video corpo

Dexcam 4HD
34Pages

{{requestButtons}}

Catalog excerpts

Dexcam 4HD - 1

Intra-oral Camera User Manual

Open the catalog to page 1
Dexcam 4HD - 3

The DEXIS DEXcam4 HD intra-oral camera provides live high definition video display, as well as freeze and capture capabilities, for dental imaging. The camera is not meant to be used for diagnostic purposes. There are no contraindications for use of the camera. • Proprietary, seven-element glass optical lens produce exceptional image quality • Uses an HD Sony CMOS sensor with resolution of 1.3 Mega-pixels • Two button capture control capability directly on the handpiece • One-touch focus adjustment • Detachable cable for easy sharing • Robust aluminum construction. Intended Use This...

Open the catalog to page 3
Dexcam 4HD - 4

DEXcam4 HD EN.book Page 2 Monday, January 16, 2017 2:12 PM Proficiency with your New Device Please become acquainted with your new DEXcam4 HD intra-oral camera, it is your tool. As with all new clinical tools, it is necessary to invest a certain amount of time for practice in order to become proficient. Learn to use it well and it will become an effective aid. Please note that this manual assumes that new users possess basic computer skills and an understanding of the Windows® operating system. Absent this experience, we highly recommend that you obtain these skills through a computer...

Open the catalog to page 4
Dexcam 4HD - 5

This device is an intra-oral camera system that provides the dentist and the patient with intra-oral and extra-oral images. This assists the dentist in describing the dental procedure being performed, as well as allowing the dentist to show the results to the patient. Both Sides

Open the catalog to page 5
Dexcam 4HD - 6

DEXcam4 HD Installation Minimum system requirements: • Microsoft Windows 7 or higher • DirectX 9.0b or higher • Pentium 4 at 2 GHz or higher • 1 GB RAM • Monitor with minimum 1024 x 768 resolution capability • Graphics / display card capable of minimum 1024 x 768 resolution at 24-bit color and 32 MB video RAM • CD-ROM drive The installation procedure has two parts: First - install DEXcapture software (refer to the DEXcapture Software manual for the software installation). Second - install the intra-oral camera. Important! Be sure to follow the instructions to install the software before...

Open the catalog to page 6
Dexcam 4HD - 7

DEXcam4 HD EN.book Page 5 Monday, January 16, 2017 2:12 PM Install the camera holster in a place that is convenient and does not exceed the maximum length of the USB cable, which is 8.9 ft. (2.7 m). The camera holster should also be positioned so that the cable from the camera does not create a tripping hazard. To install the holster, clean the mounting surface, remove the backing from the double sided-tape on the holster, then firmly press the holster against the mounting surface. After approximately two hours, the holster can be used to store the camera. Intra-oral Camera Sheath Refer to...

Open the catalog to page 7
Dexcam 4HD - 8

DEXcam4 HD EN.book Page 6 Monday, January 16, 2017 2:12 PM To use the camera, remove it from the holster. It turns on and begins to send video images. 1. Preset the camera to view either an intra-oral or extra-oral image. CAUTION Be sure to use a new sheath when capturing intra-oral images. 3 2. Position the camera. 3. Adjust the camera focus. • Press + to zoom in, which moves the LED indicator to the right. Press to zoom out, which moves the LED indicator to the left. 4. Press and release the two Capture buttons to freeze the image on your viewing screen. CAUTION Ensure that the camera is...

Open the catalog to page 8
Dexcam 4HD - 9

• The camera is not waterproof. Do not immerse in water and avoid situations where water or other foreign matter could get on or inside the product. Failure to do so may damage the product. • If the camera is wet, or if your hands are wet, do not use the product. • Avoid subjecting the camera to vibrations or shocks. When the product is not in use, make sure it is placed in the camera holster. • Examine the camera housing for visible damage. Be careful not to use the camera if it is damaged. Doing so may cause injury. • Avoid using the camera in a dusty environment. • Remove the protective...

Open the catalog to page 9
Dexcam 4HD - 10

DEXcam4 HD EN.book Page 8 Monday, January 16, 2017 2:12 PM Label Location The following label is located on the USB cable of the camera. Product Symbols Indicates the product should be used only once. The symbol is found on the packaging for the sheaths. This symbol on the product and/or accompanying documents means that used electrical and electronic products should not be mixed with general household waste. Please refer to the written instructions of this manual. The ETL listed mark indicates that Intertek has certified the product described herein under control number 4005847 to be in...

Open the catalog to page 10
Dexcam 4HD - 11

DEXcam4 HD EN.book Page 9 Monday, January 16, 2017 2:12 PM Service and Proper Disposal The service life of the camera is five years. There are no serviceable parts or maintenance on the device. Consult the device warranty for additional information. Properly dispose of the camera at the end of its useful life. Do not mix with general household waste. Specifications USB Compatibility USB 2.0 or later Electrical Ratings 5VDC, 500mA Requirements for safety approvals This device must be connected to a personal computer that meets the standard specified in IEC 60950, revision 2003. Ambient...

Open the catalog to page 11
Dexcam 4HD - 12

The intraoral camera emits electromagnetic energy at levels compliant to international standards; however, interference between the camera and other devices is possible if located in close proximity. The customer or the user of the camera can help prevent electromagnetic interference by locating sensitive devices away from the camera. This camera, classified as Medical Electrical Equipment, requires special precautions regarding EMC and must be installed and put into service according to the EMC information provided in the accompanying product documentation. Portable and mobile RF...

Open the catalog to page 12
Dexcam 4HD - 14

DEXcam4 HD EN.book Page 12 Monday, January 16, 2017 2:12 PM Guidance and manufacturer’s declaration - electromagnetic immunity for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING Guidance and manufacturer’s declaration – electromagnetic immunity This camera is intended for use in the electromagnetic environment specified below. The customer or the user of this camera should assure that it is used in such an environment. Immunity test Compliance level Electromagnetic environment - guidance Portable and mobile RF communications equipment should be used no closer to any part of this camera,...

Open the catalog to page 14

All KaVo Kerr Group catalogs and technical brochures

  1. HELIOS LED

    5 Pages

  2. UPHOLSTERY

    4 Pages

  3. Sprirt 3300

    4 Pages

  4. NX3

    2 Pages

  5. NTI Diamonds

    1 Pages

  6. Demi Ultra

    3 Pages

  7. KaVo OP3D Pro

    20 Pages

  8. OP2D

    8 Pages

  9. OP200 D

    9 Pages

  10. TempBond

    2 Pages

  11. X-Ray-Holders

    144 Pages

  12. PROGLOSS™

    2 Pages

  13. X-Ray Aprons

    2 Pages

  14. Burs catalogue

    28 Pages

  15. Demi Plus

    4 Pages