T2

T2

T2

Product catalog summary
Overview
The document describes the Provaset T2, a compact and versatile instrument designed for leak testing using absolute pressure decay. It is an updated version of the Provaset 2P, maintaining compatibility with its programs and connections. The device is suitable for manual use on benches or integration into automatic systems managed by PLCs.

Specifications
  • Leak test capability up to 6 bar with a resolution of 1 Pa (0.01 mbar).
  • 3.5” color LCD touch screen display.
  • Supports up to 100 test programs with calculation of leak flow rate in cm³/min or cm³/h.
  • Digital I/O, RS232/RS485, USB for PC connectivity.
  • Test data collection via Ethernet and USB.
  • Available in 2 bar or 6 bar full scale models.
  • Manual precision pressure regulator with display indication.

Features
  • Staubli connector for Leak Master integration.
  • Automatic data collection via USB host.
  • Optimized dimensions and unpainted anodized aluminum case suitable for controlled environments like clean rooms and laboratories.

Accessories
  • AV10 pneumatic module for blood lines leak testing, which automatically starts and releases tests based on results.

Options (T2 PLUS)
  • Enhanced features for obstruction testing in continuous mode, providing immediate results with visual and audible alerts.

Technical Specifications
  • Power supply: 85 - 264 V∼, 50-60 Hz, 18W; 24Vcc ± 10 %, 1.5 A.
  • Compressed air line: Dry, non-condensing, 5-micron filtered and oil-free air, compliant with ISO8573-1.
  • Keyboard: Touch screen with manual Start/Stop button.
  • Indicators: “Passed” and “Failed” result LEDs.
  • Audio alert: Built-in beeper.
  • Interfaces: Modbus RTU - TCP/IP, CVS ASCII, CSV.

Dimensions
  • Provaset T2: 180 mm x 135 mm x 160 mm.
See more

Catalog excerpts

T2-1

Provaset T2 COMPACT EQUIPMENT FOR LEAK TESTS BY ABSOLUTE PRESSURE DECAY OPTIMIZED DIMENSIONS AND UNPAINTED ANODIZED ALUMINIUM CASE, MAKE IT SUITABLE FOR USE IN CONTROLLED ENVIRONMENT SUCH AS CLEAN ROOM AND LABORATORIES 3.5” colour LCD display, touch screen Up to 100 test programs tables Test recording via Ethernet and USB Backward compatibility of programming with Provaset 2P DESCRIPTION PROVASET T2 is a compact, versatile limited areas but it could be integrated Provaset T2 is available with 2 bar and extremely reliable instrument on automatic systems managed by or 6 bar full scale model, with 1 Pa that applies the latest electronic and resolution on leak reading. pneumatic technologies to offer the Provaset T2 is able to communicate A Staubli connector is available to best performances. through digital I/O, Ethernet, USB host/ connect a Leak Master. This new Provaset T2 updates the slave and RS232/RS485 serial line The test pressure can be regulated with previous Provaset 2P model, with a manual precision pressure regulator; which mantains the program and The test data collection is possible on the pressure of regulation is shown on connection compatibilities. Provaset T2 USB memory and via Ethernet. is designed for manual use on bench in Ethernet port for instrument management and data collection Staubli connector on front panel 100 test programs, calculation of the leak flow rate Q in cm³/min or cm³/h Test log available on the instrument Automatic pendrive data collection (through USB host) Certificated Leak Master to be inserted in the Staubli® connector Remote control keypad

 Open the catalog to page 1
T2-2

Provaset T2 SPECIFICATIONS Power supply Dry, non-condensing, 5-micron filtered and oil-free air, compliant with ISO8573-1 Test pressure Measurement area: 0÷2 bar, resolution 1 Pa; 0÷6 bar, resolution 1 Pa Touch screen Manual Start/Stop button 3.5” colour TFT LCD touch screen display “Passed” result led, “Failed” result led Test counter PASS and REJECT totals, resettable to zero Audio alert Built-in beeper Programmable parameters Up to 100 test programs (T2 PLUS) 4 photocoupled inputs and 4 photocoupled outputs Serial Lines Modbus RTU - TCP/IP, CVS ASCII, CSV Staubli® Connector For Leak Master...

 Open the catalog to page 2

All MARPOSS catalogs and technical brochures

  1. SOFTWARE

    2  Pages

  2. DOSASET

    2  Pages

  3. LTC

    2  Pages

  4. LEAK MASTER

    2  Pages

  5. TM3P

    2  Pages

  6. TM3PF

    2  Pages

  7. TM3PD

    2  Pages

  8. T3LPQ

    2  Pages

  9. T3LPF

    2  Pages

  10. T3LP

    2  Pages

  11. T3LMF

    2  Pages

  12. T3LD

    2  Pages

  13. 2SEAL

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.