CALIBRATION BATH - 1

CALIBRATION BATH Black Body Cavity ready OB-15/2 BB (+ 40 °C …+ 250 °C) OB-15/2 BBLT (- 20 °C …+ 130 °C) Special calibration baths - metrology performance: • Perfect solution for calibrating IR thermometers and other thermographic equipment • Various shapes Black Body Cavities available • Uniformity and stability in mK range • Fluid ensures perfect heat transfer • Unique vertical flow design • Two temperature ranges • State of the art solution Kambic d.o.o. Metliska cesta 16; 8333 Semic ; Slovenia - EU www.kambicmetrology.com info@

Open the catalog to page 1
CALIBRATION BATH - 2

Over temperature shut off device” Vertical fixture – one included as standard Touch screen based controller with user friendly interface, history graph advance settings options RS-232, USB or Ethernet communication ports Special double cylinder design with unique vertical flow across entire working volume 15L working chamber with immersion depth 270 mm Black Body Cavity – available in various types Full 4 side maintenance access Extensive high performance heat insulation Air cooled conventional compressor based refrigeration system (BBLT version) All exterior in stainless steel AISI 304 SB...

Open the catalog to page 2
CALIBRATION BATH - 3

Controller functions: LCD capacitive touch display Actual measured fluid temperature in bath Display Power switch Start temperature control System elements operation indicator Stop temperature control Access to lock / unlock screen Access to info screen Access to history graph Access to factory / advance user settings Bath’s temperature stability graph at + 100 °C: Kambic d.o.o. Metliska cesta 16; 8333 Semic ; Slovenia - EU www.kambicmetrology.com info@kambicmetrology.

Open the catalog to page 3
CALIBRATION BATH - 4

External dimensions (WxHxD) in mm Internal dimensions (diameter x immersion depth) in mm Volume (L) Temperature range (°C) Temperature display resolution (°C) Temperature set resolution (°C) Temperature stability (°C) Temperature uniformity (°C) Heating rate Cooling rate Temperature Control Noise level (dBA) Cooling Power supply Wattage (W) Interface Weight (kg) ~ 15 - 20 … + 130 0.001 0.001 ± 0.006 @ - 20 °C ± 0.005 @ 0 °C ± 0.003 @ 100 °C ± 0.007 @ -20 °C ~ 55 min (+ 20 °C …+ 100 °C) ~ 35 min (+ 20 °C …- 20 °C) PID < 58 Compressor Cooling System 230 V 50/60 HZ (± 10 %) 2500 RS 232 (USB or...

Open the catalog to page 4
CALIBRATION BATH - 5

Accessories: - LIG thermometer holder - Vertical fixture - Adjustable multiple probe fixture - External fluid level for LIG - Foldable shelve - Fume hood - Evaluation report Ordering information and accessories: Description Calibration bath OB-15/2 BB Calibration bath OB-15/2 BBLT Black Body cavity ASTM standard Black Body cavity JIS T standard Black Body cavity CEN EN standard USB interface Ethernet interface Evaluation report – performed by Kambic Evaluation report (accredited ISO/IEC 17025:2017) OB-Tool (Monitoring and data collection software) External fluid level for LIG Vertical...

Open the catalog to page 5
CALIBRATION BATH - 6

Kambic d.o.o. Metliska cesta 16; 8333 Semic ; Slovenia - EU www.kambicmetrology.com info@kambicmetrology.com