1. Catalogs
  2. Mar Cor
  3. CWP 100 RO System w/ Hot Water Loop Disinfection

CWP 100 RO System w/ Hot Water Loop Disinfection

CWP 100 RO System w/ Hot Water Loop Disinfection

CWP 100 RO System w/ Hot Water Loop Disinfection

Product catalog summary
Overview: The CWP 100 H Reverse Osmosis System is designed to deliver AAMI grade water for dialysis, adhering to ANSI/AAMI/ISO 13959:2009 standards. It offers a production capacity ranging from 5,000 to 13,000 gallons per day (GPD) and is engineered to meet current dialysis water quality standards.
Key Features:
  • Cost-effective and consistent water purification.
  • Semi-automated heat disinfection for the distribution loop and automated chemical disinfection for the RO and membranes to reduce biofilm and microbial contaminants.
  • Advanced electronics ensure quiet, reliable, and safe operation.
  • Standard features include hot water loop disinfection, a medical-grade Wirsbo-cleanPEX® distribution system, and direct feed operation.
  • Integrated with the Phoenix® Hemodialysis System for end-to-end heat management.
  • Monitoring software for documentation and verification of disinfection procedures.
  • 510(k)-cleared as a water purification system for hemodialysis and ETL certified.
Optional Components: Mar Cor Purification offers additional components such as blending valves, booster pumps, multi-media filters, water softeners, carbon adsorption units, hardness and total chlorine test kits, AAMI water analysis, and chemical disinfectants.
Specifications:
  • Product Water Output: Varies by model, ranging from 3.4 to 9.3 gallons per minute.
  • Quality: Achieves >95% rejection of total dissolved salts and >99% rejection of bacteria and endotoxins, with a water conversion factor of 67%.
  • Water Supply Requirements: Minimum input flow and pressure vary by model, with a temperature range of +5˚C to +25˚C.
  • Drain Requirements: Peak flow during rinse varies by model.
  • Reverse Osmosis Membranes: Made of modified polyamide, thin film composite, with a pH tolerance of 2–11.
  • Power Supply: Main voltage is 120/208 V, 60 Hz, 3-phase, with power ratings varying by model.
  • Dimensions and Weight: Dimensions are consistent across models, with weight varying from 1014 lbs to 1290 lbs depending on the model.
Technical Data: Additional space of 20 inches (500mm) is required on both sides of the unit for installation.
Contact Information: Mar Cor Purification provides contact details for various locations across the United States for customer support and inquiries.
See more

Catalog excerpts

CWP 100 RO System w/ Hot Water Loop Disinfection-1

Competent, Consistent, & Compliant Medical Device Reverse Osmosis System 5,000 to 13,000 GPD The CWP 100 H offers cost-effective, consistent delivery of AAMI grade water standards according to ANSI/AAMI/ ISO 13959:2009, water for dialysis at the point of use. This system is designed to meet all current standards for dialysis quality water. The semi-automated heat disinfection of the distribution loop and the automated chemical disinfection of the RO and membranes simply and easily reduce biofilm development and minimizes levels of endotoxin, fungi, yeast and other microbial contaminants. Greater cost-effectiveness is ensured through system efficiency and labor savings. The CWP represents a simple, convenient, and effective disinfection strategy. Advanced electronics and top quality components also ensure quiet, reliable, and safe operation. Standard Features Optional Items Hot water loop disinfection Mar Cor Purification offers a complete line of components and Medical-grade Wirsbo-cleanPEX® distribution system accessories, including: Direct feed operation Blending Valves and Booster Pumps Semi-automated disinfection of the loop and RO membrane Multi-Media Filters Integrated end-to-end heat with the Phoenix® Hemodialysis System Water Softeners Carbon Adsorption Monitoring software for documentation and verification of disinfection procedures Hardness and Total Chlorine Test Kits 510(k)-cleared as a water purification system for hemodialysis ETL certified AAMI Water Analysis Chemical Disinfectants

 Open the catalog to page 1
CWP 100 RO System w/ Hot Water Loop Disinfection-2

Technical Data CWP 100 H Specifications Product Water - Minimum capacity at 20˚C and at a pure water outlet pressure of 29 psi(200 kPa) Model Part Number Product Water in gallons/min (l/min) Depends on inlet water quality. The system is designed to utilize drinking water that meets the standards of the Safe Drinking Water Act, 40 CFR. Additional pretreatment of inlet water is required. A properly maintained system using inlet water meeting SDWA can obtain the following rejection rates: Total dissolved salts > 95 % • Bacteria (CFU) & endotoxins > 99 % • Water conversion factor 67 % Water Supply...

 Open the catalog to page 2

All Mar Cor catalogs and technical brochures

  1. SDS

    4  Pages

  2. MINNCARE HD

    2  Pages

  3. Minncare

    2  Pages

  4. WRO 300 H

    2  Pages

  5. Millenium HX

    2  Pages

  6. 23G Series RO

    2  Pages

  7. BioPure HX2

    2  Pages

  8. CWP

    2  Pages

  9. f

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.