LKPv and LGPv Laboratory Appliances Overview

Alarm, Memory, and Safety Functions:
The appliances are equipped with high and low temperature alarms, door opening alarms, and power failure alarms, all with audio and visual signals. These alarms have adjustable delays to prevent false alerts. Internal system alarms include high condenser temperature, low evaporator temperature, and probe failure alarms, providing early warnings for maintenance.

Internal Memory Functions:
The appliances can store details of the last 30 alarms and maintain a temperature profile at 4-minute intervals, with a capacity of up to 2800 logs.

Special Safety Functions:
Features include a solid-state relay for reliability and double fan cooling to ensure continued operation even if one fan fails.

Temperature and Alarm Documentation:
An integrated temperature data logger records temperature values, which can be transferred to a PC using an optional infrared key and software.

External Monitoring Connections:
The RS 485 interface allows for external temperature and alarm monitoring, with optional signal converters available. The system can send email alerts in case of alarms.

External Alarm Contact Connection:
A volt-free contact can be integrated into external warning systems, such as GSM alarm dialers.

Product Temperature Monitoring Connection:
An optional NTC product temperature probe can be connected to monitor product temperatures, with adjustable alarm limits and delays.

Temperature Consistency and Distribution Testing:
The document outlines the testing procedures for Liebherr laboratory cabinets according to the EN 60068-3 standard, focusing on temperature consistency and distribution. The laboratory is certified to ISO 17025 standards.

Test Procedure Description:
The EN 60068-3 test conditions include an ambient temperature of +25°C and 60% relative humidity. The refrigerator and freezer set temperatures are +5°C and -20°C, respectively. Measurements are taken using nine PT 100 air temperature probes.

Temperature Gradient and Fluctuation:
The gradient measures temperature distribution, with a lower gradient indicating better distribution. Fluctuation measures the variation in temperature at a specific point over time. Results show gradients ranging from 1.8 K to 7.5 K and fluctuations from 1.0 K to 6.7 K.

External Monitoring and Networking:
The appliances support external temperature and alarm monitoring via an RS 485 interface, with optional conversion to RS 232 for PC connectivity. Software is provided for data visualization and alarm notifications, including email alerts.

Product Temperature Monitoring:
An interface for an optional NTC product temperature probe is available, allowing the display of either internal air temperature or product temperature.

Specifications and Features:
Pharmacy refrigerators comply with DIN 58345 standards, ensuring safety and reliability for pharmaceutical storage. Key features include NTC product temperature probes, internal alarm memory, min/max temperature memory, a +2°C safety thermostat, and access ports for additional probes.

Temperature Evaluation:
The MK(U)v models were evaluated according to EN 60068-3 standards using 9 PT 100 temperature probes. Results indicate temperature gradients and fluctuations, with specific values provided for models MKv 3910, MKv 3913, MKUv 1610, and MKUv 1613.

Laboratory Appliances:
The LKexv series, including models 5400, 3600, 2600, and 1800, feature analogue controls and spark-free interiors, suitable for storing explosive or highly flammable materials. These appliances conform to ATEX directive 2014/34/EU, with a temperature class rating of T6.

Compliance and Limitations:
While the MKv and MKUv models comply with DIN 58345 for pharmacy refrigerators, other Liebherr refrigerators and freezers do not meet DIN 58371 and DIN 58375 standards for blood and plasma storage.
See more

Catalog excerpts

Laboratory technical sales manual-1

Technisches Laboratory Verkaufshandbuch technical Laborgeräte sales manual

 Open the catalog to page 1
Laboratory technical sales manual-2

Contents LKPv and LGPv laboratory appliances with Profi electronic controller Alarm, memory and safety functions 03 Temperature, door opening and power failure alarms 03 Internal system alarms 03 Internal memory functions 04 Special safety functions 04 Temperature and alarm documentation 05 External temperature and alarm monitoring connections 05 External alarm contact connection 07 Product temperature monitoring connection 08 Safety feature table 09 Temperature consistency and distribution according to EN 60068-3 10 Description of the EN 60068-3 test procedure...

 Open the catalog to page 2
Laboratory technical sales manual-3

LKPv and LGPv laboratory appliances with Profi electronic controller Alarm, memory and safety functions Liebherr MediLine refrigerators and freezers with Profi electronic controller help to protect your inventory. These cabinets offer a variety of alarm, memory and safety features designed to greatly reduce the risk of loss of contents. Temperature, door opening and power failure alarms High and low temperature alarms All LKPv and LGPv models have integrated high and low internal air temperature alarms with both audio and visual signals. The factory settings for the alarm limits are + 3 /– 2 K in...

 Open the catalog to page 3
Laboratory technical sales manual-4

LKPv and LGPv laboratory appliances with Profi electronic controller Internal memory functions Internal alarm memory This function stores to memory details of the last 30 alarms: time and date of the start and the end of the alarm condition, the type of the alarm condition and the maximum and minimum internal air temperatures measured during this alarm condition period are registered. Internal temperature memory This function stores to the memory the internal temperature profile at 4-minute intervals. 2800 temperature logs are stored to memory, which corresponds to approximately a 7-day logging...

 Open the catalog to page 4
Laboratory technical sales manual-5

LKPv and LGPv laboratory appliances with Profi electronic controller Temperature and alarm documentation Liebherr MediLine refrigerators and freezers with Profi electronic controller help to protect your inventory. The appliances offer a variety of features and options that are specifically designed to allow remote monitoring of interior temperatures, product temperatures and alarm events, in order to provide comprehensive documentation of the storage conditions. External temperature and alarm monitoring connections Infrared interface and optional infrared key All LKPv and LGPv models have an integrated...

 Open the catalog to page 5
Laboratory technical sales manual-6

LKPv and LGPv laboratory appliances with Profi electronic controller RS 485 interface and optional signal converter All LKPv and LGPv models have an RS 485 interface. If an RS 485 monitoring system is not already in place, a converter kit is available as an optional extra to convert the RS 485 signal to an RS 232 signal and allow connection to a PC or laptop. The RS 485 bus connection is made using standard shielded data cables, e.g. type LiYCY 2 x 0,14 mm2. The maximum possible length of the data lines depends on the quality of the components used. The maximum range is 500 m. A maximum of 20...

 Open the catalog to page 6
Laboratory technical sales manual-7

LKPv and LGPv laboratory appliances with Profi electronic controller External alarm contact connection Volt-free contact All LKPv and LGPv models have an integrated volt-free contact. The relay contact can be wired into a warning system in the building which would advise e.g. security personnel or a janitor outside of working hours. Warning systems with audio or visual alarms are possible. The following example shows a circuit with a visual alarm. Either normal operation or alarm condition or both can be indicated. Either AC or DC circuits are possible, max. AC 230 V / 5 A. Example: visual alarm...

 Open the catalog to page 7
Laboratory technical sales manual-8

LKPv and LGPv laboratory appliances with Profi electronic controller Product temperature monitoring connection Interface for optional NTC product temperature probe All LKPv and LGPv models have an integrated interface for connecting an optional available NTC product temperature probe to the Profi electronic controller. The controller can be set to display either the internal air temperature, as registered by the air probe or the temperature registered by the product probe or both. Also the temperature alarm limits as well as the temperature alarm delay for the product temperature probe can be adjusted...

 Open the catalog to page 8
Laboratory technical sales manual-9

LKPv and LGPv laboratory appliances with Profi electronic controller with Profi electronic controller

 Open the catalog to page 9
Laboratory technical sales manual-10

LKPv and LGPv laboratory appliances with Profi electronic controller Temperature consistency and distribution according to EN 60068-3 The Liebherr laboratory cabinets are all tested in-house to the EN 60068-3 test procedure. The measurement results for the appliance setups can be found in the results section. Please note the measurements are for the specific appliances tested and are for reference only. Liebherr’s development facility is certified to ISO 17025 (general requirements for the competence of testing and calibration laboratories). Description of the EN 60068-3 test procedure The general...

 Open the catalog to page 10
Laboratory technical sales manual-11

LKPv and LGPv laboratory appliances with Profi electronic controller Cabinets are tested empty over a period of 24 hrs including defrost cycles. Test criteria for characterising the temperature consistency and distribution 1) Gradient or homogeneity: “The gradient HѲ is the maximum difference, obtained in an achieved environment, between the mean values of measurements Ѳmj increased by their expanded uncertainty U mj, within the working space during the measurement time.” Gradient: “Distribution of temperatures within the chamber” HѲ = max (Ѳmj + Umj ) - min (Ѳmj - Umj ) max = (mean value warmest...

 Open the catalog to page 11
Laboratory technical sales manual-12

LKPv and LGPv laboratory appliances with Profi electronic controller 2) Fluctuation: “The characterisation of the temperature fluctuation within the working space enables to determine the variation in maximum temperature at a measurement point of the working space during the measurement time.” Fluctuation of the point SѲ j: The difference between the maximum and minimum values measured at point j during the duration of measurements Maximum fluctuation, SѲ M : Maximum value of all fluctuation values SѲ j during the duration of measurements Example: Fluctuation Sensor with greatest difference

 Open the catalog to page 12