Calibration Bath Low Temperature - 1

Low temperature calibration baths - metrology performance: • Uniformity and stability in mK range • Calibrate sensors of any dimension • Fluid ensures perfect heat transfer • Up to 500 mm immersion depth • Unique vertical flow design • State of the art solution • Three bath sizes Kambic d.o.o. Metliska cesta 16; 8333 Semic ; Slovenia - EU www.kambicmetrology.com info@kamb

Open the catalog to page 1
Calibration Bath Low Temperature - 2

Over temperature shut off device Vertical fixture – one included as standard Special double cylinder design with unique vertical flow across entire working volume 7L, 22L, 50L working chamber with immersion depth up to 500 mm Touch screen based controller with user friendly interface, history graph advance settings options RS-232, USB or Ethernet communication ports Insulation cover included simple to drill and fix any size sensor Full 4 side maintenance access Extensive high performance heat insulation All exterior and interior in stainless steel AISI 304 SB With rounded corners Mobile...

Open the catalog to page 2
Calibration Bath Low Temperature - 3

Controller functions: LCD capacitive touch display Actual measured fluid temperature in bath Display power switch Start temperature control System elements operation indicator Access to info screen Stop temperature control Access to lock / unlock screen Access to history graph Access to factory / advance user settings Bath’s temperature stability graph at - 40 °C: Kambic d.o.o. Metliska cesta 16; 8333 Semic ; Slovenia - EU www.kambicmetrology.com info@kambicmetrology.

Open the catalog to page 3
Calibration Bath Low Temperature - 4

External dimensions (WxHxD) in mm Internal dimensions (diameter x immersion depth) in mm Volume (L) Temperature range (°C) Temperature display resolution (°C) Temperature set resolution (°C) Temperature stability (°C) Temperature uniformity (°C) Heating rate Cooling rate Temperature Control Noise level (dBA) Cooling Power supply Wattage (W) Interface Weight (kg) 0.001 ± 0.004 @ - 40 °C ± 0.005 @ 20 °C ± 0.002 @ 100 °C < ± 0.007 @ - 40 °C ~ 37 min (+ 20 °C …+ 100 °C) ~ 64 min (+ 20 °C …- 40 °C) PID < 58 Compressor Cooling System 230 V 50/60 HZ (± 10 %) 2000 RS 232 (USB or Ethernet as option)...

Open the catalog to page 4
Calibration Bath Low Temperature - 5

- Vertical fixture - Adjustable multiple probe fixture - External fluid level for LIG - Foldable shelve - Fume hood - Evaluation report Ordering information and accessories: Description Calibration bath OB-7/2 LT Calibration bath OB-22/2 LT Calibration bath OB-50/2 LT USB interface Ethernet interface Evaluation report – performed by Kambic Evaluation report (Accredited ISO/IEC 17025:2017) OB-Tool (Monitoring and data collection software) External fluid level for LIG Vertical fixture Adjustable multiple probe fixture LIG (liquid in glass thermometer) fixture Local fume hood with fan Foldable...

Open the catalog to page 5
Calibration Bath Low Temperature - 6

Kambic d.o.o. Metliska cesta 16; 8333 Semic ; Slovenia - EU www.kambicmetrology.com info@kambicmetrology.com