1. Catalogs
  2. LumiraDx
  3. CRP Test Specifications

CRP Test Specifications

CRP Test Specifications

CRP Test Specifications

Product catalog summary
Intended Use
The LumiraDx CRP Test is designed for use with the LumiraDx Platform by healthcare professionals to quantitatively determine C-Reactive Protein (CRP) levels in human whole blood and plasma samples. It aids in detecting infections, tissue injuries, and inflammatory disorders in patients aged 2 years and older.
Test Description
CRP is a key acute phase protein used as a biomarker for various diseases. The LumiraDx CRP Test is a microfluidic immunoassay that provides rapid results from a fingerstick in 4 minutes. It is traceable to certified global reference materials.
Built-in Quality Controls
The LumiraDx Platform includes quality control checks for instrument and test strip functionality, including electrical and mechanical components, test strip positioning, expiry, and performance monitoring. The CRP Test includes an Onboard Quality Control assay.
CRP Quality Controls
LumiraDx offers Multi Quality Controls at two levels for comprehensive assessment of the instrument and test strips.
CRP Test Specifications
  • Precision: A study showed precision across different CRP concentrations with total precision ranging from 5.6% to 7.4% CV.
  • Method Comparison: A comparison with the Siemens Dimension® Xpand® Plus system showed a slope of 1.00 and an intercept of -0.48, with a correlation coefficient of 0.99.
  • Matrix Equivalency: Equivalency was demonstrated across capillary fingerstick blood, whole blood, and plasma samples.
CRP Test Strip Specifications
  • Storage Temperature: 2–30°C (36–86°F)
  • Operating Temperature: 15–30°C (59–86°F)
  • Measuring Range: 5.0 - 250.0 mg/L
  • Hook Effect: No effect observed up to 1044 mg/L
  • Minimum Sample Size: 20 µL
  • Sample Types: Direct fingerstick, Venous Whole Blood, Plasma
  • Time to Result: 4 minutes
Additional Information
For more details, visit lumiradx.com or contact LumiraDx Customer Services. Note that the product is not available in all regions and is not currently available in the USA.
See more

Catalog excerpts

CRP Test Specifications-1

CRP Test Specifications For in vitro diagnostic use. The LumiraDx CRP Test is intended for use with the LumiraDx Platform. It is an automated in vitro diagnostic test for near-patient testing, for use by healthcare professionals, for the quantitative determination of C-Reactive Protein in human whole blood (capillary finger stick and venous) and plasma samples. The measurement of CRP provides information for the detection and evaluation of infection, tissue injury, inflammation disorders, and associated disease. It is intended for use in patients 2 years of age or older. Test description CRP is one of the most prominent acute phase proteins (APP), a protein whose serum concentration increases or decreases during acute or chronic inflammatory conditions. As such it has become a universal biomarker of infection and inflammation for a number of diseases and pathophysiological conditions. Built-in Quality Controls The LumiraDx Platform is integrated with several built-in quality control checks to ensure that the Instrument and Test Strips are functioning correctly for every test run. These checks include: • Electrical component operation, heater operation, battery charge state, mechanical actuators, sensors and optical system performance Test Strip positioning Test Strip expiry Monitoring of Test Strip performance and microfluidics controls during test runtime The CRP Test contains an Onboard Quality Control (OBC) assay Sufficient sample volume The LumiraDx CRP Test is an easy to use, fast microfluidic immunoassay designed to rapidly quantify CRP levels in whole blood and plasma. Hematocrit determination on the test strip The Test delivers rapid, quantitative results from a single direct fingerstick in 4 minutes at the point of care. The LumiraDx CRP Test is traceable to the ERM® - DA474/IFCC certified global reference materials. LumiraDx Multi Quality Controls come in two levels and are available from LumiraDx to complete Quality Control assessment of the Instrument and CRP Test Strips. Method comparison The method comparison was performed using 2 Test Strip lots with plasma samples (Lithium Heparin) sourced from patients presenting with symptoms of respiratory illness, inflammation or injury, at hospital Emergency Departments (ED), acute medical units or out-patient clinics. A comparison of 205 CRP measurements with the LumiraDx CRP Test to the RCRP Flex® assay on the Siemens Dimension® Xpand® Plus Integrated Chemistry System analysed by Passing Bablok regression yielded the following statistics: Slope = 1.00, Intercept = -0.48, r=0.99. Matrix Equivalency A study was carried out with 40 subjects presenting with symptoms of respiratory illness, inflammation or injury. Samples of capillary fingerstick blood (direct application and transfer tube) and paired whole blood (Lithium Heparin) and plasma (Lithium Heparin) samples were collected and tested and data was analysed by Passing Bablok regression. The results showed equivalency across all matrix types. A precision study was carried out in heparinised venous plasma on a protocol based on CLSI EP5-A3. The study was carried out at 3 concentrations of CRP, each was tested in 1 run of 5 replicates per day, for five days across 3 sites. The results of the precision study are summarised in the below table: Between Site Precision (% CV)

 Open the catalog to page 1
CRP Test Specifications-2

CRP Test Strip Specifications Displayed results Storage temperature Operating temperature Measuring range 5.0 - 250.0 mg/L Each laboratory should investigate the transferability of the expected values to its own patient population and, if necessary, determine its own reference ranges. Hook effect No hook effect is observed with the LumiraDx CRP Test at CRP concentrations up to 1044 mg/L. Minimum sample size Sample type • Direct fingerstick • Venous Whole Blood (Lithium Heparin) • Plasma (Lithium Heparin) For more information visit lumiradx.com or contact the LumiraDx Customer Services by email:...

 Open the catalog to page 2

All LumiraDx catalogs and technical brochures

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.