Client Information: The report is prepared for Natalia Papaja-Liczberska from Eurofins Polska Sp. z o.o., with the sample described as medical masks SAFE Type IIR, BFE≥98% ES/RS, manufactured by Medical Protection Sp. z o.o. Sp.K.
Sample Information: The sample was received on 5-Oct-2020, and the analysis was conducted from 5-Oct-2020 to 13-Oct-2020. The masks are tested according to EN 14683:2019+AC:2019 TYPE IIR standards.
Testing and Results: The primary test conducted was the resistance against penetration by synthetic blood, following ISO 22609:2004. All 32 samples passed the test at a pressure of 16 kPa, meeting the Acceptable Quality Limit (AQL) of 4.0%.
Conclusion: The masks passed the resistance test against synthetic blood, confirming compliance with the required standards for Type IIR masks.
Technical Details: The test involved 32 specimens with a sample size of 5.58 cm in diameter, tested at a pressure of 16 kPa. The environmental conditions were maintained between 16.7ºC and 26ºC with humidity between 82.8% and 88%.
Standards and Requirements: According to EN 14683:2019+AC:2019, Type IIR masks require a bacterial filtration efficiency (BFE) of ≥98%, differential pressure of <60 Pa/cm2, splash resistance pressure of ≥16 kPa, and microbial cleanliness of ≤30 CFU/g.
Additional Notes: The report is electronically validated by Maria Jesus Martinez Puig, Chemical Lab Manager at Eurofins Textile Testing Spain, S.L. The test was subcontracted within the Eurofins group and is accredited.
See more

Catalog excerpts

EUROFINS Laboratory report of waterproofing tests-1

Page: 1 / 6 Analytical Report Nr. ANALYTICAL REPORT Client Information Eurofins Polska Sp. z o.o. ul. Księcia Ziemowita 53 blok 3A lok. 4 WARSZAWA POLAND [email protected] For the attention of Natalia Papaja-Liczberska Sample Information Order Code: Reception Date: Analysis Starting Date: Analysis Ending Date: Information provided by the customer: 720-2020-00199226 Maski medyczne SAFE Typ IIR, BFE≥98% ES/RS 720-2020-00199226 Maski medyczne SAFE Typ IIR, BFE≥98% ES/RS, manufacturer: Medical Protection Sp. z o.o. Sp.K. Client Reference: Sample Description: Testing conducted on behalf of: Medical Protection Sp. z o.o. Sp.K. ul. Biały Kamień 2/U6 Warszawa 02-593 EN 14683:2019+AC:2019 TYPE IIR Customer requirements: Purchase Order Number: Decision rule Shared risk - Simple acceptance. Eurofins Textile Testing Spain, S.L. Calle Germán Bernácer, 4 03203 Elche SPAIN Phone+34 966 299 638 www.eurofins.com/tex

 Open the catalog to page 1
EUROFINS Laboratory report of waterproofing tests-2

Page: 2 / 6 Analytical Report Nr. SAMPLE PICTURE Eurofins Textile Testing Spain, S.L. Calle Germán Bernácer, 4 03203 Elche SPAIN Phone+34 966 299 638 www.eurofins.com/tex

 Open the catalog to page 2
EUROFINS Laboratory report of waterproofing tests-3

Page: 3 / 6 Analytical Report Nr. CONCLUSION: TEST PROPERTY Resistance against penetration by synthetic blood ISO 22609:2004 A - Mask Test has been performed as per application request Eurofins Textile Testing Spain, S.L. Calle Germán Bernácer, 4 03203 Elche SPAIN Phone+34 966 299 638 www.eurofins.com/tex

 Open the catalog to page 3
EUROFINS Laboratory report of waterproofing tests-4

Page: 4 / 6 Analytical Report Nr. COMPONENT LIST: COMPONENT ID COMPONENT NAME Eurofins Textile Testing Spain, S.L. Calle Germán Bernácer, 4 03203 Elche SPAIN Phone+34 966 299 638 www.eurofins.com/tex MATERIAL DESCRIPTION Mask COLOR Blue

 Open the catalog to page 4
EUROFINS Laboratory report of waterproofing tests-5

Page: 5 / 6 Analytical Report Nr. Sample code Nr. Date MASKS TESTING Analyses on:A - Mask Analysis Ending Date: 13/10/2020 Resistance against penetration by synthetic blood ISO 22609:2004 Result: Complete test data reported at Annex. AQL information according to ISO 22609:2004: A single sampling plan providing an AQL of 4,0 % requires 32 test specimens. An AQL of 4.0 % is met for a single sampling plan when 29 or more specimens show pass results. AQL= Acceptable Quality Limit. Eurofins Textile Testing Spain, S.L. Calle Germán Bernácer, 4 03203 Elche SPAIN Phone+34 966 299 638 www.eurofins.com/tex...

 Open the catalog to page 5
EUROFINS Laboratory report of waterproofing tests-6

Page: 6 / 6 Analytical Report Nr. Signed for and on behalf of Eurofins Textile Testing Spain: Report electronically validated by Maria Jesus Martinez Puig Chemical Lab manager EXPLANATORY NOTE u Test not covered by ENAC accreditation scope l Test is subcontracted within Eurofins group and is accredited ¢ Test is subcontracted within Eurofins group and is not accredited n Test is subcontracted outside Eurofins group and is accredited o Test is subcontracted outside Eurofins group and is not accredited Eurofins General Sales Terms and Conditions applied. Results obtained refer only to samples,...

 Open the catalog to page 6
EUROFINS Laboratory report of waterproofing tests-7

Test report Annex Analytical Report Nr. AR-20-YL-006336-01 Sample code Nr. 560-2020-00007013 DETERMINATION RESISTANCE AGAINST PENETRATION BY SYNTETIC BLOOD Test Method: ISO 22609:2004; Targeting-plate test method Number of test specimens: 32 Sample size: Circular, diameter 5,58 cm Sample area tested: 24,5 cm2 Pressure: 16 kPa (120,0 mm Hg) Stream velocity of synthetic blood: 550±10 cm/s Distance of the face mask target area surface from the tip of the cannula: 30,5 cm Angle of the pneumatic valve with respect to the face mask target area: 90º Technique used to enhance visual detection of synthetic...

 Open the catalog to page 7
EUROFINS Laboratory report of waterproofing tests-8

Test report Annex Analytical Report Nr. AR-20-YL-006336-01 Sample code Nr. 560-2020-00007013 Operating requirements for surgical masks based on EN 14683: 2019+AC: 2019 standard Bacterial filtration efficiency (BFE), (%) Differential pressure (Pa/cm 2) Splash resistance pressure (kPa) Microbial cleanliness (CFU/g)

 Open the catalog to page 8
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.