is 100 litres/min. Add the 02 flow rate for the final mixture flow rate. 100 (air entrained) + 4 (oxygen flow) = 104 litres/min. Similar calculations...

 Open the catalog to page 4