Grossing station SAFETY FIRST Elevating laboratory efficiency with unrivalled safety, enhanced ergonomics and versatility
Open the catalog to page 1Grossing station
Open the catalog to page 2WHERE INNOVATION AND FLEXIBILITY MEET SAFETY AT EVERY TURN Designed with adaptability in mind, safeGROSS 150 pro boasts unique features tailored to meet diverse needs of laboratories of any dimension and to ensure seamless integration into any workspace. From configurable settings to ergonomic design elements, every aspect is meticulously crafted to enhance safety, efficiency, operator comfort and precision in specimen handling. What truly sets our grossing stations apart is an advanced 3-way ventilation system which includes a groundbreaking patented feature (air curtain), designed to safeguard...
Open the catalog to page 3Grossing station SAFETY FIRST Unmatched safety Dual draft ventilation In the dual draft configuration, backdraft and downdraft dynamics are combined to pull formalin fumes and other dangerous vapors and aerosols both down and away from personnel, from the rear and the bottom. Airflow is drawn through the perforated work surface panels and BACKDRAFT VENTILATION a backdraft grid located vertically behind the work surface. For enhanced operator safety, the backdraft system grid is fixed to reach a maximum height of 130 cm so that harmful fumes cannot reach the operator’s respiratory tract. DOWNDRAFT...
Open the catalog to page 4Patented air curtain protection The air curtain, driven by independent fans enabled by the operator, prevents the dispersion of harmful substances, enhancing workplace safety. Integrated into the workstation, this unique solution serves dual functions: • creating a barrier of clean air to contain formalin fumes and other dangerous substances • directing carcinogenic elements downwards for extraction. This unique system safeguards both operator and surrounding environment from harmful agents and protects processed materials from contamination. safeGROSS 150 pro Quiet and gentle ventilation does...
Open the catalog to page 5Grossing station Perfect ergonomic designfor ideal operator comfort and efficiency 1 Large functional worktop with a depth of 70 cm depth (total workstation depth of 90 cm) 2 “Optimal area” workspace design and work surface equipped with a wide range of standard accessories for easier specimen preparation (see separate section for accessories) 4 High-precision adjustable elevation to accommodate all pathology lab personnel and to ensure operator-friendly use 5 Perforated removable rear panel - space saving and organized storage of accessories 6 Rounded front corners (10 mm) - safer and more comfortable...
Open the catalog to page 6Ergonomics tailored to individual needs Worktop height and optimal area The worktop height is carefully crafted for comfortable use whether sitting or standing. These features keep the “optimal area” centrally accessible, and the 150-cm width provides ease of use, allowing the operator to have everything needed at hand during a procedure. By “optimal area,” we refer to the section of the station easily reachable by the operator, mitigating the risk of repetitive strain injuries caused by unnatural movements throughout the workday. Elevation and stability Two electro-mechanical actuators, featuring...
Open the catalog to page 7Grossing station Efficient and user-friendlyprocess control, and traceability A PLC with fully decontaminable 7” touchscreen monitor is integrated in the grossing station for full process control. The digital control panel is separated from the image system to ensure the privacy of sensitive data and the remote control option makes it possible to operate also from a distance, enhancing accessibility. Main functions • LED lights control • Elevation control • Selection of preferred position • Ventilation control • Alarm display • Accessories management • Remote assistance • Software updates
Open the catalog to page 8ENDOVISION integrated imaging system safeGROSS 150 pro is equipped with an advanced imaging system designed for touchscreen displays for the digital management of operational processes and for the creation of reports in compliance with DICOM and HL7 standards. This imaging system is ideal for capturing and managing video, voice, and image content from the high-resolution camera, enabling seamless archiving for future review. With the aid of the pedals, touchscreen, and voice commands, the operator is enabled to carry out the following tasks: • Take pictures (FHD camera)* • Video recording (FHD...
Open the catalog to page 9Grossing station SAFETY FIRST Advanced modularity and flexibility Selecting the appropriate setup is a breeze with our array of pre-configured grossing station models, meticulously designed to fit a laboratory’s specific needs. safeGROSS pro is available in the standard configuration with a width of 150 cm (safeGROSS 150 pro). Other widths are available on request.
Open the catalog to page 10safeGROSS 150 pro, our premium grossing station with an elevating configuration, is equipped with the following essential features, equipment and systems: • Made entirely of stainless steel AISI 304 (without visible joints and screws) • Dual draft ventilation (backdraft+downdraft) • Air curtain ventilation system • ENDOVISION imaging system • Suction conveyor connecting the 3 suction ducts into one • Formalin collection system: 10- or 20-l collection tank. Warning lights and sound indicators warn when tank is nearly full • Formalin dispensing system: Integrated in the grossing station: pump,...
Open the catalog to page 11safeGROSS is Logibiotech brand name for DFB900 by Hiplaas LOGIBIOTECH srl Registered Office Via Fontana, 16 - 20122 Milano, Italy Operating Headquarters Via Emilia SS9 443 – 29010 Alseno (PC), Italy www.logibiotech.com e-mail: [email protected] PEC: [email protected] VAT: IT10961670964
Open the catalog to page 124 Pages