Hemiarthroplasty Surgical technique-1

The Furlong Hemiarthroplasty System OPERATIVE TECHNIQUE

 Open the catalog to page 1
Hemiarthroplasty Surgical technique-2

Furlong® Cemented Hemiarthroplasty Furlong® Cemented Instrumentation 7

 Open the catalog to page 2
Hemiarthroplasty Surgical technique-3

The Furlong® Hemiarthroplasty System • Cemented stem with bipolar, modular bipolar or physiological head options provides clinical flexibility to match surgeon preference • Fully modular system designed to allow the surgeon to select the implants which best meet the patients clinical needs • Easy to use effective instrumentation designed to enhance clinical outcome • JRI Orthopaedics Limited was established in 1969 by an orthopaedic surgeon, Mr Ronald Furlong, when he introduced the Muller cemented hip to the UK market. • Developers of the world’s first H-A.C. coated hip replacement prosthesis...

 Open the catalog to page 3
Hemiarthroplasty Surgical technique-4

Furlong® Cemented Hemiarthroplasty Clinical Evidence 72 Implants had been inserted into 71 different patients by 35 different Surgeons of all grades. No dislocations in any of the 71 patients who had bipolar hips inserted during the study period2 Comparable with results of total hip replacement (THR) but without the risk of dislocation2 Results are significantly better than would be expected with conventional Hemiarthroplasty in this group of patients2 Eleven of the sixteen patients without comorbidities who were able to walk 1 mile before fracture were able to do so at review. Mean follow up...

 Open the catalog to page 4
Hemiarthroplasty Surgical technique-5

Physiological Head The JRI Physiological Head is manufactured from HNSS and available in sizes 39mm-54mm (1mm increments), 56mm and 58mm. HNSS increases biocompatibility and provides greater fatigue strength and corrosion resistance over conventional medical grade stainless steel. Modular Bipolar Head The Modular Bipolar Head is designed to provide surgeons with improved options for management of offset and leg length. The inner articulation is achieved with the choice of 3 JRI Orthopaedics Furlong 28mm CoCr heads. NB: If using Modular Bipolar Heads please refer to the specific surgical technique....

 Open the catalog to page 5
Hemiarthroplasty Surgical technique-6

Furlong® Cemented Operative Technique 1 Correct positioning of the patient on the operating table is very important. The hip joint is exposed using a preferred surgical approach for hemiarthroplasty. The fractured femoral head is removed using the Femoral Head Extractor (10-01-27). A light tap may help engagement and purchase into the femoral head. If required, trimming of the fracture line can be performed. A trial stem can also be used to help identify where the resection should be and a line made using a diathermy probe or skin marker. The femoral head size can then be estimated using the...

 Open the catalog to page 6
Hemiarthroplasty Surgical technique-7

The smallest XXS rasp (50-00-14) is used to prepare the proximal femur. The large Tommy Bar (10-25-28) should be used to control version. Move the rasp anteriorly and posteriorly to check anterior / posterior stability. When movement only occurs between the handle and the rasp, not between the rasp and bone, the rasp is correctly sized. Care should be taken to lateralise the rasps as they are inserted. At this point, if required, the appropriate trial stem can be carefully inserted using the Stem Impactor (50-99-34). The trial stem must match the last size rasp used for a secure femoral fit The...

 Open the catalog to page 7
Hemiarthroplasty Surgical technique-8

The preferred cementing technique is now used. Modern techniques recommend the use of a cement plug, lavage, drying of the femoral canal and retrograde filling with a cement gun. Making sure the tibia is vertical, the definitive femoral stem should be inserted in neutral alignment, to ensure a continuous circumferential cement mantle, using the Stem Impactor and inserted to the depth determined by the trial prosthesis. The depth of the stem should be determined by the height of the centre of rotation of the femoral head on the contralateral side. If the same size rasp, trial and definitive stem...

 Open the catalog to page 8
Hemiarthroplasty Surgical technique-9

Intramedullary Rasp Reamer Size (mm) Implant Size Cement Plug (mm) Extra Extra Small or 10 -11 10mm Diameter 11mm Diameter 12mm Diameter 13mm Diameter 14mm Diameter 15mm Diameter 16mm Diameter 17mm Diameter 18mm Diameter 19mm Diameter 20mm Diameter

 Open the catalog to page 9
Hemiarthroplasty Surgical technique-10

CEMENTED HEMIARTHROPLASTY TRAY RASPING & TRIALING 81-00-58 Trial Stems (Threaded Spigot) Extra Extra Small Extra Small Extra Extra Small Extra Small Small Medium Large HEMIARTHROPLASTY TRAY REAMING Femoral Head Extractor HEMIARTHROPLASTY TRAY (BOTTOM LAYER)* GENERAL INSTRUMENTS 81-00-55 DESCRIPTION PRODUCT NO. Trial Head Introducer Femoral Head Impactor Trial Stem Extractor Box Chisel Stem Impactor Fin Cutter Impactor Head *The bottom layer can be used with Bipolar heads (General Instument Set upper layer) and with Physiological heads (General Instrument Set upper layer).

 Open the catalog to page 10
Hemiarthroplasty Surgical technique-12

MODULAR BIPOLAR TRIALS DESCRIPTION Femoral Head Gauge ø40mm Modular Bipolar Trial ø41mm Modular Bipolar Trial ø42mm Modular Bipolar Trial ø43mm Modular Bipolar Trial ø44mm Modular Bipolar Trial ø45mm Modular Bipolar Trial ø46mm Modular Bipolar Trial ø47mm Modular Bipolar Trial ø48mm Modular Bipolar Trial ø49mm Modular Bipolar Trial ø50mm Modular Bipolar Trial ø51mm Modular Bipolar Trial ø52mm Modular Bipolar Trial ø53mm Modular Bipolar Trial ø54mm Modular Bipolar Trial ø56mm Modular Bipolar Trial ø58mm Modular Bipolar Trial ø60mm Modular Bipolar Trial Bipolar Assembly Jig Main Body Bipolar Assembly...

 Open the catalog to page 12
Hemiarthroplasty Surgical technique-13

CEMENTED FEMORAL STEMS STANDARD Extra Extra Small Extra Small Small Medium Large

 Open the catalog to page 13
Hemiarthroplasty Surgical technique-15

MODULAR BIPOLAR HEADS (O/D) JRI CoCr FEMORAL HEADS 12/14 TAPER MODULAR BIPOLAR HEAD OPTIONS Head-neck length / Head size (mm) NB: Refer to Modular Bipolar Optech for assembly instructions.

 Open the catalog to page 15
Hemiarthroplasty Surgical technique-16

• Nurse Training • Consignment stock checks: • Factory Tours • Instruments • Educational Courses • PACS Digital X-Ray templates • In service support • 2D/3D Pre-operative planning 1The Furlong® Hemiarthroplasty hip system and the JRI Modular Bipolar Head is made in Britain. 2S Dixon & G Bannister. Cemented Bipolar Hemiarthoplasty for Displaced Intracapsular Fracture in the Mobile Active Elderly Patient. Int. J Care Injured 2004; 35:152-156. 3P Chandran et al. Mid Term Results of Furlong LOL Uncemented Hip Hemiarthroplasty for Fractures of the Femoral Neck. Acta Orthop. Belg., 2006; 72:426-433....

 Open the catalog to page 16

All JRI Orthopaedics catalogs and technical brochures

  1. Furlong

    20  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.