1. Catalogs
  2. Lepu Medical
  3. Vesscide™ Cutting Balloon System

Vesscide™ Cutting Balloon System

Vesscide™ Cutting Balloon System

Vesscide™ Cutting Balloon System

Product catalog summary
Overview: The document provides detailed technical specifications and usage guidelines for the Vesscide Cutting Balloon System, a medical device used in PCI (Percutaneous Coronary Intervention) pretreatment. It is manufactured by Lepu Medical Technology in Beijing, China.
Specifications: The cutting balloon system is available in various diameters (2.0mm to 4.0mm) and lengths (5mm, 10mm, 15mm), with each size having a specific number of blades. The system requires a minimum guiding catheter of 1.8mm/6F and a guiding wire of 0.36mm/0.014inch. The effective length of the catheter is 1350mm, with markers located at both ends of the balloon.
Technical Parameters: The document includes a compliance table showing the balloon diameter at different pressures (4 to 12 atm). The nominal pressure and rated burst pressure are specified, ensuring safe operation within these limits.
Performance and Usage: The cutting balloon system is designed to improve crossability, optimize retraction, and enhance pushability. It is particularly effective for treating stent restenosis, small vessel lesions, ostia lesions, bifurcation lesions, and calcified lesions. The balloon should be slowly dilated to 6 atm or nominal pressure and depressurized gradually to achieve optimal results.
Product Features: The system features sharp, anti-friction blades with a working height of 0.13mm to prevent vessel perforation. The semi-compliant balloon is coated with HydrosurfTM to avoid slipping and ensure precise lesion anchoring. Inlaid platinum markers reduce the risk of irregular dissection and acute occlusion.
Indications: The cutting balloon system is indicated for patients with atherosclerotic plaques in coronary arteries requiring cutting. It benefits from low-pressure dilation and features a segmented microsurgical blade design to reduce plaque migration and elastic retraction.
References: The document cites several studies and consensus papers from Chinese specialists on the treatment of calcified coronary artery lesions and the clinical application of drug-coated balloons and bioabsorbable stents.
See more

Catalog excerpts

Vesscide™ Cutting Balloon System-1

[ Structure indication and technical parameters ] Hub Side hole Inner tube Balloon 1. Adapted guiding catheter: Minimum guiding catheter 1.8mm/6F 2. Adapted guiding wire:0.36mm/0.014inch Cutting Balloon System 3. Effective length of the catheter:1350mm 4. Marker location:Both ends of the balloon Essential weapon for PCI pretreatment Rated Burst pressure Lepu Medical Technology (Beijing) Co.,LTD. No.37 Chaoqian Road,Changping District, Beijing, China, 102200 Tel:+86-10-80120666 Email:[email protected] Web:en.lepumedical.com

 Open the catalog to page 1
Vesscide™ Cutting Balloon System-2

[ Product Features ] Segmented Microsurgical blades Design . Sharp and anti-friction blades, balance cutting force and flexibility . Blades' Working height is only 0.13mm, which could avoid blood vessel perforation . Pre-made longitudinal neat blades, which could reduce the risk of intimal disordered tearing Semi-compliant balloon The cutting balloon system is indicated to be applied for patients with atherosclerotic plaques in the coronary arteries that need to be cut. . Improve crossability . Optimize retraction Inlaid platinum marker . Reduce balloon external diameter . Increase flexibility...

 Open the catalog to page 2

All Lepu Medical catalogs and technical brochures

  1. Nexor 32

    2  Pages

  2. Fluo-1800

    28  Pages

  3. Lepgen-96

    1  Page

  4. Occluder Device

    20  Pages

  5. ELISA 200

    75  Pages

  6. Nexor 96

    2  Pages