GMP RF F MTR

GMP RF F MTR

Product catalog summary
Overview: The document provides technical specifications and features of the Matryoshka Real Time Wireless Temperature Data Logger with a flexible sensor, designed for precise temperature monitoring in compliance with FDA 21 CFR Part 11.
Specifications:
  • Sensor Type: 4-wire PT-100, flexible length of 430 mm, customizable up to 5000 mm, with a sensor size of 1.5 x 8 mm and a cross-section of 1.77 mm².
  • Operating Range: Body: -90°C to +140°C; Sensor: -200°C to +200°C.
  • Accuracy and Resolution: Accuracy of ±0.1°C from -90°C to 140°C, with a resolution of 0.01°C.
  • Memory: Capacity of 32,000 data points, non-volatile memory type.
  • Sampling Rate: Adjustable from 1 second to 24 hours.
  • Transmission: Simultaneous transmission for up to 20 loggers, unlimited sequential transmission.
  • Internal Clock Drift: 4 seconds per 24 hours at 23°C.
Mechanical Features:
  • Material: Peek 1000, FDA compliant.
  • Enclosure: IP-68 rated.
  • Dimensions and Weight: Diameter of 46 mm, height of 74.5 mm, and weight of 100g.
Power and Battery:
  • Power Source: 2 x Lithium batteries, user-replaceable, classified under UN3091.
  • Battery Lifetime: Up to 6 months depending on usage.
Wireless and Connectivity:
  • Antenna: Internal.
  • Wireless Range: 121 meters (400 feet).
  • Radio Frequency: 2.4 GHz ISM band.
  • Access Point: USB to computer, supports unlimited loggers.
Calibration and Certification:
  • Factory Calibrations: Traceable NIST/COFRAC with ITS-90 coefficients stored internally.
  • User Calibrations: Closed loop calibration using XpertLog® software.
  • Certifications: Health Safety, EMC, RF, compliant with EN 60950-1 / 2006, EN 60601-1-2 / 2007, EN 300440-2 v1.3.1 (2009-03).
Additional Features:
  • Compatible with 70 ml vial-shaped logger, suitable for 150, 250, and 500 ml fake vials.
  • Real-time temperature data transmission to XpertLog Software.
  • Network ready with multiple user support, centralized server database, and active directory authentication.
  • On-site calibration capability.
Order Reference: GMP-RF-F-MTR70ml
Standard Calibration Points: -80°C, 0°C, 140°C
For more details, visit www.lives-international.com. Document version LVSC.3.2.0 - January 2022.
See more

Catalog excerpts

GMP RF F MTR-1

GMP RF F MTR Matryoshka Real Time Wireless Temperature Data Logger with Flexible Sensor -90°C to +140°C + + + + + + Sensor type - flexible - customized - Sensor size - cross section Operating Range - body - Sensor Accuracy Resolution Memory capacity Memory type Sampling rate Transmission Internal clock drift Mechanical - material - enclosure - weight - diameter - height Power Battery lifetime Antenna Wireless range Radio frequency Access point Calibrations Certification 70 ml vial shaped logger, compatible with 150, 250 and 500 ml fake vials Collects and transmits real time temperature to XpertLog Software ITS-90 calibration coefficients are stored in the internal memory Complies with FDA 21 CFR Part 11 Network ready: multiple users, centralized, server database, active directory authentication Calibration can be performed on site 4 wires PT-100 430 mm (17”) up to 5000 mm (195”) 1.5 x 8 mm (0.06 x 0.315”) 1.77 mm² (0.0027 inch²) -90°C to +140°C -200°C to +200°C ±0.1°C from -90°C to 140°C 0.01°C 32,000 data points non volatile 1 sec to 24 hours simultaneus: maximum 20 loggers sequential: unlimited 4 sec / 24 h @ 23°C Peek 1000 - FDA compliant IP-68 100g (3.53 oz) 46 mm (1.81") 74.5 mm (2.93") 2 x Lithium batteries, 0.35 g/each Lithium content classified UN3091 under IATA DGR section 3.9.2.6 compliant with ICAO/IATA packing instructions 970, user replaceable Up to 6 months, based on usage Internal 121 m (400 feet) 2.4 Ghz ISM band USB to computer Programs and reads unlimited number of loggers Factory Calibrations: traceable NIST/COFRAC - ITS-90 coefficients stored in the internal memory User Calibrations: closed loop calibration using XpertLog® software Health Safety: EN 60950-1 / 2006 EMC: EN 60601-1-2 / 2007 RF: EN 300440-2 v1.3.1 (2009-03) Visit www.lives-international.com for more details. Order Reference GMP-RF-F-MTR70ml Standard Calibration Points • -80°C • 0°C • 140°C

 Open the catalog to page 1

All Lives International catalogs and technical brochures

  1. Access Point

    1  Page

  2. GMP RF Bio

    1  Page

  3. GMP RF TP F

    1  Page

  4. GMP RF TH F

    1  Page

  5. GMP RF TP

    1  Page

  6. GMP RF F 10R

    1  Page

  7. GMP RF 5F HT

    1  Page

  8. GMP RF 5F

    1  Page

  9. Accessories

    2  Pages

  10. XperTrack

    1  Page

  11. GMP RF F LP

    1  Page

  12. GMP RF F

    1  Page

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.