OxyProbe®

OxyProbe®

OxyProbe®

Product catalog summary
Overview
The OxyProbe® Dissolved Oxygen Sensors by LTH Electronics are designed for microbial fermentation and cell culture bioprocess applications in fermentation vessels and bioreactors. They are available with various process connections to fit different port types.

Unique Features
OxyProbe® sensors use mechanical seals with o-rings, eliminating adhesives that could fail. This design enhances durability and simplifies cathode replacement.

Sensor Models
  • D400 & D405: Designed for 25 mm side ports, these sensors fit directly into the vessel, reducing contamination risks. They feature a modular design for easy maintenance.
  • D470 & D475: Fit into a 25 mm side port with a 1.5” sanitary flange, eliminating housing needs and reducing contamination risks. They require a port angle of at least 15° above horizontal.
  • D210 & D215: Fit into a 1.5” sanitary pipe tee, not intended for standard 25 mm ports but compatible with NovAseptic ports. They share interior design and parts with other 25 mm OxyProbes®.

Specifications
  • Material: 316L Stainless steel
  • Finish: RA32 with electro polish on wetted parts
  • Measurement: Amperometric – Polargraphic
  • Output: Approx. 67nA at 25°C
  • Temperature Range: 0-80°C
  • Sterilisation: Max 135°C
  • Stability: Better than 2% per week at constant conditions
  • Max Pressure: 4 bar, 58 PSI

Choosing a Sensor
Consider the process connection type, sensor extension length, connector preference, and cable requirements.

Cable Assemblies
Various shielded cable assemblies are available to connect sensors to transmitters and controllers, featuring D4 twist-lock coaxial plugs.

Spares and Accessories
  • Single Cartridge Kit: Includes membrane cartridge and o-rings for one sensor.
  • Four Cartridge Kit: Includes components for four sensors.

Order Codes
Detailed order codes are provided for different sensor models and cable assemblies.

Compliance
These products comply with current European Directives.
See more

Catalog excerpts

OxyProbe®-1

OxyProbe® OxyProbe® Dissolved Oxygen Sensors LTH Electronics can offer a series of Polargraphic Dissolved Oxygen sensors that are ideally suited for use in microbial fermentation and cell culture bioprocess applications inside fermentation vessels and bioreactors. D215 The Oxyprobe® Dissolved Oxygen sensors are available with a choice of process connections to suit 25mm threaded ports, side ports with 1.5” sanitary flange connection and a 1.5” sanitary flange

 Open the catalog to page 1
OxyProbe®-2

What makes OxyProbe® sensors unique? The internal seals of an Oxyprobe® are all mechanical seals using o-rings. There are no adhesives to leak or fail in process, making this a more rugged and reliable sensor. Replacing the cathode is also much easier and faster since there is no sealant to clean from the threads. No sealant also makes the job of replacing the cathode much faster and cleaner. D400 & D405 OxyProbe® Dissolved Oxygen Sensors The OxyProbe® dissolved oxygen sensor offers the most rugged design for 25 mm side ports. It fits directly into the vessel, eliminating the need for a housing,...

 Open the catalog to page 2
OxyProbe®-3

Specifications 316L Stainless steel Serialised for traceability RA32 finish with electro polish on wetted parts Measurement principle: Amperometric - Polargraphic Output signal in air: Approx. 67nA at 25°C Temperature range: 0-80°C Sterilisation: Maximum 135°C Stability: Better than 2 % of readout per week at constant pressure and temperature. (30°C.) Maximum pressure: 4 bar, 58 PSI. How to choose a sensor: (1) Confirm the type of process connection - standard 25 mm port, 1.5" sanitary flange port or pipe tee. (2) Determine how far the sensor should extend into the vessel. For most process media,...

 Open the catalog to page 3
OxyProbe®-4

LTH Electronics Ltd, Chaul End Lane, Luton, Bedfordshire. LU4 8EZ England Telephone: +44 (0)1582 593693 Fax: +44 (0)1582 598036 email: [email protected] web: www.lth.co.uk LTH Electronics Ltd reserves the right to make changes to this data sheet or the product without notice, as port of our policy of oontinued developments and improvements. All care has been taken to ensure the aocuracy of information contained n this data sheet. However we cannot aocept responsibility for any errors or damages resulting from errors or inaccuracies of information con fared herein.