MaxVenturi Overview
The MaxVenturi is designed to interface with high-flow humidification systems, utilizing the Venturi principle to mix ambient air and oxygen without requiring piped medical-grade air. It is ideal for adult high-flow oxygen therapy in various environments.
Part Numbers
Various configurations are available, including options with pole mounts and locking pole mounts, compatible with DISS and NIST standards, for both domestic and international use.
Specifications
  • Measurement Range: 0 to 100%
  • Display Resolution: 0.10%
  • Accuracy and Linearity: 1% of full scale at constant conditions
  • Total Accuracy: ±3% over full operating temperature range
  • Response Time: 90% of final value in ~15 seconds at 23˚C
  • Warm-up Time: None required
  • Operating Temperature: 15˚C to 40˚C (59˚F to 104˚F)
  • Storage Temperature: -15˚C to 50˚C (5˚F to 122˚F)
  • Atmospheric Pressure: 800 to 1013 mBars
  • Humidity: 0 to 95% (non-condensing)
  • Power Requirements: Two AA alkaline batteries
  • Battery Life: ~5,000 hours continuous use
  • Low Battery Indication: Battery icon on LCD
  • Sensor Type: Maxtec MAX-250 Series galvanic fuel cell
  • Expected Sensor Life: > 1,500,000 O2 percent hours (approx. two years)
  • Dimensions: 7.3˝x 6.6˝x 5.3˝ (185 mm x 167 mm x 135 mm)
  • Weight: 2.54 lbs (1152 grams)
  • Drift of Measurement: < +/-1% of full scale at constant conditions
Key Features
  • Flow Meter: Independent adjustable O2% and flow
  • Built-in Pole Clamp: For easy transport or storage
  • Room Air Inlet: Eliminates need for piped medical air
  • MAX-250E Sensor: Long-life sensor
  • Flow Control Valves: Air/O2 control valves
  • Built-in Analyzer: Displays accurate O2 measurements
  • On/Off Button: One-touch power control
  • One-Touch Calibration: Auto calibration to room air or 100% oxygen
  • Delivered Gas Outlet: For high-flow system interface
See more

Catalog excerpts

MaxVenturi®-1

MaxVenturi® SENSING Pole Mount Adjustment Knobs MAXVENTURI OVERVIEW PART NUMBERS The MaxVenturi has been specifically designed to interface w/ Pole Mount (NIST). . . . . . . . . . . . . R211P03-021 with high-flow humidification systems. Using the well- w/ Locking Pole Mount (DISS). . . . R211P03-022 established Venturi principle, this unit allows users to mix w/ Locking Pole Mount (NIST) . . . R211P03-023 ambient air and oxygen without the need for piped medicalgrade air. This is the perfect solution for adult high-flow oxygen therapy in almost any environment. w/ Pole Mount (DISS). . . . . . . . . . . . R211P03-020 INTERNATIONAL w/ Pole Mount (DISS). . . . . . . . . . . . R211P03-024 w/ Pole Mount (NIST). . . . . . . . . . . . R211P03-025 w/ Locking Pole Mount (DISS). . . . R211P03-026 w/ Locking Pole Mount (NIST) . . . R211P03-027

 Open the catalog to page 1
MaxVenturi®-2

1. Flow Meter Independent adjustable O2% and flow 2. Built-in Pole Clamp For easy transport or storage 3. Room Air Inlet No need for piped medical air 4. MAX-250E Sensor Long-life sensor 5. Flow Control Valves Air/02 control valves 6. Built-in Analyzer Displays accurate 02 measurements 7. On/Off Button One-touch on and off power 8. One-Touch Calibration Auto cal. to room air or 100% oxygen 9. Delivered Gas Outlet For high-flow system interface Specifications Measurement Range Dispaly Resolution Accuracy and Linearity when Total Accuracy Response Time Warm-up Time Operating Temperature Storage...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.