Overview: The Mobile Dental Generator is designed to provide precision diagnostics for dental care on-the-go, combining power, portability, and performance.
System Technical Specifications:- Power: 530 VA at 110V, ≤0.4 Ohm resistance.
- Tube Head: Operates at 110~240 V ± 10%, 50/60 Hz, with absorbed current of 2.1 A at 240V and 5.2 A at 110V.
- Power: 500 VA at 240V, ≤0.8 Ohm resistance.
- Fuses: F1 - F 6.3 A / 250 V, F2 - ETF- 6.3 A / 250V.
X-ray Tube Specifications:- High Voltage: Selectable 60/65/70 kV with ± 10% accuracy.
- Anodic Current: 5 mA with ± 20% accuracy.
- Exposure Time: 0.01 – 2.0 s with ± 10% +1 ms accuracy.
- Maximum Load: 70 kV, 5 mA, 2 S.
- Operation: High frequency DC, momentary operation up to 2 seconds.
- Focal Spot Size: 0.4 mm (IEC 336).
- X-ray Field: Ø < 60 mm with options for 22×32 mm and 32×42 mm.
- Minimum Focus Skin Distance: 20 cm.
- Leakage Radiation: < 0.25 mGy/h at 1 m.
Weight and Dimensions:- Gross Weight: 80 kg including packing.
- Net Weight: 55 kg in standard configuration.
- Components: X-ray head (6.3 kg), Control unit (1.8 kg), Double pantograph arm (10.4 kg), Extension arms (3.2 kg and 3.9 kg).
Operating Conditions:- Temperature: +10 °C to +40 °C.
- Relative Humidity: 30% to 75%.
- Pressure: 700 to 1060 hPa.
Wall-Mount Dental Generator: Shares the same specifications as the Mobile unit, with differences in weight distribution for components like the wall plate and extension arms.
See more