Product catalog summary
Instructions for Use: The AIR FREE STANDARD HEAD 4 HOLE BACK END (Item code: AF90S) is a dental device restricted to licensed professionals. Users must read the manual thoroughly before use and keep it for future reference.
Warnings: Only use burs from reputable manufacturers with specified dimensions. Ensure burs are not bent or damaged and are correctly seated in the chuck. Do not operate the handpiece without a bur, and replace the turbine if the handpiece becomes noisy or sluggish.
Specifications: The device operates at an air pressure of 35-40 psi and a speed of 400,000 rpm. It features a fixed back end coupling type.
Installation: Follow the provided figure for proper installation of the handpiece.
Changing Bur: To remove a bur, push the end cap with your thumb. To mount a bur, insert it fully into the head and ensure it is secure by tugging firmly.
Cartridge Replacement: Remove the cartridge by mounting a bur, turning the head cap counter-clockwise, and pushing the bur. Clean the inside with spray lubricant. To mount, insert the cartridge into the head-cap and secure it by threading the head-cap into the handpiece.
Cleaning & Lubrication: Remove the handpiece from the coupling, secure a bur, and clean with a soft cloth or brush. Do not immerse in water or use ultrasonic baths. Use a needle to clear debris from the spray hole and spray lubricant into the air drive tube. Operate the handpiece to purge debris.
Sterilization: Remove the bur, place the handpiece in a sterilization pouch, and steam autoclave for 15 minutes at 132ºC. Allow a minimum of 20 minutes for drying.
Warranty: The product is warranted against manufacturing defects for one year. The warranty covers repair or replacement but excludes damages from misuse or unauthorized alterations.
Repair: For repairs, provide contact information and nature of repair to Medidenta, Las Vegas, NV.
Replacement Parts: Various parts such as the Air-Free Standard Head Cartridge and Cap Wrench are available for replacement.
See more

Catalog excerpts

Air Free 90 Ortho – Standard Head-1

Read this manual before use. Keep it in a safe place for future reference. A WARNING Federal (US) law restricts the use of this device to licensed professionals. 1. Only use burs of reputable manufacturers with a shaft diameter of 1.5911.60mm (.06251.0629"), a total length of max. 26mm (1.02"), and head diameters of up to 2mm (.0787"). Burs which deviate from the stated measurements must not be used 2. Before each use, make sure that the bur is not bent or damaged. 3. Before each use, make sure that the bur is correctly seated and locked in the chuck. 4. Do not run the handpiece without a bur or bur blank locked in the chuck. 5. When handpiece becomes noisy or sluggish after extended length of use, replace with a new turbine. 6. Do not push the head cap button during rotation. 1. Installation Follow (Fig. 1) below to install the handpiece. Push on end cap with thumb to click, and remove the bur (Fig. 3) Fig. 2 Bur mounting Push on end cap and insert the bur into the head until it stops (5mm deep approx.) release pressure on button bur should be secure. Firmly tug on the bur to insure that it is properly secured. Caution: Always insert the bur full length of the chuck. If you attempt to extend it from the chuck, the bur may dislodge during use and cause injury to you or your patients. 3. Cartridge Replacement Removing cartridge (Fig. 3&4) Mount a bur in the chuck. Turn the head cap counter-clockwise with the cap wrench (Fig. 3), and remove the head cap. Push on the bur and the cartridge is easily removed from the back of the head (Fig. 4). Clean inside the handpiece head with spray lubricant. Mounting Cartridge (Fig. 5&6) Check for all components inside head (Fig. 5) Gently insert the rear end of Cartridge into the Head-Cap, then hold the Cartridge by the Head-Cap, bring Head of handpiece down over the top of the chuck pin and thread the Head-Cap into the Head of handpiece, as shown on (Fig. 6) 1. Remove the handpiece from the coupling. 2. Insert bur into the chuck of the handpiece and secure. 3. Remove any adherent material using only a soft cloth, soft brush, gauze, or cotton swab that has been immersed in a soapy solution. Do not use a wire brush. Wipe and dry thoroughly. Do not use ultrasonic bath, nor immerse the handpiece in water or any solution. 4. Use needle to remove any debris that may clog the spray hole (Fig. 8) ' ' “ '

 Open the catalog to page 1
Air Free 90 Ortho – Standard Head-2

5. Insert nozzle tip into air drive tube on the back end of the handpiece and spray for approximately 3 seconds. (Fig. 9) Fig. 9 6. Fully depress the button of the end cap and hold the collet open then plunge the bur in and out several times to distribute the cleaner into the chucking mechanism. Fully insert the bur, release the pressure on end cap button to secure the bur. Rotate the bur back and forth between your fingers to further distribute the cleaner & lubricant in the head of the handpiece to help loosen any debris. 7. Attach handpiece to an air supply to operate the handpiece. Before...

 Open the catalog to page 2