SDMA

SDMA

SDMA

Product catalog summary
Introduction
Symmetric Dimethylarginine (SDMA) is a crucial biomarker for assessing renal function in both veterinary and human medicine. This document evaluates the clinical consistency and reliability of SDMA assays from JZ Biotech and brand A, focusing on their agreement and correlation to support clinical decision-making and laboratory harmonization.

Objective
The primary goal is to compare the analytical performance of JZ Biotech and brand A SDMA assays by assessing result correlation and evaluating bias and agreement using statistical methods.

Materials
Data was collected from cases at the Cooperative Animal Hospital between December 2023 and September 2024.

Results
Descriptive Statistics
JZ Biotech's mean SDMA was 26.7 µg/dL (SD = 22.1), while brand A's was 24.9 µg/dL (SD = 19.4).
Correlation Analysis
Pearson’s r was 0.93 (p < 0.001), indicating a strong linear relationship. The regression equation was brand A = 0.85 × JZ Biotech + 1.5 (R² = 0.86).
Bland-Altman Analysis
The mean bias was +1.8 µg/dL, with 95% limits of agreement from -9.2 µg/dL to +12.8 µg/dL. Higher variability was noted in samples with SDMA > 60 µg/dL.

Conclusion
High Concordance
The assays showed a strong correlation (r = 0.93), meeting criteria for interchangeability.
Regression Equation
The slope (0.85) and intercept (+1.5) indicated comparability across most ranges.
Low Dispersion
Despite a slightly higher standard deviation for JZ Biotech, assay reproducibility was good.
Outliers
Outliers were rare and attributed to operational errors or special samples, not affecting overall conclusions.

Prepared by Dr. Wu Sichen, Lead Clinical Analyst, and reviewed by Prof. Yang Liheng, Director of Quality Assurance.
See more

Catalog excerpts

SDMA-1

Clinical Consistency and Reliability Validation of JZ Biotech SDMA Assay KeyWord: SDMA, JZ Biotech, Method Comparison, Clinical Agreement, High Consistency, Low Systematic Bias, Reliability 1. Introduction 4.1 Descriptive Statistics Symmetric Dimethylarginine (SDMA) is a biomarker widely used in veterinary and human medicine to assess renal function. Accurate measurement of SDMA is critical for early detection of kidney disease. JZ Biotech and brand A are two prominent diagnostic companies offering SDMA testing solutions. This report compares SDMA results obtained from both systems to evaluate their agreement and correlation, which is essential for Pearson’s r: 0.93 (p < 0.001), indicating a strong linear relationship (Figure 1). Regression Equation: brand A = 0.85 x JZ Biotech +1.5 clinical decision-making and harmonizing laboratory (R2 = 0.86). To compare the analytical performance of JZ Biotech and brand A SDMA assays by: Assessing the correlation between results. Evaluating bias and agreement using statistical methods. Data from cases collected at the Cooperative Animal Hospital from December 2023 to September 2024. Mean Bias: +1.8 pg/dL (JZ Biotech > brand A). 95% Limits of Agreement: -9.2 pg/dL to +12.8

 Open the catalog to page 1
SDMA-2

Key Insight: Higher variability observed in samples or special samples and did not affect the overall with SDMA >60 µg/dL (Figure 2) . Figure 2: Bland-Altman Plot High concordance: strong correlation with industry Pearson's correlation coefficient (r = 0.93): the data Lead Clinical Analyst demonstrated a highly linear correlation (p < 0.001) between JZ Biotech and brand A's SDMA results, which met the criteria for interchangeability both statistically and clinically (Figure 1). Prof. Yang Liheng The regression equation (brand A = 0.85 × JZ Biotech + 1.5): the slope was close to 1 (0.85) and the...

 Open the catalog to page 2

All JZ Biotech catalogs and technical brochures

  1. RB02

    2  Pages

  2. JZ-FIAS7

    25  Pages

  3. JZVET-OT01

    2  Pages

  4. JZVET-WL01

    3  Pages

  5. JZVET-060C

    2  Pages

  6. JZVet-EX01

    2  Pages

  7. JZVet-CH02

    2  Pages

  8. JZVET-EX02

    2  Pages

  9. JZVET-CX01

    2  Pages

  10. JZVETDK

    1  Page

  11. test kit

    2  Pages

  12. JZVet-EX

    1  Page

  13. JZVet-CH

    1  Page

  14. FHV AG

    2  Pages

  15. FPV AG

    2  Pages

  16. FCoV AB

    2  Pages

  17. FIV AB

    2  Pages

  18. FHV AB

    2  Pages

  19. FCV AB

    2  Pages

  20. FPV AB

    2  Pages

  21. FPV/FCV/FHV

    2  Pages

  22. FSAA

    2  Pages

  23. FPL

    2  Pages

  24. SDMA

    2  Pages

  25. JZ-FIA

    2  Pages

  26. PAG TEST

    1  Page

Archived catalogs

  1. JZVET-100B

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.