1. Catalogs
  2. Markus Klotz GmbH
  3. Particle Counter TCC

Particle Counter TCC

Particle Counter TCC

Product catalog summary
Overview: The Particle Counter TCC is a control system designed for measuring the number and size of particles in liquids and gases. It is suitable for stationary use and monitors three customizable size classes. The system outputs particle counts as an analog signal or via RS232, with settable limits for pollution monitoring and alarm signaling.
Technical Specifications:
  • Display and Control: Features an LCD display for measurements and LEDs for status information. Control is possible through SPS or BUS-interface, with a keypad for manual operation.
  • Standby Mode: Can enter sleep mode via SPS or Bus-Command (RS232, CAN, PROFI).
  • Measuring Mode: Supports single, continuous, or interval measurements with programmable measuring and interval times.
  • Channels/Measuring Range: Three channels with freely selectable size classes.
  • Measurement Output: Outputs particle numbers as a power signal (0-20mA, 4-20mA) or via RS232, with options for CAN and PROFI-BUS.
  • Alarm Gauge: Adjustable for each channel, with output via relay contact or digital exit.
  • Sensor Control: Monitors readiness and pollution.
  • Power Supply: Requires 18-36VDC, max. 10 W.
  • Dimensions/Weight: 210 x 160 x 120 mm, weighing 1.5 kg.
Particle Sensors:
  • LDS 30/30 (water): Flow rate of 30 ml/min, measuring range of 0.9 - 139 µm, cell dimensions of 300 x 300 µm, and a maximum concentration of 120,000 particles/ml.
  • LDS 45/50 (oil): Flow rate of 50 ml/min, measuring range of 1.5 - 150 (400) µm, cell dimensions of 450 x 500 µm, and a maximum concentration of 25,000 particles/ml.
See more

Catalog excerpts

Particle Counter TCC-1

The particle counter TCC is a control system for measuring the number and size of particles in liquids and gases. The measuring system is suitable for the stationary use. The TCC monitors three size classes which can be chosen freely and transfers the number of particles either as an analog signal (Power f.e. 4..20mA) to a SPS or to the RS232. For every channel limits are settable, which monitor the degree of pollution and give out an alarm signal if need be. The measuring mode of the TCC focus only on time-related measurements, so that the flow rate through the laser sensor must be held sufficiently constant. (Flowcontroller o.s.). Standby-Mode Measuring Mode Measuring Time / Interval Time Channels / Measuring Range Measurement output Alarm gauge Sensor control Power supply Dimensions / Weight LCD-Display for measurements/ LEDs for status information, drain of measuring is also controllable through SPS or BUS-interface, Keypad for manual operation Sleep-Mode over SPS- or Bus-Command (RS232, CAN, PROFI) Single measurement, continuous or interval measurement Measuring -, Interval Times freely programmable Three measuring channels with freely eligible size classes Particle number as a power signal (0-20mA, 4-20mA), over RS 232 Option: CAN-, PROFI-BUS) Adjustable for every channel, output over relay contact or rather digital exit Readiness and pollution 18-36VDC, max. 10 W 210 x 160 x 120 (L x B x H)mm / 1.5kg Particle sensors Markus Klotz GmbH Theodor-Heuss-StraUe 23 75378 Bad Liebenzell Phone 07052/92336 Fax 07052/92338 E-mail [email protected]

 Open the catalog to page 1

All Markus Klotz GmbH catalogs and technical brochures

  1. PCSS Air Net

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.